Table 2 Time to first fixation (TFF) based on eye-tracking metrics for tasks showing significant differences in at least one comparison between devices.

From: Qualitative and quantitative evaluation of the usability of transport ventilators using eye tracking

Tasks

Monnal T60

Oxylog 3000 plus

Hamilton-T1

Medumat Transport

 

[s]

[s]

[s]

[s]

Turn on the device

18.0 [14.4]

0.8 [0.4]

1.0 [0.8]

0.8 [0.4]

Set VCV mode

8.1 [5.0]

3.9 [2.8]

1.7 [1.9]

7.9 [7.6]

Set vent. Frequency to…

1.2 [1.0]

1.2 [1.0]

13.9 [13.0]

1.8 [0.9]

Set PEEP to…

1.5 [1.5]

9.8 [8.8]

1.8 [1.2]

1.6 [1.1]

Set peak pressure to…

43.5 [25.5]

2.3 [1.8]

44.3 [24.1]

1.5 [0.7]

Set I:E ratio to…

16.0 [17.2]

2.7 [2.4]

6.4 [2.9]

2.1 [0.9]

Read out Ppeak

7.2 [8.8]

19.0 [13.2]

3.4 [2.4]

12.2 [26.0]

Read out exp. MV

3.9 [2.6]

8.2 [4.7]

0.8 [0.8]

3.8 [3.1]

Set PSV mode

7.7 [1.8]

3.2 [1.8]

11.9 [10.7]

10.4 [6.2]

Open alarm menu

1.5 [2.8]

3.0 [1.7]

1.8 [1.1]

8.8 [4.6]

Set alarm volume to…

21.3 [12.3]

20.8 [19.4]

34.9 [25.8]

17.5 [11.7]

Read out end-tidal CO2

3.4 [5.4]

9.7 [1.8]

2.7 [0.8]

2.5 [1.6]

Read out exp. VT

12.1 [6.4]

10.6 [2.9]

1.5 [1.7]

3.6 [2.8]

Turn off the device

7.8 [12.6]

1.8 [2.2]

1.9 [1.1]

1.7 [0.9]