Table 4 Types of assembly defects in reusable medical devices in control and experimental groups.
From: Effect of hazard analysis critical control points system in the assembly of reusable medical devices
Groups | The number of device packages | Types of assembly defects | |||||
|---|---|---|---|---|---|---|---|
Device model number error | Dressing quantity error | Dressing type error | Device quantity error | Device function abnormality | Placement error | ||
The control group | 35,841 | 465 | 291 | 95 | 47 | 18 | 13 |
The experimental group | 36,159 | 51 | 20 | 8 | 6 | 1 | 1 |
X2 | 338.261 | 239.584 | 74.364 | 32.104 | 13.619 | 8.742 | |
P | < 0.001 | < 0.001 | < 0.001 | < 0.001 | < 0.01 | < 0.05 | |