Table 2 Modeling and evaluation of factors from GAB and MET calibration curves.

From: Green and white analytical approach for parallel quantification of gabapentin and methylcobalamin in medicinal products using inventive RP-HPLC technique

Parameter

Drugs

GAB

MET

Wavelength

210 nm

210 nm

Range (µg/mL)

3–50

3 - 50

Coefficients of determination (R2)

0.9998

0.9998

Slope

1.4487

102.66

Intercept

−0.4668

−66.369

LOD

0.80 µg/mL

0.60 µg/mL

LOQ

2.50 µg mL

2.00 µg/mL

  1. a Limit of detection (3.3× σ/Slope) and a limit of quantitation (10× σ/Slope).