Table 3 Incidences of assembly defects in reusable medical devices in control and experimental groups.

From: Effect of hazard analysis critical control points system in the assembly of reusable medical devices

Groups

The number of device packages

The number of device packages with assembly defects

Rate of assembly defects %

The control group

35,841

929

2.59

The experimental group

36,159

87

0.24

  1. X2 = 715.365; P < 0.001.