From: Clinical evaluation of the installation conditions and accuracy of wearable neonatal jaundice meter

Fig. 8

Measurement error in the serum bilirubin concentration between the wearable transcutaneous jaundice meter and JM-105. Average=TSB+device/2. Difference=TSB−device. (a) Center of the forehead with 1 N of applied pressure. (b) Center of the forehead with 2 N of applied pressure. (c) Upper eyebrow with 1 N of applied pressure. (d) Upper eyebrow with 2 N of applied pressure.