Table 2 Comparison of chlorine detection performance (LOD/LOQ) between the ultra-low energy MWE-LIBS system developed in this work and previously reported LIBS and MWE-LIBS methods.
Condition | LOD/LOQ, wt % | |
|---|---|---|
This Manuscript | Ultra-Low Energy MWE-LIBS, 0.05 mJ | 0.204 ± 0.068 / 0.679 ± 0.204 |
From References | LIBS, 3 mJ14 | 0.87 ± 0.07 / ** |
Standard Method (MWE-LIBS), 3 mJ16 | 0.17 ± 0.02 / ** | |
German standard DIN 32645*, 3 mJ (MWE-LIBS)14 | 0.03/0.11 | |
Global Interlaboratory Experiments (LIBS), 3.8–25.5 GW/cm213 | 0.01–0.10/0.03–0.33 | |
Jiaotong University, China (DP- LIBS) 15 mJ + 30 mJ41 | 0.19–0.26/ ** |