Abstract
The final rinse water pollution of endoscopes in medical institutions is a severe growing problem, posing a latent risk to nosocomial infection. This study aimed to explore the efficacy of combination of point-of-use (POU) filters and peracetic acid (PAA) disinfection in reducing total viable counts (TVCs) in final rinse water of endoscopes. Eight final rinse water faucets of endoscope centers were randomly selected and none of the faucets had POU filters prior to the study. Four faucets were assigned to experimental group and had POU filters installed, the other four faucets were set as control group and did not have POU filters installed. During monitoring, both the experimental and control groups were used 0.3% PAA to disinfect the purified water system twice. TVCs in final rinse water was monitored weekly for 13 weeks. 231 final rinse water samples were collected, of which 111 and 120 samples were in the experimental and control groups, respectively. The mean TVCs and positive rate of samples for experimental group were both significantly lower than that of control group [10 vs. 50,450 colony-forming units (CFU)/100 mL, and 27.7% vs. 98.3%, P < 0.001]. The TVCs in control group was from 2.0 × 103 to 2.5 × 105 CFU/100 mL, and the positive rates of samples were almost 100%. However, the TVCs and positive rates of samples in experiment group were always at a low-level, with a maximum TVCs of 47 CFU/100 mL, corresponding to a maximum positive rate of 50%, by the 13th week of monitoring. Our findings demonstrated that PAA disinfection alone exhibited limited efficacy in controlling TVCs in final rinse water of endoscopes, whereas its combined application with POU filters significantly enhances the control effect of TVCs.
Similar content being viewed by others
Introduction
Hospital centralized water supply systems have served as reservoirs for waterborne pathogens, and colonization of waterborne pathogens in water supplies have been correlated with the incidence of healthcare-associated infections (HAIs)1,2,3,4. As the final rinsing process after endoscope disinfection, the contamination of endoscope final rinse water can easily lead to endoscopic contamination and iatrogenic infections5. Previous study revealed that contamination of the endoscopic final rinse water increased probability of endoscopic reprocessing failure6. Endoscope reprocessing failures have been reported by the ECRI institute as one of the top 10 health technology hazards7. Ji et al. investigated 59 endoscope units in China and found that 56.52% flushing channel samples of gastrointestinal endoscopes were positive for culture, but use of purified water could reduce the risk of endoscope reprocessing failure8. Multinational guidelines recommend rinsing endoscopes and accessories with purified or sterile water after high-level disinfection9,10. Monitoring and reducing the total viable counts (TVCs) for final rinse water of endoscopy centers were essential to prevent infection caused by gastrointestinal endoscopes11.
The most common method for controlling the TVCs in water is to regularly disinfect the water systems with chemical disinfectants12. At present, the commonly used disinfectant include chlorine, chlorine dioxide, chlorine ammonia, and ozone disinfection, etc., but these methods have the problem of residual toxic disinfection by-products13,14. Peracetic acid (PAA) as a widely used alternative oxidant and disinfectant in water treatment, can rapidly and effectively eliminate trace contaminants and produce fewer harmful by-products15. However, PAA disinfection must be repeated at certain intervals, and TVCs of the final rinse water cannot be effectively controlled. Point-of-use (POU) water filter has obvious benefits to lower the number of waterborne pathogens16, and was a simple, successful, and highly cost-effective strategy to control the outbreak of nosocomial infection17,18. After an outbreak of invasive fusariosis in a children’s cancer hospital in Brazil, many control measures were instituted, but the outbreak was only controlled one year after the first case, when all faucets and showers were equipped with POU filters filtering 0.2 μm19. A large outbreak of legionella infection occurred in an American hospital despite using copper-silver ionization system, while the legionella epidemic was effectively controlled by equipping with POU water filters20.
In the present study, we aimed to explore the influence of combining the PAA chemical disinfection for the centralized water supply system with installing POU filters for the final rinse water of endoscopes on controlling the TVCs in the final rinse water.
Materials and methods
Study design
The study was performed at a endoscopic center of a tertiary general hospital in Wuhan, China, from March to June 2022. Eight final rinse water faucets of endoscope centers were randomly selected and none of the faucets had POU filters prior to the study. Four faucets were assigned to experimental group and had POU filters installed, the other four faucets were assigned to control group and did not have POU filters installed. During monitoring, the pure water storage tanks, water transmission pipelines [constructed from unplasticized polyvinyl chloride (PVC-U)], and each terminal point of both the experimental and control groups were disinfected twice by using 0.3% PAA, with each immersion cycle lasting 40 min (Fig. 1). TVCs in final rinse water was monitored weekly, starting from the day of PAA disinfection and POU filter installation, lasted for 13 weeks (Fig. 1).
The POU filter (Pall-Aqua-Safe Water Filter, AQ14F1S) is prepared with polyether sulfone membrane of 0.2 μm, whose filter element is able to intercept microorganisms and particles in the water and purify water, based on the principle of physical interception. Moreover, silver ion bacteriostatic is added to the filter shell to prevent bacteria from growing on the filter. Nurses and cleaning personnel at the endoscopic center were instructed not to touch or clean the filters, which were not changed during the monitoring period.
Sample collection
After disinfecting with 75% ethanol and 60 s water flush, all samples (100 mL per sample) were collected by using pre-sterilized glass bottles, and were stored at a low-temperature refrigerator immediately until testing within two hours of collection. We totally collected 231 samples, of which 111 collected from faucets with filters installed were assigned to the experimental group, and the other 120 collected from faucets without filters installed were set as the control group. Meanwhile, we simultaneously collected 24 source water samples from water tank of centralized water supply system and 20 endoscopic final rinse water samples from faucets without POU filters on the second day after PAA disinfection.
Microbiological analysis
Water samples were treated by the membrane filtration method. 1.0 mL of original water, diluted 10-fold and 100-fold water samples were poured culture, respectively, and the remaining 99 mL of samples were filtered (pore size of 0.45 μm, Millipore, USA). After filtration, the membranes were attached to R2A agar mediums and incubated at 17–23 °C for seven days for colony counting21. After culturing, if the filter membrane could not be counted, the result of the poured plate was adopted; otherwise, the sum of the number of bacterial colonies on the plate and the filter membrane was counted. The TVCs of 0 colony-forming units (CFU)/100 mL was considered qualified.
Statistical analysis
The distribution of characteristic variables was expressed as the mean (standard deviation, SD), number (percentage), or median (interquartile range, IQR). The differences in continuous variables between experiment and control groups were tested by Mann–Whitney U test or paired-chi square (Fisher exact). The threshold for significance was P < 0.05. In this study, all statistical analyses were performed with Statistical Analysis System 9.4 (SAS Institute, Carry, North Carolina, https://support.sas.com/software/94/) and GraphPad Prism 8 software (San Diego CA, https://www.graphpad.com/scientific-software/prism).
Results
The flowchart of the study was shown in Fig. 2. The TVCs and positive rate of samples for experimental group were both significantly lower than that of the control group (10 vs. 50,450 CFU/100 mL and 27.7% vs. 98.3%, P < 0.001, Table 1).
The TVCs for 24 water source samples was significantly lower than 20 endoscopic final rinsing water samples (6.2 vs. 128.2 CFU/100 mL, P = 0.001). During the surveillance, TVCs among experimental group and control group both gradually increased with the time after disinfection, but the former always lower than the later (Fig. 3).
As shown in Fig. 4, the TVCs for the control group ranged from 2.0 × 103 to 2.5 × 105 CFU/100 mL, and was nearly unaltered or slightly increased within 2 weeks after PAA disinfection. However, the TVCs gradually increased as the disinfection interval extended, but decreased after the second PAA disinfection, while when the 13th week of the monitoring period had the highest TVCs, which was 2.5 × 105 CFU/100 mL, followed by the 12th week, which was 1.0 × 105 CFU/100 mL. The TVCs for the experiment group ranged from 0 to 47 CFU/100 mL, and was unchanged or slightly increased during the monitoring time; it is worth noting that a reduction to 0 CFU/100 mL in TVCs occurred after the second disinfection.
As shown in Fig. 5, the positive rates for the control group were almost 100%, but the positive rates for the experiment group were always at a low to medium level, with the 13th week of the monitoring period having the highest positive rate, which was 50%.
The predominant bacteria detected in water samples for control group were Pseudomonas fluorescens, Sphingomonas paucimobilis, Ralstonia picketti, Stenotrophomonas maltophilia, Brevundimonas vesicularis, Burkholderia cepacia and Methylobacteriu. While for the experiment group, only Ralstonia pickettii and Burkholderia cepacia were detected.
Discussion
To our knowledge, this is the first study to investigate the effect of decreasing the TVCs in the final rinse water of endoscopes by combination of POU filter and PAA disinfection. Severe microbial pollution of medical water has become a difficult problem for medical institutions22,23. Contamination of water system by potentially infectious microorganisms, such as bacteria, viruses and protozoa, is a source of nosocomial infections24,25. Our research found that TVCs in the final rinse water of endoscopes easily exceed the standard, but it is practicable to combine POU filters installation at the final of branches with PAA disinfection for pipeline systems to control the bacterial count in the final rinse water.
Removing microbial contaminants from water through chemical disinfection is a way to guarantee the safe management of medical water26,27. PAA has powerful antibacterial activity and PAA chemical disinfection can effectively destroy multiple pathogenic microorganisms in water, which is regarded as the best alternative to chlorine disinfection and is widely used in disinfection of water systems in medical institutions28. However, in this study, we found that after PAA disinfection, bacterial contamination still occurred during transportation in waterway system, resulting in significantly higher TVCs in the rinse water from branch terminal than that in the source water. In our study, the TVCs in the final rinse water of endoscope centers began to increase after one week of disinfection of centralized water supply system and transmission pipelines. Therefore, due to the instability and limited duration of the effect, PAA disinfection must be repeated at a certain time interval29. Previous study indicated that high-frequency disinfection would corrode pipeline system components, cause biofilm attachment, and produce a large number of disinfection by-products; long-term use will also lead to selective resistance of bacteria to disinfectants26,30,31,32. In this study, PVC-U was selected as the pipeline material due to its documented resistance to PAA corrosion. Throughout the experimental period, no corrosion of the PVC-U material caused by PAA exposure was observed. The biofilm shedding in water system releases attached bacteria into water, resulting in altered number of bacterial colonies33, which may explain the large change in the number of colonies after the second PAA chemical disinfection among control group in this study. Consequently, the effectiveness of only using PAA chemical disinfection to control the TVCs in final rinse water is questionable.
As an efficient and rapid intervention method, POU filters can effectively reduce the bacterial content of final water34, and reduce the risk of waterborne infection35,36. A study at a cancer center in Northwestern Pennsylvania revealed that no legionella was detected in POU filtered samples within 12 weeks, and the total bacteria was decreased by an average of 1.86 logarithm compared with controls37. Parkinson et al. found that POU filters installed on both faucets and showers eliminated legionella and reduced heterotrophic plate count concentrations for 12 weeks38, which was consistent with our results that the bacterial count in the final rinse water of the experiment group was significantly lower than that of the control group. Bielefeldt et al. found that the initial disinfection efficiencies of POU filters ranged from 3 to 4.5 logarithm experimentally, but the effectiveness decreased to 0.2–2.5 logarithm after loading multiple batches of highly contaminated water, indicating that POU filters were effective, but showed loss of effectiveness with time39. In addition, water chemistry (pH, OM, inorganic ions), adsorbent characteristics (pore distribution, particle size, functional group) can affect the adsorption efficiency of POU filters, and lead to variable result in real world application40.
The combination of POU filters and PAA disinfection obtain better disinfection effects and economic applicability. Previous studies indicated that the number of bacteria in the downstream of POU filters and risk of reverse pollution both gradually increased after a period of use; downstream of filter membranes and filter outlets were susceptible to the pollution of liquid splashing, drain aerosol backwash, and other ways, resulting in the increase of the bacterial41. It is debatable whether it is economical to invest medical resources in preventing healthcare-associated infections related to opportunistic aquatic pathogens, especially since their short maximum lifetime and relatively high prices have limited their use36,37. In this study, we found that installation of POU filter combined with PAA chemical disinfection could significantly reduce the number and variety of bacteria for lasting about 12 weeks, this duration exceeded the 30-day timeframe typically advised in the product manual. Therefore, combination of regular chemical disinfection of central water supply systems with POU filters in essential departments such as endoscopy center is an economical and feasible method control the bacteria in downstream of POU within the required range and prevent nosocomial infection36,38,42.
Most of the bacteria detected in the study belonged to the Proteobacteria, which are widely distributed in hospital environments and water, and some of them are opportunistic pathogens. Burkholderia cepacia can cause bloodstream infections and has high infectivity43,44. Stenotrophomonas maltophilia can lead to pneumonia, bloodstream infections, or catheter-related infections in immunocompromised individuals45. Pseudomonas fluorescens is occasionally reported to cause pneumonia46 and bloodstream infections47. Ralstonia pickettii and Brevundimonas vesicularis have low virulence and rarely cause nosocomial infections. The literature on nosocomial infections reported by Methylobacterium and Sphingomonas paucimobilis is limited. Stenotrophomonas maltophilia and Burkholderia cepacia have prominent resistance, among which Stenotrophomonas maltophilia has significant resistance to carbapenems, quinolones, tetracyclines, β-lactams and other antibiotics; Burkholderia cepacia is a multidrug-resistant bacterium that exhibits extensive resistance to β- lactam and aminoglycoside antibiotics. In our study, bacterial identification revealed no significant changes in antibiotic resistance profiles before and after the combination of POU filters and PAA disinfection. We acknowledge that prolonged PAA exposure may theoretically induce biofilm adaptation. But our integrated POU filtration system’s regular replacement protocol effectively disrupts potential resistance development cycles. Nevertheless, further long-term studies are necessary to conduct more detailed analyses on strain typing and antibiotic resistance monitoring.
This study is the first to introduce the concept of controlling the TVCs in the final rinse water of endoscopes by combining POU filters with PAA chemical disinfection of pipeline systems, and provides an important reference for controlling bacterial reproduction and biofilm formation in water system. In addition, this study used the R2A culture method for the detection of bacteria in water, which has higher sensitivity and stronger reliability of results21,48,49. In some countries the microbial level is expected to be < 10 CFU/100 mL50 or 0 CFU/100 mL51, and in our study we have adopted strict standards that the TVCs = 0 CFU/100 mL was considered as qualified. However, several limitations of our study should be acknowledged. First, the quality of purified water is affected by many factors, such as source water, storage temperature and environment, and season26,52, hence the interval of disinfection in pipe network should be further studied and discussed according to regional and seasonal differences. Second, due to the hospital’s routine use of PAA chemical disinfection, we cannot test without PAA disinfection to evaluate the impact of filtration. However, the TVCs in the control group measured one week before PAA disinfection was significantly higher than the TVCs measured one week after PAA disinfection (62,625 vs. 2354 CFU/100 mL, P < 0.001), which could be used as a before-and-after comparison. Third, as a widely used chemical disinfectant, the side effects of PAA cannot be ignored, which although were not analyzed in the current research but need to be accessed in the further studies. Forth, for the branches of central pipeline water supply system, only endoscope center was selected. It is necessary to further study the long-term effects of installing POU filters on the number of pathogenic microorganisms and related nosocomial infections in other departments or hospitals.
Conclusion
The central water supply system is prone to excessive microorganisms at the branches. Combination of POU filters with PAA chemical disinfection of pure water pipelines is able to effectively reduce the TVCs detected in final rinse water of endoscopes. Our study provides an effective reference for medical institutions with central water supply systems to control the bacterial content in final rinse water.
Data availability
The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.
Abbreviations
- CFU:
-
Colony-forming units
- IQR:
-
Interquartile range
- PAA:
-
Peracetic acid
- POU:
-
Point-of-use
- SD:
-
Standard deviation
- TVCs:
-
Total viable counts
- PVC-U:
-
Unplasticized polyvinyl chloride
References
Hayward, C. et al. Hospital water as the source of healthcare-associated infection and antimicrobial-resistant organisms. Curr. Opin. Infect. Dis. 35(4), 339–345 (2022).
Walker, J. & Moore, G. Pseudomonas aeruginosa in hospital water systems: Biofilms, guidelines, and practicalities. J. Hosp. Infect. 89(4), 324–327 (2015).
Zhou, W. et al. Microbial diversity in full-scale water supply systems through sequencing technology: A review. RSC Adv. 11(41), 25484–25496 (2021).
Thacharodi, A. et al. Carbapenem-resistant Acinetobacter baumannii raises global alarm for new antibiotic regimens. Science 27(12), 111367 (2024).
Rutala, W. A. & Weber, D. J. Disinfection, sterilization, and antisepsis: An overview. Am. J. Infect. Control 47, A3–A9 (2019).
Ji, X. Y. et al. The importance of sampling technique and rinse water for assessing flexible gastrointestinal endoscope reprocessing: A 3-year study covering 59 centers. Am. J. Infect. Control 48(1), 19–25 (2020).
ECRI Institute. Top 10 Health Technology Hazards for 2019. https://assets.ecri.org/PDF/White-Papers-and-Reports/2019-Top-10-Hazards.pdf (Accessed 1 August 2019).
Ji, X. Y. et al. Microbiologic assessment of flexible gastrointestinal endoscope reprocessing using a pump-assisted sampling technique: An investigation involving all endoscopy units in Tianjin, China. Am. J. Infect. Control 46(7), e43–e48 (2018).
Speth, J. Guidelines in practice: Processing flexible endoscopes. AORN J. 118(3), 169–178 (2023).
Beilenhoff, U. et al. Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: Position Statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA)—Update 2018. Endoscopy 50(12), 1205–1234 (2018).
Blázquez-Garrido, R. M. et al. Microbiological monitoring of medical devices after cleaning, disinfection and sterilisation. Enferm. Infecc. Microbiol. Clin. 36(10), 657–661 (2018).
O’Donnell, M. J. et al. Management of dental unit waterline biofilms in the 21st century. Future Microbiol. 6(10), 1209–1226 (2011).
Richardson, S. D. & Ternes, T. A. Water analysis: Emerging contaminants and current issues. Anal. Chem. 94(1), 382–416 (2022).
Tomlinson, A., Drikas, M. & Brookes, J. D. The role of phytoplankton as pre-cursors for disinfection by-product formation upon chlorination. Water Res. 102, 229–240 (2016).
Ao, X. W. et al. Peracetic acid-based advanced oxidation processes for decontamination and disinfection of water: A review. Water Res. 188, 116479 (2021).
Florentin, A. et al. Water and surface microbiologic quality of point-of-use water filters: A comparative study. Am. J. Infect. Control 44(9), 1061–1062 (2016).
Oh, Y. et al. Electrically heatable carbon nanotube point-of-use filters for effective separation and in-situ inactivation of Legionella pneumophila. Chem. Eng. J. 366, 21–26 (2019).
Totaro, M. et al. Cost analysis for electron time-flow taps and point of use filters: A comparison of two methods for Legionnaires’ disease prevention in hospital water networks. J. Hosp. Infect. 103(2), 231–232 (2019).
Litvinov, N. et al. An outbreak of invasive fusariosis in a children’s cancer hospital. Clin. Microbiol. Infect. 21(3), 268.e1–7 (2015).
Demirjian, A. et al. The importance of clinical surveillance in detecting legionnaires’ disease outbreaks: A large outbreak in a hospital with a Legionella disinfection system-Pennsylvania, 2011–2012. Clin. Infect. Dis. 60(11), 1596–1602 (2015).
Cao, X. et al. Comparing the effects of two culture methods to determine the total heterotrophic bacterial colony count in hospital purified water. J. Epidemiol. Glob. Health 14(1), 184–192 (2024).
Richardson, S. D. & Kimura, S. Y. Water analysis: Emerging contaminants and current issues. Anal. Chem. 92(1), 473–505 (2020).
Hutchins, C. F. et al. Contamination of hospital tap water: The survival and persistence of Pseudomonas aeruginosa on conventional and “antimicrobial” outlet fittings. J. Hosp. Infect. 97(2), 156–161 (2017).
Inkster, T. et al. Investigation and control of an outbreak due to a contaminated hospital water system, identified following a rare case of Cupriavidus pauculus bacteraemia. J. Hosp. Infect. 111, 53–64 (2021).
Lytle, D. A. et al. A comprehensive evaluation of monochloramine disinfection on water quality, Legionella and other important microorganisms in a hospital. Water Res. 189, 116656 (2021).
Lindmark, M. et al. Passive in-line chlorination for drinking water disinfection: A critical review. Environ. Sci. Technol. 56(13), 9164–9181 (2022).
Wang, J. et al. Disinfection technology of hospital wastes and wastewater: Suggestions for disinfection strategy during coronavirus Disease 2019 (COVID-19) pandemic in China. Environ. Pollut. 262, 114665 (2020).
Zhang, C. et al. Inhibition of regrowth of planktonic and biofilm bacteria after peracetic acid disinfection. Water Res. 149, 640–649 (2019).
Nocker, A. et al. Thermal and chemical disinfection of water and biofilms: Only a temporary effect in regard to the autochthonous bacteria. J. Water Health 19(5), 808–822 (2021).
Busse, M. M. et al. Comparative life cycle assessment of water disinfection processes applicable in low-income settings. Environ. Sci. Technol. 56(22), 16336–16346 (2022).
Jones, C. H. et al. Life cycle environmental impacts of disinfection technologies used in small drinking water systems. Environ. Sci. Technol. 52(5), 2998–3007 (2018).
Hsu, M. S. et al. Efficacy of chlorine dioxide disinfection to non-fermentative Gram-negative bacilli and non-tuberculous mycobacteria in a hospital water system. J. Hosp. Infect. 93(1), 22–28 (2016).
Luo, X. et al. Advances in application of ultraviolet irradiation for biofilm control in water and wastewater infrastructure. J. Hazard Mater. 421, 126682 (2022).
Marois-Fiset, J. T. et al. Laboratory evaluation of a new coagulant/disinfectant point-of-use water treatment product for emergencies. J. Appl. Microbiol. 121(3), 892–902 (2016).
Wolf, J. et al. Effectiveness of interventions to improve drinking water, sanitation, and handwashing with soap on risk of diarrhoeal disease in children in low-income and middle-income settings: A systematic review and meta-analysis. Lancet 400(10345), 48–59 (2022).
Zhou, Z. Y. et al. Removal of waterborne pathogens from liver transplant unit water taps in prevention of healthcare-associated infections: A proposal for a cost-effective, proactive infection control strategy. Clin. Microbiol. Infect. 20(4), 310–314 (2014).
Baron, J. L. et al. Field evaluation of a new point-of-use faucet filter for preventing exposure to Legionella and other waterborne pathogens in health care facilities. Am. J. Infect. Control 42(11), 1193–1196 (2014).
Parkinson, J. et al. Point-of-use filters for prevention of health care-acquired Legionnaires’ disease: Field evaluation of a new filter product and literature review. Am. J. Infect. Control 48(2), 132–138 (2020).
Bielefeldt, A. R. et al. Bacterial treatment effectiveness of point-of-use ceramic water filters. Water Res. 43(14), 3559–3565 (2009).
He, A. et al. Exploring the origin of efficient adsorption of poly- and perfluoroalkyl substances in household point-of-use water purifiers: Deep insights from a joint experimental and computational study. Sci. Total Environ. 831, 154988 (2022).
Yetiş, Ö. et al. Failure of a hollow-fibre shower filter device to prevent exposure of patients to Pseudomonas aeruginosa. J. Hosp. Infect. 130, 1–6 (2022).
Baranovsky, S. et al. Tracking the spread routes of opportunistic premise plumbing pathogens in a haematology unit with water points-of-use protected by antimicrobial filters. J. Hosp. Infect. 98(1), 53–59 (2018).
Saran, A. et al. Burkholderia cepacia complex nosocomial outbreaks in India: A scoping review. Indian J. Med. Res. 160(6), 593–605 (2024).
Fomda, B. et al. An outbreak of Burkholderia cepacia bloodstream infections in a tertiary-care facility in northern India detected by a healthcare-associated infection surveillance network. Infect. Control Hosp. Epidemiol. 44(3), 467–473 (2023).
Duan, Z. et al. Molecular epidemiology and risk factors of Stenotrophomonas maltophilia infections in a Chinese teaching hospital. BMC Microbiol. 20(1), 294 (2020).
Ishii, H. et al. Pseudomonas fluorescens pneumonia. Int. J. Infect. Dis. 140, 92–94 (2024).
Oba, Y. et al. Longitudinal nosocomial outbreak of Pseudomonas fluorescens bloodstream infection of 2 years’ duration in a coronary care unit. Am. J. Infect. Control 45(8), e75–e79 (2017).
Ahn, Y. et al. Oligotrophic media compared with a tryptic soy agar or broth for the recovery of Burkholderia cepacia complex from different storage temperatures and culture conditions. J. Microbiol. Biotechnol. 29(10), 1495–1505 (2019).
Maltais, J. B. et al. Comparison of techniques for culture of dialysis water and fluid. Hemodial. Int. 21(2), 197–205 (2017).
Marek, A. et al. Endoscopy supply water and final rinse testing: five years of experience. J. Hosp. Infect. 88(4), 207–212 (2014).
Walker, J. T. et al. Final rinse water quality for flexible endoscopy to minimize the risk of post-endoscopic infection. Report from Healthcare Infection Society Working Party. J. Hosp. Infect. 124, 79–96 (2022).
Gavaldà, L. et al. Role of hot water temperature and water system use on Legionella control in a tertiary hospital: An 8-year longitudinal study. Water Res. 149, 460–466 (2019).
Acknowledgements
We deeply appreciate the contribution to this thesis made in various ways by all members in the Department of Nosocomial Infection Management, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. In addition, we are also very grateful for the help of Pall Medical.
Funding
This work was supported by the Hubei Province Public Health Leadership Talent Program, and Scientific Cultivation Fund, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology [Grant Number F016010032200318706]. The funding agreements ensured the authors’ independence in designing the study, interpreting the data, writing, and publishing the report.
Author information
Authors and Affiliations
Contributions
XC, HM and LC contributed equally to this work. XC, YF and LX made substantial contributions to the conception or design of the work; XC, HM and YW took responsibility for the integrity of the data and the accuracy of the data analysis. XC and HM drafted the manuscript. LC, FG and HX conducted the statistical analyses. LX managed the project and provided guidance. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Cao, X., Meng, H., Cai, L. et al. Exploring the efficacy of combination of point-of-use filters and peracetic acid disinfection in reducing total viable counts in final rinse water of endoscopes. Sci Rep 15, 16570 (2025). https://doi.org/10.1038/s41598-025-01701-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-025-01701-z