Table 3 Key parameters of the MINFLUX 2D iteration sequence

From: MINFLUX reveals nonnormal distribution of linkage errors in immunofluorescence labeling

2D imaging

Center frequency ratio limits (CFR)

Dwell time (ms)

L-size (nm)

Minimal photon count

Laser power factor

TCP

Pattern repeat

Background threshold (kHz)

Pre-scan

Off

\(\ge\)1

288

160

1

Hexagon

1

15

1st iteration

0.5

\(\ge\)1

288

150

1

Hexagon

5

10

2nd iteration

Off

\(\ge\)1

151.2

100

2

Hexagon

5

10

3rd iteration

0.8

\(\ge\)1

75.6

100

4

Hexagon

5

10

4th iteration

Off

\(\ge\)1

39.6

150

6

Hexagon

5

10