Table 2 Regression statistics for Al I (396.4 nm) and Fe I (373.5 nm) calibration curves obtained with and without MW enhancements using multifiber LIBS.

From: Improvement of SNR in laser-induced breakdown spectroscopy using microwave and multifiber synergy

 

Condition

Slope (a)

Intercept (b)

sy (a.u.)

LOD (wt.%)

LOQ (wt.%)

Al I

With MW

1.58 × 10⁶

− 1.47 × 10⁸

310,500.74

0.59

1.966

Without MW

1.94 × 105

− 1.83 × 10⁷

65,772.13

1.016

3.386

Fe I

With MW

8.86 × 105

7.79 × 103

15,084.83

0.323

0.510

Without MW

3.14 × 104

3.07 × 104

1412.12

0.378

0.678