Table 1 \(k_{\text {axis}}\) values for each probe.

From: Error analysis of a heating oven for a spin-exchange relaxation-free magnetometer

Type

\(k_x\)

\(k_{y_1}\)

\(k_{y_2}\)

\(k_z\)

Exp.

22.30

16.21

-12.53

-10.37

S.w.w.

13.27

10.82

-8.64

3.18

S.o.w.

0.16

-0.13

-0.24

0.33

  1. Exp.: Experimental result; S.w.w.: Simulation with twisted wire pair; S.o.w.: Simulation without wires. All units are in \(\mathrm {pT/mA}\).