Table 1 Parameters for eFP data simulations. The notation of the aberration parameters follows that given by Saxton52. Positive \(\:{C}_{1}\) values are defined as overfocus.

From: Electron fourier ptychography for phase reconstruction

 

Gold particle

Apoferritin

Cry11Aa

Simulation box size (nm3)

13.5 × 13.5 × 0.8

102.4 × 102.4 × 30.0

153.6 × 153.6 × 30

Slice thickness (nm)

0.14

1.00

1.00

Number of slices

6

31

31

Accelerating voltage (kV)

300

300

300

Illumination semiangle (mrad)

0.02

0.02

0.02

Focal spread (nm)

4.3

8.5

8.5

\(\:{C}_{1}\) (nm)

2.5

−2000

−2172

\(\:{C}_{3}\) (mm)

−1.9 × 10−5

2.7

2.7

Pixel size (nm/pixel)

0.013

0.05

0.25

Tilt magnitudes (mrad)

10 and 15

5

1.9

Total fluence (e/nm2)

4.6 × 105

1 × 103, 3 × 103, 9 × 103, and infinite fluence

4.5 × 102

Number of tilts

7

4, 5, 9, and 13

4