Table 1 Acquisition protocols.

From: Comparison of SS-EPI DWI and one-minute TGSE-BLADE DWI for diagnosis of acute infarction

Parameter

SS-EPI DWI

TGSE-BLADE DWI

T2WI

b value (s/mm2)

0, 1000

0, 1000

NA

TR (ms)

3900

3300, 3200*

3540

TE (ms)

63, 71*

46, 62*

79

FA (degrees)

NA

120

120

FOV (mm)

220 × 220

220 × 220

220 × 220

Matrix

160 × 160

160 × 160

448 × 448

Slice thickness (mm)

5

5

5

Number of slices

22

22

22

Voxel size (mm3)

1.4 × 1.4 × 5.0

1.4 × 1.4 × 5.0

0.5 × 0.5 × 5.0

Bandwidth (Hz/pixel)

1202

520

189

NEX

2

1

1

Parallel imaging

(Phase Encoding × Slice Encoding)

GRAPPA

3 × 1

GRAPPA 2 ×

Slice acceleration 2

GRAPPA

3 × 1

Turbo factor

NA

13

11

EPI factor

128

3

NA

Acquisition time (s)

52

59

51

  1. * Parameters are for MAGNETOM Skyra.