Table 1 3D ablation radii (minimum and maximum in mm) for each A-V distance and all flow rates (median (min / max); *p ≤ 0.008; (*)p ≥ 0.008 - ≤ 0.05.

From: Optimizing microwave ablation planning with the ablation success ratio

 

0 ml/min

1 ml/min

2 ml/min

5 ml/min

10 ml/min

100 ml/min

500 ml/min

A-V distance: 2.5 mm

 rmin

9.3 (8.7/10.4)

9.3 (7.5/9.9)

8.2(*) (5.6/9.5)

7.6(*) (7.1/8.9)

6.7* (4.8/8.3)

5.8 * (4.3/6.9)

5.3 * (4.8/6.0)

 rmax

15.4 (13.5/16.9)

16.0 (14.5/18.1)

14.3 (12.8/14.5)

14.2 (13.4/16.2)

14.9 (14.2/20.2)

14.5 (13.5/15.3)

14.4 (13.5/16.2)

A-V distance: 5.0 mm

 rmin

9.2 (8.4/10.7)

8.3 (6.6/9.8)

5.9(*)(5.0/9.6)

5.4*(3.5/5.7)

4.9*(4.6/5.9)

4.2* (3.9/4.9)

4.8* (3.7/6.9)

 rmax

14.7 (12.8/18.4)

15.3 (12.9–16.1)

16.4 (14.9/20.2)

14.3 (12.4/15.6)

13.4 (11.4/16.9)

14.5 (12.1/16.8)

14.6 (12.6/17.5)

A-V distance: 10.0 mm

 rmin

8.8 (8.7/10.5)

8.5 (6.9/9.6)

7.0* (6.5/7.5)

7.0* (6.6/8.4)

7.0* (5.7/7.9)

6.6* (6.2/7.5)

6.4* (5.5/7.2)

 rmax

15.8 (12.8/18.3)

15.6 (13.7/16.8)

16.2 (14.8/16.8)

16.4 (13.1/18.3)

16.3 (12.0/20.1)

16.9 (14.4/18.8)

14.6 (12.2/18.1)