Table 1 Key questionnaire results pooled across study days.

From: Benchmarking the utility of dry-electrode electroencephalography for clinical trials

Device

Set-up (technicians)

Clean-up (technicians)

Comfort (participants, rating 0–10)

Time (min)

Ease (rating 0–10)

Time (min)

Ease (rating (0–10))

Standard EEG

23.57 (21.46–25.81)

6.65 (6.28–7.03)

16.59 (14.99–18.21)

5.24 (4.87–5.65)

8.22 (8.02–8.41)

DSI-24

10.98 (9.19–13.46)

8.66 (8.38–8.93)

3.6 (3.05–4.21)

9.06 (8.66–9.35)

7.86 (7.66–8.06)

Quick-20r

17.42 (15.04–20.01)

6.44 (5.85–6.98)

4.24 (3.55–5.03)

8.86 (8.63–9.09)

5.98 (5.67–6.29)

zEEG

15.48 (13.04–17.98)

6.87 (6.32–7.43)

3.82 (3.17–4.46)

8.67 (8.37–8.96)

4.88 (4.54–5.18)

  1. Data shown reflect the mean and the 95% confidence level.