Table 2 Summary of verification.

From: Verification, analytical validation, and clinical validation (V3): the foundation of determining fit-for-purpose for Biometric Monitoring Technologies (BioMeTs)

Who?

Engineers, data & computer scientists

What?

Generation and preliminary processing of sample-level data

When?

Prior to testing the technology in human subjects

Where?

At the bench

Why?

To evaluate the performance of a sensor technology (1) against pre-specified criteria and (2) to demonstrate that the sample-level data generated is correct within the limits of the pre-specified conditions.