From: Accre 8 emerging point of care CLIA system for vitamin B12 assessment compared with three established assays

Fig. 2The alternative text for this image may have been generated using AI.

Correlation Analysis of Vit B12 levels between different assays. Figure 2 illustrates the correlation analysis of Vit B12 levels between different assays, assessing the strength of the relationship between LC-MS/MS and three CLIA-based methods using Spearman’s rank correlation coefficients (r). Plot A presents the correlation between Accre 8 and LC-MS/MS (r = 0.94, p < 0.001), while Plot B shows the correlation between Accre 8 and Roche (r = 0.94, p < 0.001). Plot C displays the correlation between Accre 8 and Abbott (r = 0.94, p < 0.001). Plot D illustrates the correlation between Abbott and LC-MS/MS (r = 0.94, p < 0.001), whereas Plot E presents the strongest correlation between Roche and LC-MS/MS (r = 0.98, p < 0.001). Each plot includes a diagonal reference line (y = x) for visual assessment of agreement.