Abstract
Repeated exposure to iodinated contrast media (ICM) may induce thyroid dysfunction but supporting radiological and histological evidence remains limited. This study aimed to evaluate whether repeated ICM administration induces iodine accumulation and related functional and histological changes in the thyroid. This prospective 8-week study involved 30 male New Zealand white rabbits (0.8–1.2 kg, < 1 year) assigned to four groups. The control group (n = 8) received 0.9% sodium chloride solution, while experimental groups received intravenous iohexol three times weekly at 1.6 mL/kg (n = 7), 6.0 mL/kg (n = 7), or 12.0 mL/kg (n = 8). Non-contrast thyroid dual-energy computed tomography (DECT) was performed weekly to quantify iodine accumulation, expressed as the thyroid enhancement ratio (TER) relative to paraspinal muscle. Serum free thyroxine (fT4), free triiodothyronine (fT3), thyroid-stimulating hormone, and creatinine were measured at baseline, week 4, and week 8. Histological analysis was performed at week 8. At week 8, TER was significantly higher in the 6.0 mL/kg (2.59 ± 0.89) and 12.0 mL/kg (2.79 ± 0.77) groups than in controls (1.15 ± 0.42) and the 1.6 mL/kg group (1.32 ± 0.46) (all P < 0.05). While transient fT4 and fT3 elevations occurred and nodular hyperplasia, follicular atrophy, and inflammation were more frequent in high-dose groups, neither showed statistically significant intergroup differences (both P > 0.05). In conclusion, repeated intravenous administration of high-dose ICM in rabbits was associated with increased iodine accumulation in the thyroid gland, as quantified by DECT. However, a transient rise in serum thyroid hormone levels, as well as more frequent histopathological changes, did not reach statistical significance.
Data availability
The datasets generated or analyzed during the study are available from the corresponding author on reasonable request.
Abbreviations
- CT:
-
Computed tomography
- DECT:
-
Dual-energy computed tomography
- H&E:
-
Hematoxylin and eosin
- HU:
-
Hounsfield unit
- ICC:
-
Intraclass correlation coefficient
- ICM:
-
Iodinated contrast media
- ROI:
-
Region of interest
- VMI:
-
Virtual monoenergetic imaging
- TER:
-
Thyroid enhancement ratio
- TSH:
-
Thyroid-stimulating hormone
References
Beckett, K. R., Moriarity, A. K. & Langer, J. M. Safe use of contrast media: what the radiologist needs to know. Radiographics 35, 1738–1750. https://doi.org/10.1148/rg.2015150033 (2015).
Andreucci, M., Solomon, R. & Tasanarong, A. Side effects of radiographic contrast media: pathogenesis, risk factors, and prevention. Biomed. Res. Int. 2014 (741018). https://doi.org/10.1155/2014/741018 (2014).
Lightfoot, C. B. et al. Survey of radiologists’ knowledge regarding the management of severe contrast material-induced allergic reactions. Radiology 251, 691–696. https://doi.org/10.1148/radiol.2513081651 (2009).
US Food and Drug Administration. FDA recommends thyroid monitoring in babies and young children who receive injections of iodine-containing contrast media. https://www.fda.gov/drugs/drug-safety-and-availability/fda-recommends-thyroid-monitoring-babies-and-young-children-who-receive-injections-iodine-containing (2022).
Dillman, J. R. et al. ACR statement on use of iodinated contrast material for medical imaging in young children and need for thyroid monitoring. J. Am. Coll. Radiol. 19, 849–853. https://doi.org/10.1016/j.jacr.2022.05.014 (2022).
Jick, S. S. et al. Iodinated contrast agents and risk of hypothyroidism in young children in the united States. Invest. Radiol. 54, 296–301. https://doi.org/10.1097/RLI.0000000000000541 (2019).
Gilligan, L. A. et al. Primary thyroid dysfunction after single intravenous iodinated contrast exposure in young children: a propensity score matched analysis. Pediatr. Radiol. 51, 640–648. https://doi.org/10.1007/s00247-020-04881-0 (2021).
Belloni, E. et al. Effect of iodinated contrast medium on thyroid function: a study in children undergoing cardiac computed tomography. Pediatr. Radiol. 48, 1417–1422. https://doi.org/10.1007/s00247-018-4163-3 (2018).
Kubicki, R. et al. Frequency of thyroid dysfunction in pediatric patients with congenital heart disease exposed to iodinated contrast media - a long-term observational study. J. Pediatr. Endocrinol. Metab. 33, 1409–1415. https://doi.org/10.1515/jpem-2020-0032 (2020).
Thaker, V. V. et al. Hypothyroidism in infants with congenital heart disease exposed to excess iodine. J. Endocr. Soc. 1, 1067–1078. https://doi.org/10.1210/js.2017-00174 (2017).
Dechant, M. J. et al. Thyroidal response following iodine excess for cardiac catheterisation and intervention in early infancy. Int. J. Cardiol. 223, 1014–1018. https://doi.org/10.1016/j.ijcard.2016.08.292 (2016).
McCollough, C. H., Leng, S., Yu, L. & Fletcher, J. G. Dual- and Multi-Energy CT: Principles, technical Approaches, and clinical applications. Radiology 276, 637–653. https://doi.org/10.1148/radiol.2015142631 (2015).
Kim, H., Goo, J. M., Kang, C. K., Chae, K. J. & Park, C. M. Comparison of iodine density measurement among Dual-Energy computed tomography scanners from 3 vendors. Invest. Radiol. 53, 321–327. https://doi.org/10.1097/RLI.0000000000000446 (2018).
Pelgrim, G. J. et al. Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT. Eur. Radiol. 27, 3904–3912. https://doi.org/10.1007/s00330-017-4752-9 (2017).
Nievelstein, R. A., van Dam, I. M. & van der Molen, A. J. Multidetector CT in children: current concepts and dose reduction strategies. Pediatr. Radiol. 40, 1324–1344. https://doi.org/10.1007/s00247-010-1714-7 (2010).
Lee, Y. Dual-energy computed tomography-based volumetric thyroid iodine quantification: correlation with thyroid hormonal status, pathologic diagnosis, and Phantom validation. Diagn. Interv Radiol. 31, 226–233. https://doi.org/10.4274/dir.2025.243132 (2025).
Lee, Y. J. et al. Effects of iodine status on thyroid volume and goiter in children living in an iodine-replete area. Eur. Thyroid J. 12 https://doi.org/10.1530/ETJ-23-0219 (2023).
Koo, T. K. & Li, M. Y. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J. Chiropr. Med. 15, 155–163. https://doi.org/10.1016/j.jcm.2016.02.012 (2016).
Lerner, R. K., Gruber, N. & Pollak, U. Congenital heart disease and thyroid dysfunction: Combination, Association, and implication. World J. Pediatr. Congenit Heart Surg. 10, 604–615. https://doi.org/10.1177/2150135119857704 (2019).
Elbaz, B. A., Elmarsafawy, H. M. & Laimon, W. N. The effect of cardiac catheterization on thyroid functions in infants with congenital heart diseases: a prospective observational study. Eur. J. Pediatr. 184, 137. https://doi.org/10.1007/s00431-024-05934-4 (2025).
Agostini, A. et al. New frontiers in oncological imaging with computed tomography: from morphology to function. Semin Ultrasound CT MR. 44, 214–227. https://doi.org/10.1053/j.sult.2023.03.009 (2023).
Rapp, J. B., Biko, D. M. & Siegel, M. J. Dual-Energy CT for pediatric thoracic imaging: A review. AJR Am. J. Roentgenol. 221, 526–538. https://doi.org/10.2214/AJR.23.29244 (2023).
Iwano, S., Ito, R., Umakoshi, H., Ito, S. & Naganawa, S. Evaluation of lung cancer by enhanced dual-energy CT: association between three-dimensional iodine concentration and tumour differentiation. Br. J. Radiol. 88, 20150224. https://doi.org/10.1259/bjr.20150224 (2015).
Jacobsen, M. C. et al. Intermanufacturer comparison of Dual-Energy CT iodine quantification and monochromatic attenuation: A Phantom study. Radiology 287, 224–234. https://doi.org/10.1148/radiol.2017170896 (2018).
Sellerer, T. et al. Dual-energy CT: a Phantom comparison of different platforms for abdominal imaging. Eur. Radiol. 28, 2745–2755. https://doi.org/10.1007/s00330-017-5238-5 (2018).
Dabli, D. et al. Optimization of image quality and accuracy of low iodine concentration quantification as function of kVp pairs for abdominal imaging using dual-source CT: A Phantom study. Phys. Med. 88, 285–292. https://doi.org/10.1016/j.ejmp.2021.07.008 (2021).
Dabli, D. et al. Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A Phantom study. Diagn. Interv Imaging. 103, 31–40. https://doi.org/10.1016/j.diii.2021.08.004 (2022).
Euler, A., Solomon, J., Mazurowski, M. A., Samei, E. & Nelson, R. C. How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a Phantom study. Eur. Radiol. 29, 2069–2078. https://doi.org/10.1007/s00330-018-5736-0 (2019).
Clemens, P. C. & Neumann, R. S. The Wolff-Chaikoff effect: hypothyroidism due to iodine application. Arch. Dermatol. 125, 705 (1989).
Markou, K., Georgopoulos, N., Kyriazopoulou, V. & Vagenakis, A. G. Iodine-Induced hypothyroidism. Thyroid 11, 501–510. https://doi.org/10.1089/105072501300176462 (2001).
Binh, D. D., Nakajima, T., Otake, H., Higuchi, T. & Tsushima, Y. Iodine concentration calculated by dual-energy computed tomography (DECT) as a functional parameter to evaluate thyroid metabolism in patients with hyperthyroidism. BMC Med. Imaging. 17, 43. https://doi.org/10.1186/s12880-017-0216-6 (2017).
Szkudlinski, M. W., Fremont, V., Ronin, C. & Weintraub, B. D. Thyroid-stimulating hormone and thyroid-stimulating hormone receptor structure-function relationships. Physiol. Rev. 82, 473–502. https://doi.org/10.1152/physrev.00031.2001 (2002).
Beindorff, N. et al. Normal values of thyroid uptake of 99mTechnetium pertechnetate SPECT in mice with respect to Age, Sex, and circadian rhythm. Nuklearmedizin 57, 181–189. https://doi.org/10.3413/Nukmed-0978-18-05 (2018).
Funding
This work was supported by Grant No. 0620215500 from Encormac (Seoul, Republic of Korea) and in part by the National Research Foundation of Korea (NRF) funded by the Ministry of Science and ICT (MSIT) (Grant No. RS-2023-00250062).
Author information
Authors and Affiliations
Contributions
S.L. and C.R. designed and led the study. C.R., S.L., and J.K. performed data curation and formal analysis. S.L. acquired funding and provided resources. C.R. drafted the original manuscript, and C.R., S.L., Y.H.C., J.Y.H., and J.E.C. reviewed and edited the final version.
Corresponding author
Ethics declarations
Competing interests
Seunghyun Lee is a Scientific Reports editorial board member but has no role in the review process or decision to publish this article. All remaining authors have declared no conflicts of interest.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Rhee, C., Lee, S., Koh, J. et al. Assessment of thyroid iodine accumulation following repeated iodinated contrast media administration using dual-energy computed tomography in a rabbit model. Sci Rep (2026). https://doi.org/10.1038/s41598-026-37701-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-026-37701-w