Table 4 Statistical parameters for analysis of OC and GU binary mixture at pH (7) using DPV at SPMWCNTS.

From: Green analytical method for determination of oxeladin citrate using advanced electrochemical modified sensors in pure form, pharmaceuticals, human serum and binary mixtures with guaifenesin

Parameter

Values*

OC

GU

Linear range conc (µg/mL)

0.63–4

0.33–3.13

Intercept

0.2459

0.2337

Slope

0.1317

0.3249

R2

0.99

0.99

LOD(µg/ml)

0.27

0.06

LOQ(µg/ml)

0.63

0.20

  1. * Obtained from an average of six experiments.