Table 3 Results of the injections for the evaluation of the accuracy of the recovery of the QC negative sample spiked at the LOQ concentration.

From: A proposed limit test for p-chloroaniline impurity in paracetamol pharmaceutical formulations

Accuracy of LOQ

QC negative sample

Spike (ppb)(ng/mL)

Range

Calculated conc. ppb (ng/mL)

Recovery

1

5

100%

5.367

107.34%

2

5

5.221

104.42%

3

5

5.413

108.26%

Average

106.98

Standard deviation

0.0188

RSD%

1.88