Table 2 Analytical performance parameters of the developed spectrofluorimetric method including linearity range, regression equation, correlation coefficient, and limits of detection and quantification.

From: Validated Erythrosin B spectrofluorimetric method for ganciclovir bioanalysis in rabbit plasma following valganciclovir bioconversion and Pharmacokinetic application

Parameter

Ganciclovir

Linearity range (µg/mL)

0.05–3.0.05.0

Intercept (a)

1.0749

Slope (b)

1.4373

Coefficient of determination (r2)

0.9991

SE of intercept (Sa)

0.0269

SE of slope (Sb)

0.0188

LOD (µg/mL)

0.0157

LOQ (µg/mL)

0.0472

LLOQ (µg/mL)

0.05

ULOQ (µg/mL)

3.0