Table 2 Specifications of wireless biolectronic actuator.

From: Development of a wireless bioelectronic actuator for wound healing in a porcine model

Parameter

Specifications

Channels

2\(\times\) 4-actuation (I-DACs), 2\(\times\) 8-sensing (ADCs)

Applied voltage \(V_i\)

0 to 4.7 V

Channel current \(I_{i}\)

0 to 50 \(\mu\)A

Actuation resolution

8-bit I-DACs (0.2 \(\mu\)A)

Sensing resolution

12-bit ADCs (1.1 mV)

Averaging factor

16 (0.5 ms apart)

Sampling period between averages

100 ms

Power management

Active/light sleep/standby

Typical delivery duration

24 h per day (up to 7 d)

microSD card size

4 to 32 GBytes

Current sensing resistors \(R_{i}\)

10 k\(\Omega \pm\)0.1%