Table 1 Comparison of apoferritin data collection and reconstruction parameters between this paper and that in Naydenova et al.1.

From: MeasureIce: accessible on-the-fly measurement of ice thickness in cryo-electron microscopy

 

This paper

Naydenova et al.16

Grids

Gold Ultrafoil 1.2/1.3

Functionalized graphene-coated gold Ultrafoil 1.2/1.3

Microscope

Titan Krios G4

Titan Krios

Accelerating voltage (kV)

300

300

Magnification

120,000

120,000

Pixel size (Å)

0.637

0.6495

Exposure time (s)

1.52

30.0

Dose (e/Å2)

16

37

Number of Particles used in final refinement

68,025

41,202

Gold standard FSC map resolution (Å)

1.88

2.14

B-factor (Å2)

−54

−54