Table 2 Performance characteristics for the PLS and iCONIC models (* indicates SECV is the correct error metric for PLS and is used here)
From: Spectroscopic approach for dynamic bioanalyte tracking with minimal concentration information
| Â | Â | Â | Â | Regression analysis of the prediction data | ||
|---|---|---|---|---|---|---|
Model | SDP (mg/dL) | SEP (mg/dL) | Model size | β0 (mg/dL) | β1 | R2 |
PLS | 35.57 | 15.92* | 208 | 13.57 | 0.9057 | 0.9001 |
iCONIC | 35.57 | 9.76 | 208 | 6.33 | 0.9893 | 0.9718 |