Table 1 Detection sensitivity, specificity, and accuracy of QIAamp DNA blood Mini kit, K-SL DNA extraction kit, and GraBon™ methods for Escherichia coli and Staphylococcus aureus.

From: Comparison evaluation of bacterial DNA extraction methods for improved molecular diagnostic accuracy of sepsis-causing pathogens in clinical whole blood samples

Pathogen

Method

Sensitivity (%)

(95% CI)

Specificity (%)

(95% CI)

Accuracy (Positive/Total)

E. coli

QIAamp DNA Blood Mini Kit

30 (16.56–46.53)

100 (91.19–100.0)

65.0% (12/40)

K-SL DNA Extraction Kit

55 (38.49–70.74)

100 (91.19–100.0)

77.5% (22/40)

GraBon™

52 (36.13–68.49)

100 (91.19–100.0)

76.5% (21/40)

S. aureus

QIAamp DNA Blood Mini Kit

35 (20.63–51.68)

100 (91.19–100.0)

67.5% (14/40)

K-SL DNA Extraction Kit

35 (20.63–51.68)

100 (91.19–100.0)

67.5% (14/40)

GraBon™

55 (38.49–70.74)

100 (91.19–100.0)

77.5% (22/40)