Table 2 SNR as captured by the GFP and amounts of stimulus information during peak and later trial periods.
From: Benchmarking the utility of dry-electrode electroencephalography for clinical trials
Device | 100 ms around P300 peak | 300–600 ms post stimulus | ||
|---|---|---|---|---|
GFP (95% CI) | AUC (95% CI) | GFP (95% CI) | AUC (95% CI) | |
Standard EEG | 86.26 (84.39–88.18) | 0.70 (0.68–0.72) | 53.82 (52.95–54.70) | 0.66 (0.64–0.67) |
Std. EEG (1020) | 98.92 (96.39–101.40) | 0.68 (0.67–0.70) | 67.63 (66.25–69.01) | 0.65 (0.64–0.66) |
DSI-24 | 118.39 (114–122.29) | 0.69 (0.68–0.7) | 71.54 (70.02–73.29) | 0.65 (0.64–0.66) |
Quick-20r | 108.2 (105.42–110.84) | 0.7 (0.68–0.72) | 57.66 (56.16–59.08) | 0.63 (0.62–0.65) |
zEEG | 81.23 (79.13–83.33) | 0.69 (0.68–0.7) | 67.36 (65.99–68.78) | 0.61 (0.6–0.62) |