Table 3 Average latency and amplitude features for the P300 per device.

From: Benchmarking the utility of dry-electrode electroencephalography for clinical trials

Device

P300 latency (95% CI)

P300 amplitude (95% CI)

Standard EEG

0.31 s (0.30 s–0.32 s)

5.46 μV (4.88 μV–6.28 μV)

DSI-24

0.32 s (0.31 s–0.33 s)

6.81 μV (6.16 μV–7.50 μV)

Quick-20r

0.31 s (0.30 s–0.33 s)

8.01 μV (7.13 μV–8.91 μV)

zEEG

0.32 s (0.30 s–0.33 s)

6.54 μV (5.88 μV–7.26 μV)