Table 7 Comparison of anti-interference capacity of our kit and commercial kits.
Baicalein(µg/mL) | 0 | 10 | 20 | 30 | 40 | 60 | 80 | 100 |
|---|---|---|---|---|---|---|---|---|
Our kit | 365.6 | 356.1 | 358.9 | 365.5 | 370 | 359 | 344 | 325.6 |
Bias (%) | – | – 2.59 | – 1.83 | – 0.027 | 1.20 | – 1.8 | – 5.9 | – 10.9 |
Commercial kit A | 365.3 | 334.2 | 304 | 278.9 | 258 | 221.4 | 184.8 | 155.4 |
Bias (%) | – | – 8.51 | – 16.78 | – 23.65 | – 29.37 | – 39.39 | – 49.41 | – 57.4 |
P value a | 0.7565 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 |
Commercial kit B | 365.1 | 333.1 | 306 | 280.1 | 259.2 | 223.2 | 186.3 | 157.2 |
Bias (%) | – | – 8.76 | – 16.18 | – 23.28 | – 29.01 | – 38.86 | – 48.97 | – 56.94 |
P value b | 0.7022 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 |