Table 4 The results of subjective image analysis comparing different image protocols.

From: Image quality and radiation dose of reduced-dose abdominopelvic computed tomography (CT) with silver filter and deep learning reconstruction

 

Standard dose with IR

SilverBeam with DLR

SilverBeam with DLR + CE-boost

P value

P1 vs. P2

P1 vs. P3

P2 vs. P3

Observer 1

  Overall image quality

4.91 ± 0.28

4.96 ± 0.17

4.62 ± 0.58

0.001

0.320

0.001

0.001

  Image noise

4.78 ± 0.41

4.84 ± 0.39

4.84 ± 0.39

0.510

0.570

0.570

0.999

  Artifact

4.77 ± 0.42

4.94 ± 0.22

4.44 ± 0.82

0.001

0.070

0.001

0.001

  Diagnostic acceptability

4.98 ± 0.10

4.99 ± 0.10

4.70 ± 0.50

0.001

0.999

0.001

0.001

Observer 2

  Overall image quality

4.93 ± 0.24

4.94 ± 0.22

4.94 ± 0.22

0.910

0.980

0.980

0.990

  Image noise

4.92 ± 0.26

4.90 ± 0.29

4.91 ± 0.29

0.880

0.990

0.990

0.999

  Artifact

4.95 ± 0.20

4.94 ± 0.20

4.93 ± 0.24

0.750

0.999

0.811

0.783

  Diagnostic acceptability

4.95 ± 0.22

4.93 ± 0.24

4.93 ± 0.23

0.960

0.970

0.970

0.999

  1. Data are reported as mean ± standard deviation. IR iterative reconstruction; DLR deep learning reconstruction; CE contrast-enhancement.