Abstract
Soft X-ray experiments at synchrotron light sources are essential for a wide range of research fields. However, commercially available detectors for this energy range often cannot deliver the necessary combination of quantum efficiency, signal-to-noise ratio, dynamic range, speed, and radiation hardness within a single system. While hybrid detectors have addressed these challenges effectively in the hard X-ray regime, specifically with single photon counting pixel detectors extensively used in high-performance synchrotron applications, similar solutions are desired for energies below 2 keV. In this work, we introduce a single photon counting hybrid pixel detector capable of detecting X-ray energies as low as 550 eV, utilizing the internal amplification of Low Gain Avalanche Diode (LGAD) sensors. This detector is thoroughly characterized in terms of Signal-to-Noise Ratio and Detective Quantum Efficiency. We demonstrate its capabilities through ptychographic imaging at MAX IV 4th-generation synchrotron light source at the Fe L3-edge (707 eV), showcasing the enhanced detection performance of the system. This development sets a benchmark for soft X-ray applications at synchrotrons, paving the way for significant advancements in imaging and analysis at lower photon energies.
Similar content being viewed by others
Introduction
Several imaging techniques rely on low energy X-rays, including Scanning Transmission X-ray Microscopy (STXM), full-field Transmission X-ray Microscopy (TXM), anomalous diffraction, X-Ray Magnetic Circular and Linear Dichroism (XMCD and XMLD) and Coherent Diffractive Imaging (CDI)1,2,3,4,5. Working at the absorption edges within the soft X-ray range between 250 eV and 2 keV offers enhanced sensitivity to specific chemical species. For instance, the K-edges of light elements such as C, O, N, F, and P, which are common in many organic and biological systems6,7, as well as the L-edges of 3d transition metals like Cu, Ni, and Fe, as well as the M-edges of rare earth elements, which are crucial for research on magnetic, ferroelectric, and electrode materials8,9,10,11,12, fall within this range.
In the early 2000s, hybrid photon-counting pixel detectors with silicon sensors revolutionized tender and hard X-ray applications at synchrotrons, replacing previously used Charge-Coupled Devices (CCDs). These detectors greatly advanced a range of experimental techniques, including macromolecular crystallography (MX), small-angle scattering (SAXS), CDI, and powder diffraction13,14, thanks to their superior dynamic range and the absence of read-out noise. Moreover, their high frame rate enabled methods such as fine φ-sliced and time-resolved crystallography15, as well as scanning microscopy techniques such as ptychography16.
To date, the use of single photon counting detectors has been largely limited to hard and tender X-ray energy ranges. The lowest reported energy at which Single Photon Counting (SPC) hybrid pixel detectors have been used is 1.75 keV17. To the best of our knowledge, position-sensitive SPC detectors for lower energy photons have not yet been available. The primary reason is that the small signals produced by single photons are nearly indistinguishable from the electronic noise of the detector18.
Hybrid detectors consist of a sensor, which absorbs radiation and converts it into electric charge, connected to an Application Specific Integrated Circuit (ASIC) via high-density flip-chip bonding on a pixel-by-pixel basis, as illustrated in Fig. 1a. Sensors must be optimized to maximize the fraction of incident photons that are absorbed, converted into electric charge, and successfully collected. This is referred to as the Quantum Efficiency (QE). However, standard silicon sensors are typically not ideal for achieving high QE for low-energy photons. This is due to their limited ability to collect charge carriers generated just below the entrance window of the sensor. The primary advantage of the hybrid approach is that the sensor and readout electronics can be optimized independently, with few technological constraints, as demonstrated in this work. This is a significant advantage over monolithic detectors, in which the sensor and readout electronics are integrated onto the same ASIC19.
a Layout of the hybrid pixel detector used. The iLGAD sensor, where X-rays are converted into electric charge, is connected pixel-by-pixel to the EIGER readout electronics using bump bonding. The readout chip is biased, controlled, and read out through wire-bond pads. b Sketch of the cross-section of the inverse-LGAD sensor. X-rays enter from the top through a thin entrance window composed of passivation and a shallow n+ doping layer. The p+ gain layer forms a junction with a high electric field, enabling charge multiplication via impact ionization. The signal is induced in the p+ pixels, which are connected to the readout electronics using bump bonding. c Multiplication factor M as a function of photon absorption depth for standard (red) and shallow (black) iLGAD variations45. The blue lines and secondary y-axis axis represent the transmission of 500 eV (dashed) and 1000 eV (dash-dotted) photons.
Monolithic detectors with an optical entrance window generally have higher QE in the soft X-ray range compared to hybrid detectors and exhibit lower electronic noise thanks to their low input capacitance20. Consequently, low-energy applications often use CCDs21,22,23 and CMOS sensors24,25, despite their limitations in frame rate, dynamic range, and radiation hardness. Recent developments, however, aim to address and overcome these drawbacks26,27,28. Ptychography in the soft X-ray energy range is mostly performed using fast CCD devices29,30 or CMOS sensors28,31,32,33,34.
Charge integrating hybrid detectors with low electronic noise can operate at lower energies, down to a few hundred eV35 or even into the EUV range without single photon resolution36,37, but they come with the drawback of limited dynamic range and more complex operation. The development of hybrid single photon counting detector systems for soft X-rays, with performance comparable to those currently achieved at higher energies, could significantly improve many experimental techniques currently limited by available detector performance1.
This study presents the results achieved by combining the EIGER single photon counting readout chip38 with Low Gain Avalanche Diode (LGAD) sensors optimized for soft X-ray detection39. The system takes advantage of the hybrid architecture’s flexibility, substituting the standard silicon p-in-n pixel sensor with inverse LGAD devices and using their internal signal amplification to detect single low-energy photons18,40. To the best of our knowledge, this work marks the first demonstration of a single photon counting pixel detector capable of detecting X-rays down to 550 eV. We present the performance of the system along with initial experimental results in ptychographic imaging at the Fe L3-edge at 707 eV.
Results
Detector description
LGAD sensors are based on an n-in-p silicon junction and incorporate an additional layer, moderately doped with the same polarity as the substrate. This region, typically created via ion implantation and referred to as the gain implant, is located just beneath the shallow surface junction. When fully depleted, this p+ region exhibits an electric field high enough to enable the generation of secondary charge carriers by impact-ionization when electrons or holes travel through it. While this concept has been widely employed in avalanche photodiodes (APDs) and single-photon avalanche detectors (SPADs), LGADs are specifically designed to target a low multiplication factor M of ~5–20, avoiding dark counts while allowing segmentation. Originally proposed in the early 2010s to achieve high timing resolution in tracking detectors for high-energy physics experiments41,42,43, this effect can also be used to amplify signals from low-energy photons, boosting their signal above the noise of the readout electronics18, similar to the noise reduction achieved in Electron-Multiplying CCDs (EM-CCD)44. However, compared to the EM-CCD, the multiplication happens before charge collection and storage instead of at the time of readout, without affecting the frame rate. The LGAD sensors used in this study have a uniform gain layer directly beneath the entrance window, as shown in Fig. 1b, enabling a 100% fill-factor, while the fine pixel segmentation is fabricated on the opposite side of the sensor, allowing the hybridization with the read-out electronics. These sensors are referred to in the literature as inverse-LGADs (iLGADs)40.
Maximum charge multiplication occurs when an incident photon is absorbed in the sensor bulk, behind the gain layer. The generated electrons then drift toward the entrance window, crossing the gain layer and producing secondary electrons and holes that drift towards the entrance window and the pixel side, respectively. When photons are absorbed at shallower depths, i.e., in the n+ region or within the gain layer, charge multiplication is initiated either partially or entirely by the holes crossing the gain layer and drifting toward the pixel contacts. Since the ionization coefficient of holes is lower than that of electrons in silicon, this results in a lower effective multiplication factor M, as illustrated in Fig. 1c, which shows M as a function of the depth of the photon absorption. To address this, iLGAD variations with a shallower gain layer were investigated. In the standard design, the gain layer extends to a depth of 800 nm, while in the shallow design, it reaches about half of that depth, thus increasing the fraction of electron-initiated events45.
To overcome the poor QE of conventional silicon sensors in the soft X-ray range, a custom entrance window process has been developed39. This optimized entrance window is based on a customized doping profile and includes a thin SiO2 and Si3N4 passivation layer on the surface, enhancing charge collection efficiency and achieving a QE of up to 55% at 250 eV, limited only by the thickness of the passivation layer45,46. In comparison, conventional sensors typically achieve a QE of less than 5% at this energy17.
These LGAD sensors are combined with the 75 μm pitch single photon counting EIGER ASIC, widely used in hard X-ray applications worldwide47,48,49,50,51. Additional details about the EIGER ASIC are available in “Methods” section. The system comprises 2 × 2 ASICs bump-bonded to a 275 μm thick iLGAD sensor with 512 × 512 pixels, covering an area of 4 × 4 cm2, as shown in Fig. 2a. The sensors used in this study, fabricated by Fondazione Bruno Kessler (FBK, Trento, Italy), to the best of our knowledge, are the largest LGAD sensors reported to date. Figure 2b demonstrates the detector’s capability to detect photons down to energies of 550 eV by imaging the far field diffraction from a Fresnel Zone Plate focusing optic at this low energy. The Detective Quantum Efficiency (DQE) is estimated to be about 30% (see section 5) and logarithmic color scale emphasizes both the absence of noise and the large dynamic range.
a Picture of a 4 × 4 cm2 inverse-LGAD sensor bump-bonded to an EIGER ASIC, mounted on a vacuum-compatible flange. b Image of the far-field diffraction from a Frenel Zone Plate focusing optics projected on the detector surface acquired with a 550 eV photon beam (single frame with 1 s exposure time). Additional diffraction and scattered X-rays are visible in the background. The set threshold corresponds to ~480 eV. Fewer than 1% noisy pixels have been masked out.
Performance of the system
Figure 3a shows the average calibrated pulse-height distributions measured for a shallow iLGAD sensor variation, with multiplication factor 4.8, with photon energies between 550 eV and 900 eV. Although detectable, photon energies below 550 eV are not shown since at these energies the Signal-to-Noise Ratio (SNR) and the DQE decreases considerably. Additionally, at lower energies the multiplication is mostly triggered by holes, resulting in a lower effective gain, since the majority of the photons convert before or in the gain layer. Consequently, signals from hole-triggered multiplication are often indistinguishable from noise. The main peak in the curves results from photons absorbed in the sensor bulk behind the gain layer undergoing electron-initiated multiplication with the totality of the charge collection in a single pixel. The flat region preceding the peak is caused by two factors: i) photons that convert at a shallower depth, which results in a lower charge multiplication factor, and ii) charge-sharing effects, where the generated charge is distributed across multiple readout pixels52. In the graph, the threshold level at with a Signal-to-Noise ratio of 5 is reached is highlighted, which is equal to 447 eV, for this variation.
a Calibrated pulse-height distribution as a function of the X-ray photon energy for the shallow iLGAD variation, M = 4.8. The spectra are obtained as average after a pixel-wise calibration, and normalized to the peak integral. The gray band and horizontal arrow indicate the level at which a SNR of 5 is achieved, considering the ASIC settings optimized for 550 eV. The ASIC settings have been optimized for the individual energies. b Comparison of the calibrated pulse-height distributions obtained using a standard and a shallow iLGAD sensor variation at 900 eV, 700 eV, and 550 eV. The ASIC settings have been optimized for the individual energies and iLGAD sensor variations. Vertical lines indicate the levels at which a SNR = 5 is reached for each of the two sensor variations, considering the ASIC settings used for 550 eV. The curves have been normalized at the peak integral.
Figure 3 b compares the average pulse-height distributions for standard and a shallow iLGAD sensors (with M = 6.1 and M = 4.8) respectively at 550 eV, 700 eV, and 900 eV. The standard iLGAD variation shows an increased flat region below the full-energy peak, as fewer photons convert within the thinner gain layer. The shift towards lower energies of the peak position, in the case of the standard variation, can be attributed to the same effect, as a smaller fraction of the events undergoes an optimal multiplication. Thanks to the higher M, a SNR of 5 is reached at a marginally lower threshold, 425 eV. However, a smaller fraction of photons will be detected, compared to the shallow variation. The ratio of photons undergoing electron-triggered multiplication to those absorbed before or within the gain layer influences the DQE of the detector. Charge-sharing effects are expected to be consistent across iLGAD variations, as they are influenced primarily by pixel size and charge collection time, which are determined by sensor thickness, bias voltage, and X-ray energy53. The calibration method and the procedure for obtaining the average pulse-height distributions shown in Fig. 3 are detailed in “Methods” section.
Figure 4a shows the noise as a function of the calibration gain \({{\mathcal{G}}}\) of the detector, which includes the combined gain of the readout electronics and the iLGAD multiplication factor M. The calibration gain is the conversion factor between signal amplitude and photon energy and it is expressed in mV/eV. \({{\mathcal{G}}}\) depends on the preamplifier gain settings of the ASIC and on the LGAD multiplication factor M (see “Methods”). The noise is measured as Equivalent Noise Charge (ENC), expressed in electrons, which is the input signal required to produce an output equal to the root mean square (rms) noise. The gain and noise values are extracted for each pixel and averaged. The error bars indicate the standard deviation of the pixel distribution (see Supplementary Notes 1 and 2). A higher \({{\mathcal{G}}}\) is generally linked to a lower effective noise, as visible in Fig. 4a, and the use of LGADs significantly improves performance compared to an EIGER ASIC paired with a conventional silicon sensor under the same settings, as the signal is further amplified by the multiplication process in the sensor. A fit with a reciprocal function is added to the plot.
a Effective noise as a function of calibration gain \({{\mathcal{G}}}\), of four iLGADs samples of both standard and shallow variations, compared to a conventional silicon sensor. The data points are fitted with a function y = a/x, as the Effective Noise is expected to have an inverse relation with the calibration gain \({{\mathcal{G}}}\). For the data points two different settings of the EIGER ASIC are used, differentiated in the plot by filled or empty markers. b Noise reduction factor as a function of the LGAD multiplication factor M, with respect to a conventional silicon sensor, for the four iLGADS samples of both variations. The dashed line indicates the ideal case where the noise reduction scales linearly with M; the data points fall below this trend and are tabulated in the Supplementary Table 1. In both plots, the error bars indicate the standard deviation of the pixel distribution.
Ideally, the noise reduction of an LGAD sensor with respect to a conventional Si sensor scales linearly with M when the same ASIC settings are used. For instance, with an LGAD sensor providing a M ~ 3, the effective noise is reduced from 90e−rms to less than 30e−rms. However, with higher M, the noise reduction becomes less efficient, due to factors such as the excess noise from the multiplication process and the increased leakage current54. In this work, effective noise levels as low as 23e−rms were achieved. As shown in Fig. 4b, for M lower than 4, the noise is effectively reduced by a similar factor. However, as M increases, the noise improvement becomes sub-linear. For instance, at a gain of around 5, the noise reduction is approximately a factor of 4, whereas for M ~ 11, the noise is reduced by less than a factor of 6. There is no noticeable difference in noise reduction between the shallow and standard iLGAD designs. However, a thicker gain layer typically enables a higher multiplication factor.
The SNR as a function of photon energy is shown in Fig. 5a. The average of the SNR calculated for each individual pixel is shown with the error bars indicating the standard deviation of the distribution. Both standard and shallow sensor variations achieve an SNR above 10 at about 850 eV. At 550 eV, the estimated SNR values are about 5.4 for both variations. The SNR varies only marginally between the two variations over the explored energy range. The optimal performance of a single photon counting detector is obtained for an SNR ≥ 1055 and the threshold set at half of the photon energy. An optimal threshold at half of the photon energy, and SNR ≥ 5, can therefore be set only above 800 eV. While the detector can operate with a lower SNR, this compromises the Counting Efficiency (CE) because the threshold must be set higher than half the photon energy. Additional details are provided in the Supplementary Note 2.
a Measured Signal-to-Noise ratio for both iLGAD variations. The SNR is calculated for each single pixel and averaged. b Calculated Detective-Quantum-efficiency at zero spatial frequency as a function of photon energy. The reported DQE represents the fraction of incoming photons that can be counted; the threshold is considered set at 50% of the incoming photon energy for 800 eV and higher, and at SNR = 5 for lower energies. The ideal Counting Efficiency (CEideal) takes into account only the loss of counts due to charge-sharing, when the threshold is set at half of the incoming energy, considering an ideal QE and noise-less electronics. QELGAD is the measured Quantum Efficiency45. DQEideal is the product between CEideal and QELGAD, i.e., the ideal Detective Quantum efficiency achievable by an LGAD sensor, without losses due to the multiplication process and considering a noiseless electronics. In both plots, the error bars indicate the standard deviation of the pixel distribution. Additional details on the SNR estimation and DQE calculation can be found in the Supplementary Notes 1 and 2.
Figure 5b shows the calculated DQE of the detector as a function of photon energy, at zero spatial frequency. The DQE combines the QE of the sensor and the CE with the threshold set at the greater of either half the incoming energy or the energy at which SNR=5. It is a key figure of merit that describes how effectively a photon is converted into useful information in the final image. The calculation is based on the iLGAD parameters, measured in ref. 45, and on the measured noise levels and multiplication values from this work. The ideal Counting Efficiency CEideal is defined as the fraction of total incident photons detectable with a threshold at half of the photon energy, assuming a noiseless electronics and 100% QE, it depends on the charge sharing and is only affected by photon loss in the pixel corners56. For a detector with pixel pitch as the ones used in this study, the maximum CE is ~93%; the charge-sharing effect has been estimated from the average pulse-height distributions in Fig. 2, with a procedure reported in Supplementary Note 3.
For SNR > 10, i.e., down to 800 eV, the DQE of the detector is close to the ideal value, given by the product between the QE of the sensor and the ideal CE, indicated as DQEideal in Fig. 5b. The DQE decreases with the energy, due to the lower QE of the sensor, as well as to a reduced counting efficiency. This reduction in CE arises from the inability to set the threshold at half the photon energy because of noise constraints. Additionally, at low energies, a higher proportion of photons are absorbed before or within the gain layer. Consequently, the shallow variation with its thinner and shallower gain layer achieves a higher DQE over the whole energy range.
Ptychographic imaging
Ptychography is a scanning CDI method that has been highly successful with hard X-rays, with resolution far surpassing the pixel size of the detector and the beam spot size limitations of direct imaging16. Single photon counting detectors are routinely used for recording diffraction patterns in far-field geometry, providing the high dynamic range needed to detect both the strong signals from the central undiffracted light cone and the weak signals from scattered photons at high Q-values. A fast frame rate is essential to match the high flux provided by modern X-ray sources and to facilitate rapid scanning57. Moreover, burst ptychography using a multi-kHz framing detector promises to further improve the resolution limits of this technique by reducing the effects of mechanical instabilities51. We focus on ptychography because, along with diffraction, it is one of the principal experimental techniques that has significantly benefited from single photon counting pixel detectors in the hard X-ray regime. In contrast, due to the stringent spatial resolution requirements, TXM primarily relies on CCD and CMOS sensors, while STXM does not benefit from the use of position sensitive detectors and usually relies on a point detector. The development of the EIGER-iLGAD detector now enables the extension of high-throughput, high-resolution ptychographic imaging into the soft X-ray energy range.
An EIGER-iLGAD detector with shallow sensor variation was installed at the Surface/Interface Microscopy (SIM-X11MA) beamline of the Swiss Light Source (SLS), Paul Scherrer Institut, Villigen, Switzerland58 for commissioning and proof-of-principle experiments until the facility shut down for the upgrade to SLS 2.0 in September 2023. The detector was then installed in the SOPHIE endstation, which went into operation at the SoftiMAX beamline of the MAX IV synchrotron in Lund (Sweden), where it was available to the synchrotron user community until the end of June 2025.
In order to demonstrate the suitability of the detector for ptychographic imaging, a Siemens star imaged by soft X-ray ptychography at 707 eV, where the DQE of the detector is ~50%. The Siemens star was fabricated on a Si3N4 membrane by electron beam lithography using negative-tone hydrogen silsesquioxane (HSQ) resist. The developed HSQ resist was then coated with a 10 nm Ir film by atomic layer deposition (ALD). Details about the lithographical patterning method can be found in ref. 59. The diffraction patterns for the ptychographic scan were obtained with an exposure time of 2.5 ms. The comparatively large pixel size of 75 μm does not represent a hindrance for ptychography, as the sampling criterion of diffraction patterns in regular CDI is not as stringent for this technique60,61.
The resulting amplitude and phase images are displayed in Fig. 6a and b, respectively. The spatial resolution achieved in this image was estimated using Fourier Ring Correlation (FRC) between two identical ptychographic scans62, suggesting a resolution of 6.5 nm with a 1 bit threshold.
a Amplitude and b Phase ptychographic images of a Siemens star. These images were obtained with an exposure time of 2.5 ms. c X-ray linear dichroism image (phase contrast) showing the in-plane projection of the ferroelectric polarisation (P). Mosaic-like multiferroic domains are visible, which contain spin cycloids (64 nm period) in an 80 nm thin freestanding BiFeO3 (001) film. The crystallographic axes of the BiFeO3 film are indicated above the color bar. The contrast indicates both the in-plane axis of P in the BiFeO3 film and delineates the spin cycloid coupled within the multiferroic domains (see refs. 63,64 for details). All images were acquired with an X-ray energy of ~707 eV (Fe L3-edge). At this energy, the maximum of the XLD contrast originating from the BFO ferroelectric domains is found in the phase information.
Figure 6c demonstrates the detector’s performance in an X-ray linear dichroism (XLD) image of a freestanding BiFeO3 (001) film. Additional images of the multiferroic domains structure in freestanding BiFeO3, acquired with the detector presented here, have been published in refs. 63,64. The results in ref. 64 demonstrate that the EIGER-iLGAD detector is also capable of operation at the O K-edge (530 eV), even below 550 eV. However, while it remains functional at these lower energies, there is a reduction in DQE, combined with a lower X-ray scattering cross section from the ferroelectric domains of BFO, resulting in a lower spatial resolution in the reconstructed image. These results were obtained with a longer exposure time of 200 ms, which ensures high spatial resolution. Soft X-ray ptychography with the LGAD EIGER has also been employed to study magnetization dynamics in permalloy (Ni81Fe19) microstructures by XMCD imaging in pump-probe mode65.
Discussion
The presented system, combining the EIGER single-photon counting pixel ASIC with custom-developed LGAD sensors, successfully extends hybrid detector capabilities into the soft X-ray energy range, pushing the lower limit for single-photon resolution down to ~550 eV. This system achieves a SNR above 5, without compromising performance in frame rate, dynamic range, or noise-free operation, that are intrinsic of the photon-counting architecture. The energy range of the presented detector can be even pushed down to lower energies, by sacrificing its DQE. In general, the shallow iLGAD design offers higher DQE across all X-ray energies. Further increasing the M while maintaining low noise performance could be beneficial for extending detection to even lower energies.
Despite the promising results, further developments are needed before this detector can fully meet the requirements of the entire soft X-ray imaging community. Improvements in both pixel and sensor yield are essential. Currently, the pixel yield is around 97%, with significant variability observed between different sensors, likely due to non-uniformity of the multiplication factor, as discussed in ref. 66. The pixel yield declines rapidly at higher temperatures and lower bias voltages, suggesting that leakage current may be causing saturation in the analog chain. Reducing the leakage current through sensor technology advancements would enable detection at lower photon energies and reduce effective noise. Enhanced cooling and optimized biasing could similarly improve performance. Additionally, the fabrication of large-area iLGAD of 4 × 4 cm2 with high production yield still has to be demonstrated, considering that the first prototyping batch showed a yield limited to about 30%. A higher yield is necessary to reliably build multi-megapixel detectors, as is standard practice with conventional silicon sensors for hard X-rays.
Additional improvements can be achieved through advances in both readout electronics and iLGAD sensor technology. While reducing noise in single photon counting detectors to improve the minimum detectable energy is challenging, and likely limited to a 30-50% improvement in the short term, significant progress is being made to increase count-rate capabilities67, a crucial step for operating single photon counting detectors at diffraction-limited light sources currently under construction worldwide. The high gain provided by LGADs enables faster signal shaping, which is essential for achieving the required count rate capabilities.
In the future, even lower energies could be achievable through the development of a lower-noise photon-counting ASIC paired with LGAD sensors. Simultaneously, advancements in LGAD technology are underway to increase the M up to 20, enabling detection at lower energies and improving sensor QE by further thinning the entrance window. From the sensor perspective, the QE can be enhanced by reducing the passivation layer on the entrance window, while a higher counting efficiency may be achieved with even shallower and thinner gain implants. However, increasing M to access lower energies is effective only if the leakage current, which raises shot noise and decreases pixel yield, remains controlled.
Single photon counting LGAD detectors will be instrumental for applications at soft and tender X-ray beamlines with the goal of extending to the EUV energy range with sensors with higher multiplication factors. Moreover, iLGAD sensors combined with charge-integrating readout find applications at X-ray Free Electron Lasers66. The possibility of applying position interpolation methods opens perspectives also for high resolution imaging, including Resonant Inelastic X-ray Scattering (RIXS), where single photon detection combined with high spatial resolution is essential and high frame rate detectors are lacking. The advancement of single photon counting detectors for soft X-rays enables new applications in low-energy photon science, offering high sensitivity and precision that promise to expand capabilities in fields from biological microscopy to materials science.
Methods
iLGAD sensors
Low Gain Avalanche Diodes (LGADs) are silicon-based sensors that include an additional moderately-doped implanted region (gain implant) of the same type of the substrate, where charge carriers are accelerated by a strong electric field (of the order of 300 kV/cm), initiating impact ionization and achieving a charge multiplication of approximately a factor of 10. The multiplication factor is mostly determined by the doping profile and concentration of the gain implant. The original LGADs developed for high-energy physics43 feature a coarse segmentation, with about 1 mm pitch pixels and the gain implant matching the pixel layout. To prevent breakdown at pixel edges, termination structures are needed to create gain-free regions, which effectively reduces the fill-factor i.e., the proportion of the pixel where the charge produced by X-rays undergoes multiplication68. Various LGAD technologies aim to minimize or eliminate this limitation 69. For the soft X-ray sensor presented here, an inverse-LGAD (iLGAD) design was used, featuring a uniform gain layer on the entrance window of the sensor40. The sensors used in this study are 275 μm thick. The sensors are fabricated using p silicon wafers, with p+ pixel implants. A n+ implant on the entrance window side forms the collecting junction, and the additional p+-type gain implant is formed just underneath the junction. Such a gain implant is unsegmented and it extends through the full sensor area, as shown in Fig. 1b. The sensors operate with hole collection, where holes drift toward the pixel while electrons move in the opposite direction (see Fig. 1b). The sensors were operated with a bias voltage of 300 V, although due to possible voltage drops along the bias line, the effective voltage reaching the sensors may be lower. The bias voltage has a small effect on the multiplication factor, while it can affect charge sharing.
The impact ionization can be triggered by electrons, when a photon is absorbed after the gain layer in the p-type bulk of the sensor or by the holes if a photon is absorbed before the gain layer. A mix of these two processes occurs when a photon is absorbed within the gain layer. Holes have a multiplication factor that is 2–4 times lower than electrons, due to their lower impact ionization coefficient, depending on the electric field and the doping profile of the gain layer. The multiplication factor values M for iLGAD sensors refer specifically to electron-triggered multiplication45. The standard iLGAD variation has a gain layer thickness and doping profile similar to LGADs designed for tracking applications, while the shallow variation features a thinner gain layer closer to the surface to reduce hole-initiated multiplication. The iLGAD sensors presented here have an entrance window optimized for soft X-rays, featuring a thin SiO2 and Si3N4 passivation layer for a total of about 80 nm instead of the aluminum layer used in conventional silicon sensors. Additionally, the doping species, profile, and concentration in the n+-implant have been optimized. A new batch of sensors with even thinner passivation is under development, which is expected to further enhance the QE.
EIGER single-photon counting read-out
The EIGER is a photon-counting ASIC, developed at PSI for diffraction experiments at synchrotron light sources. It features a 75 μm pixel pitch, with each ASIC comprising 256 × 256 pixels and covering an area of 2 × 2 cm238. Each pixel includes a low-noise charge preamplifier and shaper with adjustable gain and speed, a comparator with a globally set threshold, a 6-bit DAC (trim bits) to correct pixel-by-pixel inhomogeneities, and a 12-bit counter. The comparator in each pixel is triggered whenever the shaper signal exceeds the threshold, counting events in the pixel counter, which is read out after the exposure time. The gain of the analog chain is controlled by adjusting the preamplifier feedback resistance, set through a gate voltage. This adjustment allows a trade-off between noise (as low as ~100e− ENC rms70) and count-rate capability (up to ~1 Mcounts/pixel/s71). The low noise settings used in this work for low energy detection allow to achieve a count rate of the order of ~100kcounts/pixel/s. The 12-bit counter can be extended to 32-bits in firmware with minimal dead time between sub-frames, and the bit depth can be adjusted to reach frame rates up to 22 kHz in 4-bit mode.
In this study, the detector is built by arranging 2 × 2 ASICs (4 × 4 cm2, 512 × 512 pixels) bump-bonded to a single sensor of the same area. A 150 μm gap between adjacent ASICs creates a cross-shaped blind area, visible in Fig. 2a. Additionally, along the sensor edges, the outer 9 pixels are replaced by guard-ring structures to prevent breakdowns, making them blind. In this work, the threshold voltage is set uniformly across all four ASICs in the detector system, and the trim-bits correction is not used. The sensor and electronics are operated in vacuum at pressures lower than 10−5 mbar and at −28 °C, using a liquid chiller. Lower temperatures help reduce sensor leakage current and increase the LGAD multiplication factor, thereby enhancing the SNR39,66.
The detector is connected to the readout board via flat-band cables and a vacuum flange patch panel, limiting the readout clock speed to half of its nominal value and capping the maximum frame rate at 10 kHz.
Calibration
The characterization and calibration of the detector are performed via threshold scans, where counts are measured as a function of the comparator threshold using a constant-flux monochromatic photon beam. An example of the resulting s-curves is shown in Fig. 7a. The uncalibrated pulse-height distributions shown in Fig. 7b are obtained as the derivative of the threshold scans. Various iLGAD variations with different multiplication factor M, at 900 eV, averaged over a large number of pixels are compared. The preamplifier settings of the EIGER ASIC are the same for all the datasets.
a Uncalibrated average threshold-scan (s-curves) for three different sensor variations, with photon energy of 900 eV, and same settings of the EIGER ASIC. The markers represent experimental data and the continuous lines the fit with the function described by Supplementary Equation 1. b Uncalibrated average pulse height distribution, obtained as a derivative of the measured s-curves. c Correlation between photon energy and signal amplitude for three iLGAD variations of different M; the values are calculated as average over multiple pixels and the error bars represent their standard deviation; the energy calibration is performed via a linear fit (dashed lines), where the slope is \({{\mathcal{G}}}\).
The threshold scans can be fitted on a pixel-by-pixel basis using an s-curve function, as described in ref. 72. This fit enables extraction of the photon count, full-energy peak position, noise level, and fraction of shared charge. Pixel-wise calibration is performed with a linear fit of the peak position in the pulse-height distribution as a function of the beam energy. Figure 7c shows the position of the full-energy peak as a function of the photon energy, with linear fits; the data points represent the average and the error bars the standard deviation of the distribution of the pixels. The calibration gain of the detector \({{\mathcal{G}}}\) is defined by the slope obtained from this fit and depends on both the gain settings of the pixel preamplifier and the multiplication factor of the LGAD sensor M.
The noise value, derived from the s-curve fit is converted using the calibration gain \({{\mathcal{G}}}\). It represents the standard deviation of the calibrated photon peak and is expressed in units of electron-hole pairs, which correspond to 3.6 eV per pair in silicon. With a fixed M, a higher \({{\mathcal{G}}}\) generally yields a lower noise level (see Fig. 4a), though it comes at the expense of count-rate capability. The M of an LGAD sensor can be estimated as the ratio of the average conversion gains to that measured with a conventional silicon sensor, under the same preamplifier gain settings (see Supplementary Note 1).
Ptychography
Ptychographic imaging with the EIGER-iLGAD detector was performed using the SOPHIE (Soft X-ray Ptychography Highly Integrated Endstation) endstation. The sample to detector distance was 96 mm. The X-ray beam, tuned to the Fe L3-edge (707 eV), was focused to a 400 nm FWHM spot on the sample using a 500 μm diameter line doubled Ir Fresnel zone plate with 50 nm outer zone width. A step size of 50 nm was chosen for the ptychographic scan on a Fermat spiral trajectory. The ptychographic scan for the Siemens star in Fig. 6a, b was carried out with an exposure time of 2.5 ms, which led to a total scan duration of 15 s with a dwell time of 22.5 ms. The XLD phase image of the BiFeO3 film shown in Fig. 6c was acquired with a 220 ms dwell time for each position, resulting in a total acquisition time of ~3 min for a 1.5 × 1.5 μm2 scan area. A 500 μm diameter line doubled Ir zone plate with an outer zone width of 20 nm was used for imaging the multiferroic domains in the BFO thin film.
The reconstruction of the images was performed with 1200 iterations of difference-map73 and 200 iterations of maximum-likelihood refinement74 implemented in the Ptychoshelves software package75. The illuminating wavefront was reconstructed with three probe modes76.The XLD phase image in Fig. 6c was computed by subtracting the drift-corrected phase images obtained with horizontal and vertical linear X-ray polarization.
Data availability
All the data that support the findings of this study are publicly available in the following repository: [http://doi.psi.ch/detail/10.16907%2Fc0b2168f-d6de-4e3d-bd96-b20a9b505944]. Scripts to access the data, as well as examples of analysis, are available in the repository: [https://doi.org/10.5281/zenodo.15740655].
Change history
25 August 2025
In this article the affiliation details for Author Nicholas W. Phillips were incorrectly given as ‘5Present address: Mineral Resources, CSIRO, Toowoomba, QLD, Australia’ but should have been ‘5Present address: CSIRO, Mineral Resources, Clayton, VIC 3168, Australia’.
References
Hitchcock, A. P. Soft X-ray spectromicroscopy and ptychography. J. Electron Spectrosc. Relat. Phenom. 200, 49–63 (2015).
Wagner, A., Duman, R., Henderson, K. & Mykhaylyk, V. In-vacuum long-wavelength macromolecular crystallography. Acta Crystallogr. Sect. D Struct. Biol. 72, 430–439 (2016).
Van der Laan, G. & Figueroa, A. I. X-ray magnetic circular dichroism-a versatile tool to study magnetism. Coord. Chem. Rev. 277, 95–129 (2014).
Elnaggar, H. et al. Probing the local distortion of Fe sites in Fe3O4 thin films using enhanced symmetry selection in XMCD. Phys. Rev. Mater. 4, 024415 (2020).
El Omari, K. et al. Experimental phasing opportunities for macromolecular crystallography at very long wavelengths. Commun. Chem. 6, 219 (2023).
Liu, F., Brady, M. A. & Wang, C. Resonant soft X-ray scattering for polymer materials. Eur. Polym. J. 81, 555–568 (2016).
Collins, B. A. & Gann, E. Resonant soft X-ray scattering in polymer science. J. Polym. Sci. 60, 1199–1243 (2022).
Heron, J. et al. Deterministic switching of ferromagnetism at room temperature using an electric field. Nature 516, 370–373 (2014).
Scagnoli, V. et al. Observation of orbital currents in CuO. Science 332, 696–698 (2011).
Wise, A. M. et al. Nanoscale chemical imaging of an individual catalyst particle with soft X-ray ptychography. ACS Catal. 6, 2178–2181 (2016).
Olejník, K. et al. Terahertz electrical writing speed in an antiferromagnetic memory. Sci. Adv. 4, eaar3566 (2018).
Yu, Y.-S. et al. Three-dimensional localization of nanoscale battery reactions using soft X-ray tomography. Nat. Commun. 9, 921 (2018).
Henrich, B. et al. Pilatus: a single photon counting pixel detector for X-ray applications. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 607, 247–249 (2009).
Brönnimann, C. & Trüb, P. Hybrid pixel photon counting X-ray detectors for synchrotron radiation. In Synchrotron Light Sources and Free-Electron Lasers: Accelerator Physics, Instrumentation and Science Applications (eds Jaeschke, E., Khan, S., Schneider, J. & Hastings, J.) 1191–1223. https://doi.org/10.1007/978-3-319-04507-8_36-2 (Springer, 2020).
Förster, A., Brandstetter, S. & Schulze-Briese, C. Transforming X-ray detection with hybrid photon counting detectors. Philos. Trans. R. Soc. A 377, 20180241 (2019).
Pfeiffer, F. X-ray ptychography. Nat. Photonics 12, 9–17 (2018).
Wernecke, J., Gollwitzer, C., Müller, P. & Krumrey, M. Characterization of an in-vacuum PILATUS 1M detector. J. Synchrotron Radiat. 21, 529–536 (2014).
Andrä, M. et al. Development of low-energy X-ray detectors using LGAD sensors. J. Synchrotron Radiat. 26, 1226–1237 (2019).
Snoeys, W. Monolithic pixel detectors for high energy physics. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 731, 125–130 (2013).
Radeka, V. Low-noise techniques in detectors. Annu. Rev. Nucl. Part. Sci. 38, 217–277 (1988).
Strüder, L. et al. Large-format, high-speed, X-ray pnCCDs combined with electron and ion imaging spectrometers in a multipurpose chamber for experiments at 4th generation light sources. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 614, 483–496 (2010).
Soman, M. et al. Developing a CCD camera with high spatial resolution for RIXS in the soft X-ray range. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 731, 47–52 (2013).
Popescu, H. et al. COMET: a new end-station at SOLEIL for coherent magnetic scattering in transmission. J. Synchrotron Radiat. 26, 280–290 (2019).
Correa, J. et al. The PERCIVAL detector: first user experiments. J. Synchrotron Radiat. 30, 242–250 (2023).
Mille, N. et al. Ptychography at the carbon K-edge. Commun. Mater. 3, 8 (2022).
Rota, L. et al. Design of ePixM, a fully-depleted monolithic CMOS active pixel sensor for soft X-ray experiments at LCLS-II. J. Instrum. 14, C12014 (2019).
Desjardins, K. et al. Characterization of a back-illuminated CMOS camera for soft X-ray coherent scattering. AIP Conf. Proc. 2054, 060066 (2019).
Desjardins, K. et al. Backside-illuminated scientific CMOS detector for soft X-ray resonant scattering and ptychography. J. Synchrotron Radiat. 27, 1577–1589 (2020).
Shapiro, D. et al. Chemical composition mapping with nanometre resolution by soft X-ray microscopy. Nat. Photonics 8. https://doi.org/10.1038/NPHOTON.2014.207 (2014).
Shapiro, D. A. et al. An ultrahigh-resolution soft X-ray microscope for quantitative analysis of chemically heterogeneous nanomaterials. Sci. Adv. 6, eabc4904 (2020).
Di Pietro Martínez, M. et al. Magnetic X-ray imaging using a single polarization and multimodal ptychography. Phys. Rev. Lett. 134, 016704 (2025).
Vijayakumar, J. et al. Soft X-ray spectro-ptychography of boron nitride nanobamboos, carbon nanotubes and permalloy nanorods. J. Synchrotron Radiat. 30, 746–757 (2023).
Zhang, C. et al. Copper carbon dioxide reduction electrocatalysts studied by in situ soft X-ray spectro-ptychography. Cell Rep. Phys. Sci. 4, 101665 (2023).
Hitchcock, A. P. et al. Comparison of soft X-ray spectro-ptychography and scanning transmission X-ray microscopy. J. Electron Spectrosc. Relat. Phenom. 276, 147487 (2024).
Hinger, V. et al. Advancing the JUNGFRAU detector toward low-energy X-ray applications. J. Intrumentation 17, C09027 (2022).
Redford, S. et al. Calibration status and plans for the charge integrating JUNGFRAU pixel detector for SwissFEL. J. Instrum. 11, C11013 (2016).
Langer, M. et al. Development of a new soft X-ray ptychography spectro-microscope at the Swiss Light Source (SLS). Microsc. Microanal. 24, 54–55 (2018).
Dinapoli, R. et al. EIGER characterization results. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 731, 68–73 (2013).
Zhang, J. et al. Development of LGAD sensors with a thin entrance window for soft X-ray detection. J. Instrum. 17, C11011 (2022).
Paternoster, G. et al. Novel strategies for fine-segmented low gain avalanche diodes. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 987, 164840 (2021).
Sadrozinski, H.-W. et al. Ultra-fast silicon detectors. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 730, 226–231 (2013).
Pellegrini, G. et al. Technology developments and first measurements of low gain avalanche detectors (LGAD) for high energy physics applications. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 765, 12–16 (2014).
Cartiglia, N. et al. LGAD designs for future particle trackers. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 979, 164383 (2020).
Tutt, J. H. et al. The noise performance of electron-multiplying charge-coupled devices at soft X-ray energy values. IEEE Trans. Electron Devices 59, 2192–2198 (2012).
Liguori, A. et al. Characterization of iLGADs using soft X-rays. J. Instrum. 18, P12006 (2023).
Carulla, M. et al. Quantum efficiency measurement and modeling of silicon sensors optimized for soft X-ray detection. Sensors 24, 942 (2024).
Casanas, A. et al. EIGER detector: application in macromolecular crystallography. Acta Crystallogr. Sect. D Struct. Biol. 72, 1036–1048 (2016).
Zinn, T. et al. Ultra-small-angle X-ray photon correlation spectroscopy using the eiger detector. J. Synchrotron Radiat. 25, 1753–1759 (2018).
Holler, M. et al. Three-dimensional imaging of integrated circuits with macro-to nanoscale zoom. Nat. Electron. 2, 464–470 (2019).
Hocine, S. et al. Operando X-ray diffraction during laser 3D printing. Mater. Today 34, 30–40 (2020).
Aidukas, T. et al. High-performance 4-nm-resolution X-ray tomography using burst ptychography. Nature 632, 81–88 (2024).
Mathieson, K. et al. Charge sharing in silicon pixel detectors. Nucl. Instrum. Meth. A 487, 113–122 (2002).
Bergamaschi, A. et al. Looking at single photons using hybrid detectors. J. Instrum. 10, C01033 (2015).
Dalla Betta, G. F. et al. Design and TCAD simulation of double-sided pixelated low gain avalanche detectors. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 796, 154–157 (2015).
Bendat, J. Principles and applications of random noise theory. Phys. Today 12, 46–49 (1977).
Fröjdh, E., Bergamaschi, A. & Schmitt, B. Single-photon counting detectors for diffraction-limited light sources. Fronti. Phys. 12. https://doi.org/10.3389/fphy.2024.1304896 (2024).
Cipiccia, S. et al. Fast X-ray ptychography: towards nanoscale imaging of large volume of brain. Eur. Phys. J. Plus 139. https://doi.org/10.1140/epjp/s13360-024-05224-w (2024).
Quitmann, C. et al. A beamline for time resolved photoelectron microscopy on magnetic materials at the Swiss Light Source. Surf. Sci. 480, 173–179 (2001).
Vila-Comamala, J. et al. Dense high aspect ratio hydrogen silsesquioxane nanostructures by 100 keV electron beam lithography. Nanotechnology 21, 285305 (2010).
Edo, T. B. et al. Sampling in X-ray ptychography. Phys. Rev. A 87, 053850 (2013).
Batey, D. J. et al. Reciprocal-space up-sampling from real-space oversampling in X-ray ptychography. Phys. Rev. A 89, 043812 (2014).
Banterle, N., Bui, K. H., Lemke, E. A. & Beck, M. Fourier ring correlation as a resolution criterion for super-resolution microscopy. J. Struct. Biol. 183, 363–367 (2013).
Butcher, T. A. et al. Ptychographic nanoscale imaging of the magnetoelectric coupling in freestanding BiFeO3. Adv. Mater. 36, 2311157 (2024).
Butcher, T. A. et al. Imaging ferroelectric domains with soft-X-ray ptychography at the oxygen k-edge. Phys. Rev. Appl. 23, L011002 (2025).
Butcher, T. A. et al. Nanoscale domain-wall dynamics in micromagnetic structures with weak perpendicular anisotropy. Phys. Rev. B 111, L220409 (2025).
Hinger, V. et al. Resolving soft X-ray photons with a high-rate hybrid pixel detector. Front. Phys. 12, 1352134 (2024).
Zambon, P. et al. High-frame rate and high-count rate hybrid pixel detector for 4D STEM applications. Front. Phys. 11. https://doi.org/10.3389/fphy.2023.1308321 (2023).
Carulla, M. et al. 50 μm thin low gain avalanche detectors (LGAD) for timing applications. Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 924, 373–379 (2019).
Centis Vignali, M. & Paternoster, G. Low gain avalanche diodes for photon science applications. Front. Phys. 12. https://doi.org/10.3389/fphy.2024.1359179 (2024).
Tinti, G. et al. Performance of the eiger single photon counting detector. J. Instrum. 10, C03011–C03011 (2015).
Radicci, V. et al. EIGER a new single photon counting detector for X-ray applications: performance of the chip. J. Instrum. 7, C02019 (2012).
Kraft, P. et al. Performance of single-photon-counting Pilatus detector modules. J. Synchrotron Radiat. 16, 368–375 (2009).
Thibault, P. et al. High-resolution scanning X-ray diffraction microscopy. Science 321, 379–382 (2008).
Odstrčil, M., Menzel, A. & Guizar-Sicairos, M. Iterative least-squares solver for generalized maximum-likelihood ptychography. Opt. Express 26, 3108–3123 (2018).
Wakonig, K. et al. PtychoShelves, a versatile high-level framework for high-performance analysis of ptychographic data. J. Appl. Cryst. 53, 574–586 (2020).
Thibault, P. & Menzel, A. Reconstructing state mixtures from diffraction measurements. Nature 494, 68–71 (2013).
Acknowledgements
Measurements with soft X-rays were performed at the Surface/Interface Microscopy (SIM-X11MA) beamline of the Swiss Light Source (SLS), Paul Scherrer Institut, Villigen, Switzerland. Additionally, soft X-ray ptychography measurements were carried out with the SOPHIE endstation installed at the SoftiMAX beamline of the MAX IV Laboratory. The SOPHIE endstation was designed and assembled at the SLS, PSI, Villigen, Switzerland. Research conducted at MAX IV, a Swedish national user facility, is supported by the Swedish Research council under contract 2018-07152, the Swedish Governmental Agency for Innovation Systems under contract 2018-04969, and Formas under contract 2019-02496. T.A.B. acknowledges funding from the Swiss Nanoscience Institute (SNI) and the European Regional Development Fund (ERDF). N.W.P. received funding from the European Union’s Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie grant agreement no. 884104. We thank Benedikt Rösner for the fabrication of the Siemens star, Chia-Chun Wei, Shih-Wen Huang and Jan-Chi Yang for the fabrication and characterization of the freestanding BiFeO3 film. We thanks Armin Kleibert and Carlos Vaz for the support at the SIM beamline of the Swiss Light Source. We additionally thank Karina Thånell and Igor Beinik for support at the SoftiMAX beamline at Max IV.
Author information
Authors and Affiliations
Contributions
Conceptualization: F.B. and A.B. Methodology: F.B., A.B., M. Bo., M. Br., M.C., M.C.V., R.D., D.G., E.F., A.M., G.P., and J.Z. Investigation: F.B., A.B., T.A.B., S.F., E.F., and N.W.P. Visualization: F.B., T.A.B., and S.F. Funding acquisition: A.B. and B.S. Project administration: A.B., B.S. Supervision: A.B., J.R. and B.S.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Peer review
Peer review information
Communications Physics thanks the anonymous reviewers for their contribution to the peer review of this work.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Baruffaldi, F., Bergamaschi, A., Boscardin, M. et al. Single-photon counting pixel detector for soft X-rays. Commun Phys 8, 321 (2025). https://doi.org/10.1038/s42005-025-02240-9