Abstract
The chemical shift of many molecules changes with temperature, which enables non-invasive magnetic resonance imaging (MRI) thermometry. Hyperpolarization methods increase the inherently low 13C MR signal. The commonly-used hyperpolarized probe [1-13C]pyruvate, and its metabolic product [1-13C]lactate, exhibit temperature and concentration dependent chemical shift changes that have not previously been reported. These effects were characterized at 7 T and 11.7 T in vitro and applied for in vivo thermometry both preclinically at 7 T and to human data at 3 T. Apparent temperature values from mouse abdomen and brain were similar to rectally measured temperature. Human brain and kidney apparent temperatures from 13C MRSI were lower than known physiological temperatures, suggesting that additional effects may currently limit the use of this method for determining absolute temperature in humans. The temperature dependent chemical shift changes also have implications for sequence design and for in vitro studies with hyperpolarized pyruvate.
Similar content being viewed by others
Introduction
In all living systems, temperature is a critical component of homeostasis. Changes in body temperature were already recognized to be related to disease more than 2000 years ago by Hippocrates1,2. Since the first temperature measurements by Galileo, methods for temperature measurements have expanded beyond the classical thermometer3. During the development of magnetic resonance imaging (MRI), it was found that the resonant frequency of protons in many molecules depends on the temperature of their environment, enabling non-invasive temperature measurements in living organisms4,5,6,7. MR thermometry has both diagnostic and therapeutic applications and has been used clinically7,8,9,10,11,12,13,14,15. Most prominently, the temperature-dependence of the water proton resonance frequency (-0.01 ppm/°C)5,7,11,16,17 has been used to measure relative temperature in vivo without use of contrast agents. Other MR thermometry methods make use of 19F, hyperpolarized 129Xe, T1, T2, diffusion, magnetization transfer or contrast agents8,18,19,20.
Hyperpolarized (HP) 13C MRI has evolved as a valuable method21 to non-invasively measure metabolic activity22,23, pH24,25,26, and perfusion24,27. Most commonly, a solution containing HP [1-13C]pyruvate (PA) is injected intravenously. The hyperpolarized 13C label is exchanged with the larger endogenous lactate pool in the reaction catalysed by lactate dehydrogenase (LDH) enabling measurements of metabolic flux28. Since tumors exhibit an increase of glycolytic activity, together with elevated lactate production in the presence of oxygen (Warburg effect29), there is an increasing research interest in studying cancer using HP PA30,31,32,33,34. Hyperpolarized 13C MRI using dissolution dynamic nuclear polarization (d-DNP) has been translated from a preclinical setting towards patients in over 60 studies33,35. Lately, other hyperpolarization methods such as parahydrogen-induced polarization (PHIP)36,37,38,39 and signal amplification by reversible exchange (SABRE)40,41 have gained interest and are currently employed in preclinical settings.
In this work, the temperature and concentration dependent chemical shift of the [1-13C]lactate (LA) and [1-13C]pyruvate resonant frequencies are investigated. We characterize the dependencies at two magnetic fields (7 T and 11.7 T), temperatures ranging from 18 to 42 °C, metabolite concentrations of 5–600 mM and in aqueous solutions and whole blood. To investigate the translatability from a controlled phantom setting to living organisms, a preclinical study in healthy mice was conducted. Each animal was injected twice within one experimental session using HP PA, first at an elevated body temperature42 of 38 °C and then at a lowered temperature (31 or 34 °C). Additionally, we retrospectively analyze data from three published human brain HP [1-13C]pyruvate magnetic resonance spectroscopic imaging (MRSI) studies using our novel thermometry method, which investigated healthy brain metabolism, kidney metabolism and glioblastoma (GBM)30,43,44,45.
Results
Thermally polarized calibration measurements
The temperature-dependent chemical shift changes of [1-13C]pyruvate and [1-13C]lactate with reference to [13C]urea (UR) were characterized in vitro in aqueous solution at five temperatures at 11.7 T ranging from 20 to 42 °C, resulting in the 13C spectra shown in Fig. 1a–c. Signals from lactate, pyruvate, urea and two peaks from parapyruvate (PP) were detected. The chemical shift difference between lactate and pyruvate, at a concentration of 5 mM, decreases by 0.013 ± 0.002 ppm/°C (R2 = 0.99, see Table 1) with increasing temperature. The two PP peaks shift in the same direction as pyruvate and lactate, respectively. Spectra shown in Fig. 1 have been referenced to urea. During initial experiments, a concentration-dependent chemical shift change between pyruvate and lactate was also observed, which is characterized in Fig. 1b–d for eight concentrations ranging from 5 to 600 mM (pH = 7.73 ± 0.35). At higher concentrations, lactate and pyruvate shift away from each other. However, at the expected low in vivo concentrations of the metabolites (below 10 mM), this effect is negligible compared to the magnitude of temperature dependent shifts.
a Spectra of aqueous solutions containing the 13C-labeled molecules [1-13C]pyruvate (PA), [1-13C]lactate (LA) and [13C]urea (UR) at 50 mM concentration each. Measurements at different 5 temperatures show their temperature dependent chemical shifts, with respect to a reference ([13C]urea, UR). During the sample preparation process, [1-13C]parapyruvate (PP) was also formed, due to the acidic environment of the original solution before titration61 (pH 2.3). b Eight different solutions, ranging from 5 mM to 600 mM 13C at 34.85 °C, highlighting the concentration-dependent chemical shifts of pyruvate, lactate and parapyruvate. Urea and pyruvate spectra subplots (a, b) are multiplied by a factor of 4 for better visibility. c Relative chemical shift between pyruvate and lactate for 5 different temperatures, fitted to a linear regression model. d Relative chemical shift between pyruvate and lactate for 8 different concentrations, fitted to a power function, as described in Table S4 and equation S2. Temperatures were set in Kelvin on the spectrometer and converted to Celsius for comparability with in vivo data.
Hyperpolarized in vitro spectroscopy
HP experiments at 7 T using LDH phantoms are compared to thermal measurements in Fig. 2 (pH = 8.5 ± 0.2) and HP blood results are plotted in red (pH = 7.0 ± 0.2). Linear calibration fits to phantom data show similar trends, with deviations in blood and LDH experiments due to worse temperature control and concentration effects. Linear fit parameters from calibrations plotted in Fig. 2 are shown in Table 1, with all measurements shown in the Table S3. Using the linear dependency relationship, absolute temperatures can be calculated from chemical shift differences between lactate and pyruvate for given concentrations. Similarly, the concentration for a fixed temperature can be computed from the concentration fit functions given in Table S4 derived from fits to equation S2.
Blood data points (red) show frequency difference of co-polarized [1-13C]pyruvate and [1-13C]lactate mixed with rodent blood that was preheated to five different temperatures (20–36 °C). For LDH data points, HP [1-13C]pyruvate was injected into a buffered solution containing LDH and nicotinamide adenine dinucleotide (NADH) to facilitate lactate production at nine different temperatures ranging from 18–40 °C. Both HP experiments were conducted at in vivo concentrations of 8 mM pyruvate and a linear regression model (not shown) was fit to both datasets. Line plots (5 mM and 100 mM from 11.7 T experiments) show linear regression models to data presented in Fig. 1d for two exemplary concentrations to demonstrate similarity between measurements at 7 T and 11.7 T and different media. Corresponding fit parameters and uncertainties are displayed in Table 1, confidence intervals for 5 mM and 100 mM linear regression fits are too small to depict.
Healthy mouse abdomen apparent temperature mapping using HP pyruvate
In vivo 2D free induction decay chemical shift imaging sequence (FID-CSI) from an exemplary healthy animal (cohort 1) shows a clear difference of apparent temperature for mean slice values (single slice placed over the kidneys in axial orientation (see Fig. 3a–c for anatomical reference, pyruvate and lactate intensity maps), mean over all voxels with a fit error below 0.5 °C within the slice) between the first injection at 38.2 °C rectal temperature and the second injection at 34.5 °C (see apparent temperature maps in Fig. 3d/e, exemplary spectrum in Fig. 3f and corresponding histograms in Fig. 3g). While the absolute values of abdominal apparent temperature from CSI (TCSI) are higher than rectal temperature (Trec) as measured by a PT-100 probe, the relative changes in apparent temperature for these two measurements are the same within error bounds (TCSI = 40.0 ± 1.7 °C and 36.1 ± 1.8 °C vs. Trec = 38.2 ± 0.1 °C and 34.5 ± 0.1 °C, (i.e.,: ΔTCSI = 3.9 °C, vs. ΔTrec = 3.7 °C).
T2w anatomical reference image (a) and both pyruvate (b) and lactate (c) peak area integral maps from the CSI at a decreased rectal temperature. d 2x-interpolated apparent temperature map (average temperature: 36.1 ± 1.8 °C, native resolution shown as white square) calculated from CSI data for the injection at a decreased rectal temperature. e Elevated rectal temperature map (average temperature: 40.0 ± 1.7 °C, kidney ROI shown as white dotted line). f Exemplary spectrum (black) and fit of a kidney voxel (red: 38.2 °C, blue: 34.5 °C, white squares in d/e) for both elevated and decreased rectal temperature injections, with pyruvate and lactate peak areas as insets. g Histogram of apparent temperatures for both datasets shown in (d, e).
Comparing data from all eight animals, differences between the injections at elevated and decreased rectal temperature measurements become apparent (Fig. 4 and Table 2).
a, c Average slice apparent temperature for the injection at an elevated rectal temperature (38.0 ± 0.3 °C, n = 8) as well as the following injection at decreased rectal temperature (34.5 ± 0.1 °C, 31.6 ± 0.1 °C, n = 4 each). Error bars are one standard deviation obtained from the fit-error thresholded temperature map, as shown in Fig. 3d–f. Apparent temperatures calculated from CSI (TCSI) are on average higher than rectal temperature (Trec). b, d Average kidney apparent temperatures obtained from segmentation in the anatomical image and application of that mask to the temperature map plotted against rectal temperature. The bisectrix is drawn for clarification.
For the elevated temperature measurements, TCSI agrees with Trec for 4 of 8 animals within error bounds (Fig. 4a). For the decreased rectal temperature scans, the two cohorts need to be distinguished (cohort 1 at Trec = 34.4 °C and cohort 2 at Trec = 31.6 °C). For the former, the abdominal apparent temperature from CSI is within the error bounds of the rectal temperature in 2 out of 4 animals (Fig. 4c). For the latter, TCSI appears to be increased with respect to Trec in 3 out of 4 animals (Fig. 4c). Relative changes in apparent temperature vs. rectal temperature are similar (cohort 1: ΔTCSI = 3.8 ± 1.7 °C vs. ΔTrec = 3.7 ± 0.3 °C, p = 0.88, cohort 2: ΔTCSI = 5.4 ± 2.2 °C vs. ΔTrec = 6.3 ± 0.3 °C, p = 0.40). To assess intra-slice variability, kidneys were segmented and mean apparent temperatures of both organs compared to respective rectal temperatures (Fig. 4b, d). For measurements at elevated rectal temperature and for mice in cohort 1 imaged at a rectal temperature of 34.4 °C, kidney apparent temperature agrees with rectal temperature within error margins (Fig. 4b, d). Similarly to whole abdominal apparent temperature (Fig. 4c), animals in cohort 2, imaged at 31.6 °C, appear to show an increased kidney apparent temperature compared to their rectal temperatures (Fig. 4d).
Differences in metabolism at lower temperatures
Dynamic slice selective spectroscopy (Fig. 5, slice placement over kidneys as seen in Fig. 5a, b) showed a 44 ± 1% and 46 ± 1% decrease in lactate production at lower rectal temperatures (area under the curve ratio (AUCR) = 0.94/0.52 at 37.8 °C/34.3 °C and AUCR = 0.74/0.40 at 37.8 °C/31.5 °C, Fig. 5b, e). Absolute apparent temperatures computed are slightly higher than the rectal measurements and also vary over time (TC13,high1 = 39.0 ± 1.3 °C and TC13,high2 = 36.4 ± 1.4 °C vs. Trec,high1 = 37.8 ± 0.1 °C, TC13,low1 = 34.2 ± 0.9 °C vs Trec,low1 = 34.3 ± 0.1 °C, TC13,low2 = 32.0 ± 2.2 °C vs Trec,low2 = 31.5 ± 0.1 °C, Fig. 5c, f). This effect is especially pronounced in the animal injected at 31.5 °C (Fig. 5f), where the apparent temperature increases after injection. For the warmer injection, the apparent temperature approaches the rectal temperature after 15 s, while, for the colder injection, it continues increasing, consistent with the core temperature being maintained higher than the directly measured rectal temperature. The difference in AUCR for hot and cold measurements is consistent with reduced activity of LDH at lower temperatures46, which results in lower lactate production. Other contributing effects to this difference, such as perfusion changes or temperature of the injected bolus, cannot be excluded but are difficult to isolate as contributing factors. Similar results (decrease by 55% and 75% in AUCR) were acquired for two animals imaged using a 3D-bSSFP sequence at elevated and decreased rectal temperatures (Figure S1).
a, d T2w anatomical reference images with slice placement over kidneys (yellow lines). b, e Time domain fit FID magnitude amplitude curves for pyruvate (solid) and lactate (dashed) for two successive injections at elevated and decreased rectal temperatures in the same animals. AUCR decreases with decreased rectal temperature (Animal 1: 37.8/34.3 °C: AUCR = 0.94/0.52, Animal 2: 37.9/31.5 °C: AUCR = 0.74/0.40, while mean apparent temperature calculated from spectra is slightly higher than the rectal temperature (TC13,high1 = 39.0 ± 1.3 °C, TC13,low1 = 34.2 ± 0.9 °C, TC13,high2 = 36.4 ± 1.4 °C, TC13,low2 = 32.0 ± 2.2 °C). c, f MR-measured mid-abdominal / kidney apparent temperature drifts over time after injection for repetitions with high SNR in both pyruvate and lactate channels as well as the effectively stable recorded rectal temperature. g Exemplary spectra and fits for both injections in (e) at approximate time of lactate peak with pyruvate and lactate peak areas as inserts.
Healthy mouse brain apparent temperature
Whole brain apparent temperature from two animals scanned twice each correlates well with monitored rectal temperature (TCSI = 38.2 ± 2.4 °C and Trec = 37.8 ± 0.3 °C, n = 4). Brain apparent temperature maps are shown in Figure S9 and S10.
Apparent temperature from human 13C MRS in the brain and abdomen
Using calibration functions for 5 mM metabolite concentration, brain apparent temperature maps from nine healthy volunteers (data published by Kaggie et al.43) were generated, with an example given in Figure S3 (apparent temperature maps and mean slice apparent temperatures from all volunteers are found in Figures S4 and S6 and Table S5). Pyruvate signal per patient as a perfusion-weighted quantity was plotted against apparent temperature resulting in no systematic correlation (Figure S5). Similarly, mean apparent brain temperature values were generated from 7 injections of hyperpolarized [1-13C]pyruvate into four healthy volunteers (data published by Ma et al.44, all values shown in Figure S6 and Table S7).
Mean apparent temperature values computed from slice selective 13C spectroscopy in GBM patients (data published by Zaccagna et al.30, all values shown in Figure S6 and Table S6) are higher than those of healthy volunteers (Healthy 143: T = 31.7 ± 2.5 °C (n = 9), Healthy 244: T = 31.7 ± 1.7 °C (n = 7) vs. GBM30: T = 35.1 ± 2.1 °C (n = 6), Fig. 6b). For glioblastoma, an increase in metabolic activity in the whole brain has been reported30, which we hypothesized could also lead to a localized increase in temperature. However, no correlation between AUCR and brain temperature was found (Fig. 6a, R2 = 0.01, slope = 3.8 ± 7.5 °C-1).
a Slice-selective spectroscopy data from patients with GBM tumors in three slices each (total 18 slices with fit accuracy of 3 °C or better as well as SNR above 5). No correlation between AUCR and apparent temperature could be estimated (R2 = 0.01, slope = 3.8 ± 7.5 °C-1), nor any difference in apparent temperature between slices containing a tumor and tumor-free slices (p > 0.6). b Comparing apparent temperatures from healthy human brain CSI datasets (Healthy 143: T = 31.7 ± 2.5 °C, n = 9), healthy brain slice selective spectroscopy (Healthy 244: T = 31.7 ± 1.9 °C, n = 7) and GBM slice selective spectroscopy (GBM30: T = 35.1 ± 2.1 °C, n = 6), shows an increase in temperature for tumor patients (p < 0.01). Healthy human kidney apparent temperature is below physiological values (T = 33.3 ± 1.6 °C, n = 11) and, when compared to GBM apparent temperature, a significant difference is found (p < 0.05).
Finally, human abdominal slice-selective MRS data from 12 volunteers scanned a total of 14 times was analyzed (Fig. 6b, all data shown in Figure S6 and Table S8). Slice mean values are below physiological values but slightly higher than brain apparent temperature (Renal45: 33.3 ± 1.6 °C (n = 10)).
Discussion
For the first time, the temperature dependent chemical shifts of [1-13C]pyruvate and [1-13C]lactate have been reported and used for in vivo temperature sensing in addition to acquiring metabolic information. The relative chemical shift difference between pyruvate and lactate at 18–42 °C was characterized in aqueous and blood solutions and at concentrations ranging from 5 mM to 600 mM of 13C-labeled molecules. We found a linear temperature-dependent chemical shift of 0.013 ppm/°C, where pyruvate and lactate peaks diverge at lower temperatures.
We assessed the dependency of the chemical shifts of [1-13C]pyruvate and [1-13C]lactate on pH as well using published data47,48, and found that lactate shifts in the relevant in vivo pH range by maximally 0.006 ppm, while pyruvate does not exhibit a relevant shift in this range (see Supplemental Information, Figure S8). Therefore, the pH related uncertainty is on the order of ±0.5 °C. Dependence of blood oxygenation on [1-13C]pyruvate and [1-13C]lactate chemical shifts was assessed to be of minor concern (Supplemental Information, Figure S7).
Mouse abdominal CSI-derived apparent temperature was higher compared to simultaneously acquired rectal probe temperatures (Fig. 4), which is unlikely to be only due to a variation between core and rectal temperature49. This difference was greater at lower rectal temperatures, hinting towards internal biological temperature drop compensation mechanisms (e.g., brown fat activation50 and modulation of peripheral vs. core blood flow when body temperature is reduced51). Variations in apparent temperature due to differences in concentration of 13C-labeled metabolites were determined to be negligible in the range from 5–50 mM (see Fig. 1c, d). For variations on the order of 5–50 mM, this would lead to a maximum deviation of around 4 °C (assuming a chemical shift difference of 12.2 ppm, the apparent respective temperatures are: Tapparent, 5mM = 37.6 °C and Tapparent, 50mM = 41.5 °C). Importantly, when assuming that the amount of injected pyruvate does not exceed 10% of total blood volume of the animal, concentrations of the metabolite in the bloodstream are expected to be below 8 mM, since pyruvate from the d-DNP device has a concentration of 80 mM. Concentration-dependent differences for animal data should therefore be negligible.
Possible explanations for the apparent temperature drifts (Fig. 5 cf) could be bolus related concentration effects or variations in injected pyruvate temperature, which was not recorded.
Apparent temperature computed from human 13C CSI datasets is consistently lower than physiologically reasonable (Fig. 6 and S6) and shows a large variation across the brain (31.6 ± 5.4 °C). Multiple reasons could be the cause for this deviation. Firstly, variations in susceptibility due to tissue microstructure lead to local brain B0 changes52, which would affect pyruvate and lactate differently, since they are expected to primarily reside in different microstructural compartments with their relative contributions changing dynamically over time. While HP pyruvate is injected intravenously, HP lactate is initially produced intracellularly and is then transported outside of the cell53,54,55. The preclinical data is consistent with this explanation, for which also a large variation across mouse brain apparent temperature maps is observed (37.7 ± 2.8 °C).
Secondly, the uncertainty of the frequency, and therefore temperature, measurement is limited by the accuracy of the spectral fit, which is directly linked to the spectral resolution and noise level. For the four human MRS datasets analyzed in this work, nominal spectral resolution varied between 2 and 20 Hz/pt. To minimize errors from fitting, iterative time-domain fitting algorithms were developed and benchmarked against artificial spectra during this study (see Figure S11) and have been made available for public use on GitHub (https://github.com/Schilling-Lab/pyr-lac-thermometry-study.git). At 3 T, 1 ppm corresponds to 32 Hz for 13C nuclei, which then results in a resolution of 48 °C/pt for a spectrum acquired with 20 Hz/pt spectral resolution. Therefore, the fitting algorithm developed in this work is able to achieve sub-resolution accuracy for the resonance frequencies of pyruvate and lactate (see Figure S11d). Lower SNR also leads to a lower accuracy of the fitted temperature, as do worse shims, which was confirmed by simulations (Figure S11b, c). In addition, line broadening of spectra before fitting leads to an error in the fitted frequency and therefore a temperature bias (Figure S11a). Despite those limitations, low SNR spectra acquired at 7 T of thermally polarized pyruvate and lactate in blood at constant temperature (18 °C) and concentration (220 mM) were analyzed and the fitted temperature agrees well with temperature from a Pt100 probe (Figure S11e, f). Thirdly, the concentration of [1-13C]pyruvate in clinical d-DNP systems is higher than in the preclinical system used in this work (80 mM vs. 250 mM), which could lead to bolus effects in humans in which concentration-dependent chemical shift difference changes could occur due to slower blood circulation and higher bolus concentration. We tried to investigate this by correlating pyruvate intensity and apparent temperature for healthy human brain data (Figure S5), but found no correlation. Future studies should address this question and investigate the concentration of hyperpolarized pyruvate within a bolus and subsequently in the target tissue explicitly.
Finally, an effect changing the apparent temperature could be due to the fact that previous work has shown that extracellular lactate has a higher frequency than intracellular lactate56 by 0.02 ppm, which corresponds to a temperature difference of 1.8 °C (intracellular lactate signal appearing warmer).
For these reasons, this method has limited application for determining accurate absolute temperatures in humans, while it warrants further investigation as an imaging biomarker especially since apparent temperature can be easily retrospectively obtained from hyperpolarized [1-13C]pyruvate MRSI datasets. Nevertheless, for spectroscopically-estimated brain temperature from 13C MRSI between healthy human brains and patients with GBM, an increased mean brain apparent temperature was found in the GBM case (Fig. 6b). This could also be due to the fact that all patients with GBM were taking high doses of dexamethasone at the time of scanning, which could mildly increase the body temperature57, or due to differences in brain microstructure for GBM patients.
Besides an application as a thermometry method, the effects characterized in this work have implications for 13C imaging sequences that assume the resonance frequencies to be stable during a scan. For example, in a balanced SSFP sequence as described by Skinner et al.58, the relative chemical shift between pyruvate and lactate was assumed to be constant. However, a shift by a few Hz, which could occur for a change in local tissue temperature of a few °C, could lead to a metabolite moving along the response curve, therefore skewing the resulting AUCRs or leading to signal bleeding from one metabolite channel to the other. To avoid this, body and even local organ or tissue temperature needs to be taken into account when using 13C MRI sequences that are based on the assumption of constant frequencies of pyruvate and lactate. Additionally, in the setup of sequences for 13C MRI, a phantom containing thermally polarized compounds, such as 13C lactate, may be used for reference frequency and transmit B1 calibrations. If this phantom is placed on the outside of the animal or patient, its temperature will likely be substantially lower than body temperature, and therefore its resonance frequency will also be offset from that in the subject59.
To summarize, our method enables temperature or concentration measurements of NMR spectra containing [1-13C]pyruvate and [1-13C]lactate. Given that both of the metabolites exhibit a temperature and concentration dependent chemical shift, they can also be used independently if another reference molecule, such as [13C]urea, is present at the same location in vivo and calibration functions for that molecule have been established, which was outside the scope of this work. Especially for quality control of novel hyperpolarization devices and translation into an automated workflow for clinical use or in vitro studies, this technique would be highly beneficial, since only one NMR spectrum is needed to measure temperature or concentration, assuming the other is known independently. Furthermore, since the temperature information is gained essentially for free with any hyperpolarized pyruvate injection and following spectroscopic sequence, one could investigate changes in apparent temperature due to tumor treatment via chemo- or radiotherapy, with the mentioned limitations to studies.
Methods
MR hardware
Preclinical measurements and calibrations were performed on a 7 T preclinical MRI scanner (Discovery MR901 magnet and gradient system, Agilent, Santa Clara, CA, USA; AVANCE III HD electronics, Bruker, Billerica, MA, USA). For mouse abdominal imaging, a 31 mm 1H/13C dual-tuned volume resonator (RAPID Biomedical, Rimpar, Germany) was used. Mouse brain imaging was conducted using a cryogenically cooled (temperatures: coil 30 K, preamplifier 77 K) linear receive-transmit 20 mm surface coil (Bruker). Phantom measurements at 7 T were carried out in a 10 mm 13C solenoid coil (RAPID Biomedical). In vitro calibration measurements at 11.7 T were done on a 500 MHz NMR spectrometer (Bruker). Human brain MRI and MRS was performed on 3 T MR systems (MR750 or MR750w, GE Healthcare, Waukesha WI, USA), using a dual-tuned 1H/13C quadrature head coil (RAPID Biomedical), as described in the original work30,43, a custom built dual-frequency 1H/13C quadrature head coil (Clinical MR Solutions, LLC, Brookfield WI, USA)44 or a 8-channel 13C receive coil (Rapid Biomedical) paired with a 13C clamshell coil for transmission45.
Hyperpolarization
For preclinical abdominal and in vitro MRI, 26–35 mg 14 M [1-13C]pyruvate (Merck, Darmstadt, Germany) or 50–85.7 mg 3 M [1-13C]lactate (Merck, Darmstadt, Germany) were polarized together60 with 15 mM OX063 trityl radical (GE Healthcare) and 1 mM gadoteric acid (Dotarem, Guerbet, France) in a commercial d-DNP polarizer (Hypersense, Oxford Instruments, Oxford, United Kingdom) at 3.35 T / 1.2 K / 94.139 GHz for more than 45 min. The compound was then rapidly dissolved in 2.9–5.0 ml hot buffered saline solution (80 mM Tris, 80 mM NaOH, 0.1 g/l Na2-EDTA) to neutralize pH and reach an average concentration of 80 mM for the 13C compound.
For mouse brain MRI, a PHIP-sidearm hydrogenation (SAH)-specific precursor of [1-13C]pyruvate was polarized and dissolved in 0.8 ml buffered aqueous solution (16 mM phosphate buffer) using an alpha device of the preclinical polarizer POLARIS (NVision, Germany), yielding a final concentration of 75 mM.
For human MRI, [1-13C]pyruvate was polarized in a clinical d-DNP polarizer (SPINlab, 5 T, GE Healthcare), as described in the original work30,43,44,45.
Temperature reference measurements
Animal and phantom temperature was monitored and recorded using a rectal MR-compatible Pt100 probe (SA Instruments, Stony Brook, NY, USA) with a resolution of ±0.1 °C. For 11.7 T NMR measurements, temperature was set through the NMR software and kept for 5 min to equilibrate before starting of the data acquisition.
In vitro MRS
Thermal in vitro measurements at 11.7 T used a 13C-decoupled sequence (TR = 90 s, FA = 90°, spectral resolution = 1 Hz/pt, averages = 16 (5–50 mM) or 8 (100–600 mM)). Hyperpolarized in vitro calibration measurements at 7 T used a non-selective excitation sequence (FA = 10°, TR = 1.5 s, spectral resolution = 0.5 Hz/pt).
Thermal phantoms
An aqueous solution containing 13C-labeled molecules at a concentration of 600 mM each was prepared by dissolving 62.9 mg of good manufacturing practice (GMP) grade [13C]urea in 1200 µl H2O. To this solution, 226.7 mg 50% w/w [1-13C]Na-lactate and 73 µl 14 M [1-13C]pyruvic acid were added (pH 2.3). Due to the low pH, some of the pyruvic acid was converted into parapyruvate, which is also visible in the spectra61. Then, the pH was adjusted to 7.88 by titration with NaOH or HCl solution (final volume of 1625 µl). From this solution, seven samples of decreasing concentrations of 13C molecules (400, 200, 100, 50, 20, 10, 5 mM) were obtained through serial dilution with pure water (pH = 7.73 ± 0.35, exact values for concentrations and pH given in Table S1). 600 µl of 13C-labeled solution was mixed with 60 µl of D2O for spectrometer shimming and lock adjustments and filled into 5 mm NMR tubes.
Samples for experiments with varying blood oxygenation were prepared by mixing 38.9 mg [1-13C]Na-pyruvate powder (Merck, KGaA, Darmstadt, Germany), 33 µl 9 M [13C]urea solution and 76 µl 13C 55% w/w Na-lactate solution with 1.5 ml of human blood (drawn into lithium-heparin syringes to prevent blood clotting and be compatible with blood gas analyzers) for a final concentration of approximately 200 mM for each of the metabolites (see Supplemental Information section 4). [1-13C]Na-pyruvate was chosen to avoid strongly acidic conditions, which would lead to blood clotting and formation of parapyruvate61.
Hyperpolarized phantoms
For hyperpolarized calibration experiments at different temperatures 152 µl HP [1-13C]pyruvate was injected into a 10 mm glass NMR tube, preheated in a water bath and containing 1572 µl fresh PBS buffer (pH 7.35) + 97 µl nicotinamide adenine dinucleotide hydrate (NADH, 250 mM) + 3-9 µl lactate dehydrogenase (LDH) solution (rabbit muscle, Roche, Basel, Switzerland), to reflect in vivo conditions (8 mM, pH = 8.58 ± 0.20, n = 8). The amount of LDH was increased at lower temperatures to generate sufficient lactate signal.
To mimic in vivo conditions, 136 µl HP [1-13C]pyruvate and [1-13C]lactate (in TRIS buffer as described in hyperpolarization section) were injected into a preheated 10 mm glass NMR tube filled with 1.5 ± 0.2 ml of rodent blood, which was prepared in 2 ml blood collection tubes containing ethylenediaminetetraacetic acid (EDTA) to prevent clotting (pH = 7.0 ± 0.2, n = 5). Rodent blood gas levels were not controlled.
For both hyperpolarized experiments, NMR acquisitions were started at time of dissolution. Bore temperature was kept stable during the acquisition using a PET Dryer Model B-8 (XPower, City of Industry, CA, USA). See Table S2 for pH and temperature values. After pipetting the HP solutions into the NMR tube close to the bore, a PT-100 temperature sensor was inserted and continuously recorded sample temperature during the acquisition.
In vivo animal MRSI
12 healthy mice (C57BL/6 mixed with 129S background, age = 144 ± 8 days, weight = 31 ± 4 g, 7 female, 5 male) were anesthetized with 2% isoflurane by volume in 2 l/min 100% O2 and manually injected twice with 300 μl 80 mM HP PA via a tail vein catheter. For a previously reported total blood volume for mice of 72 ml/kg62, this leads to an upper limit of the pyruvate concentration on the order of 10 mM (10.8 ± 1.4 mM, n = 10), which is in line with previously reported values63. Apparent temperatures estimated from 13C spectra were computed using the linear calibration function for the relative frequency difference of [1-13C]pyruvate and [1-13C]lactate at 5 mM concentration, which was chosen by assuming that metabolites occur in a lower concentration in perfused organs compared to blood. The animals were divided into two cohorts differing with respect to their rectal temperature (Trec) at the second injection. First, animals were kept at around Trec = 38 °C, while for the second injection the temperature was decreased by turning down the bore heating over the course of 1 h, resulting in a Trec = 34–35 °C (first cohort, n = 6) or 31–32 °C (second cohort, n = 6) before being sacrificed by cervical dislocation (under anesthesia 5% isoflurane, 2 l/min 100% O2). All experiments were conducted in accordance with a valid animal license (ROB 55.2-2532.Vet_02-18-91, ROB-55.2-2532.Vet_02-23-70).
Eight animals (four per cohort) underwent a single-slice static 2D free induction decay chemical shift imaging sequence (FID-CSI) placed over the kidneys, starting 17 s post start of the injection, as this is where lactate signal peaking in the kidneys was expected (TR/TE = 235/1 ms, acquisition matrix = 14 × 12, voxel size = 2.2 × 2.2 × 6 mm3, FA = 12°, spectral bandwidth = 40 ppm, spectral resolution = 2.2 Hz/pt, acquisition points = 700). One animal per cohort received a dynamic broad excitation bandwidth slice spectroscopy scan to determine differences in metabolism at varying body temperature and time dependent temperature following an injection with HP PA (FA = 10°, TR = 2 s, spectral resolution = 0.55 Hz/pt, acquisition points = 2048). One animal per cohort was scanned using a spectrally-selective 3D balanced steady-state free precession (3D-bSSFP) sequence39,58, see Supplemental Information Figures S1, S2.
Additionally, two healthy (C57BL/6, age = 142 ± 4 days, weight = 19.9 ± 0.1 g, both female) mice were imaged using a static 2D FID-CSI (center-out k-space sampling, TR/TE = (88.5, 88.5, 100.4, 174.2/ 1.3) ms, acquisition matrix = (40 × 18, 33 × 18, 35 × 18, 30 × 24) voxel size = (1 x 1 x 3, 1 x 1 x 3, 1 x 1 x 3, 1.33 × 1.33 × 3) mm3, FA = (6, 5, 7, 7)°, spectral bandwidth = (40, 40, 70.5, 40) ppm, spectral resolution = (11.8, 11.8, 10.3, 5.9) Hz/pt, acquisition points = 256, 256, 512, 512). CSI slices were positioned horizontally through the mouse brain.
In vivo human MRSI
Data from four published studies on healthy human brain43,44, glioblastoma (GBM)30 and kidney metabolism45 was re-analyzed and temperature values computed using calibration functions for 5 mM 13C metabolites, since an upper limit of the pyruvate concentration in blood could be computed to be on the order of 2 mM (total blood volume of healthy adult humans computed using the Nadler and Lemmens-Bernstein-Brodsky equations64) and 5 mM was the lowest calibration function obtained in in vitro characterization experiments. Additionally, at low concentrations the chemical shift difference of PA and LA related to concentration is negligible compared to the shift due to temperature. GBM and renal patients underwent a dynamic broad excitation bandwidth slice spectroscopy scan interleaved with an IDEAL spiral CSI acquisition (the latter being not used for this study)65 after intravenous injection with 0.4 ml/kg of 250 mM hyperpolarized [1-13C]pyruvate (GBM: TR = 500 ms, FA = 15°, slices = 3, slice thickness = 30 mm, receive bandwidth = 62.5 kHz, acquisition points = 4095, spectral resolution = 15 Hz/point, renal: TR = 500 ms, FA = 15°, slices = 5, slice thickness = 3 cm, Rx bandwidth = 62.5 kHz, acquisition points = 3008, spectral resolution = 20 Hz/point), while healthy volunteers in study 143 were scanned using a static 2D chemical shift imaging sequence (slices = 5, slice thickness = 20 mm, Rx bandwidth = 5 kHz, acquisition points = 256, spectral resolution = 20 Hz/point, matrix size = 10 × 10 with corners of k-space omitted66, voxel size = 2×2 cm2,) and in study 244 using a slice spectroscopy sequence (TR = 3 s, FA = 6.5°, slices = 4, slice thickness = 15 mm, Rx bandwidth = 5 kHz, acquisition points = 2048, spectral resolution = 2.4 Hz/point). Ethical approvals are found in the respective publications.
Data analysis
The data was analyzed using Python 3.10. The spectroscopic data was fit in time-domain by iteratively minimizing the sum-of-squares of the difference of the measured data and the modeled free induction decay (FID) data. ChatGPT (OpenAI, San Francisco, CA, USA) was used in assisting with writing analysis code.
Linear fits to in vitro data provided calibration functions for lactate-pyruvate frequency difference to temperature. Calibration curves for 5 mM concentration were applied to compute temperature values from metabolite peak frequencies, which is referred to as apparent temperature in this work due to the other influences on the metric that are described here. For purposes of thresholding, single voxel or repetition temperature uncertainties were calculated from the metabolite peak frequency fit uncertainties via equation S1. For mouse kidney CSI data, resulting apparent temperature maps from fits were 2x interpolated and masked by thresholding pixels with a fit uncertainty below 0.5 °C and an ROI drawn around the animal’s abdomen on the axial T2w slice corresponding to the CSI slice. For mouse slice selective spectroscopy data, spectra with pyruvate and lactate peaks above noise level were selected and then spectra with a fit uncertainty above 2 °C excluded. This larger threshold was chosen to account for the difference in SNR between the CSI data and slice-selective spectroscopy data, with CSI spectra having higher SNR due to increased flip angle and signal averaging due to inherent repeated excitations. Mouse brain CSI data was similarly thresholded to pixels with a fit uncertainty below 0.5 °C. For human apparent temperature analysis, values from spectra with a fit uncertainty below 3 °C and peak SNR above 5 were included, due to the lower spectral resolution compared to the animal data. ROI masking was not performed for human brain CSI data.
Uncertainties of averaged subject apparent temperature values were calculated as one standard deviation from the mean of included voxel values or of repeated spectral acquisitions. Uncertainties of mean values for the four different human datasets (healthy brain CSI, healthy brain slice-spectroscopy, healthy kidney and GBM brain) shown in Figure S6 were computed by computing the error of the mean over all subjects in each study. Statistical analysis comparing rectal vs. abdominal temperature from 13C spectroscopy was performed using a paired t-test67.
It was also found that line broadening changes the resonance frequencies such that the apparent temperature decreases (Figure S11a). Due to this reason, only spectra without line broadening were used in this work.
Data availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
Code availability
The underlying code for this study in the form of Jupyter Notebooks and a custom Python package are available on GitHub (https://github.com/Schilling-Lab/pyr-lac-thermometry-study.git).
References
Grodzinsky, E. & Sund Levander, M. History of Body Temperature. in Understanding Fever and Body Temperature: A Cross-disciplinary Approach to Clinical Practice (eds. Grodzinsky, E. & Sund Levander, M.) 7–22 (Springer International Publishing, Cham, 2020). https://doi.org/10.1007/978-3-030-21886-7_2.
Otsuka, K. & Togawa, T. Hippocratic thermography. Physiol. Meas. 18, 227 (1997).
Ring, E. F. J. The historical development of temperature measurement in medicine. Infrared Phys. Technol. 49, 297–301 (2007).
Hindman, J. C. Proton Resonance Shift of Water in the Gas and Liquid States. J. Chem. Phys. 44, 4582–4592 (1966).
Poorter, J. D. et al. Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human Muscle. Magn. Reson. Med. 33, 74–81 (1995).
De Poorter, J. Noninvasive MRI thermometry with the proton resonance frequency method: study of susceptibility effects. Magn. Reson. Med. 34, 359–367 (1995).
Ishihara, Y. et al. A precise and fast temperature mapping using water proton chemical shift. Magn. Reson. Med. 34, 814–823 (1995).
Rieke, V. & Butts Pauly, K. MR thermometry. J. Magn. Reson. Imaging 27, 376–390 (2008).
de Senneville, B. D. et al. MR thermometry for monitoring tumor ablation. Eur. Radiol. 17, 2401–2410 (2007).
Roujol, S., Ries, M., Quesson, B., Moonen, C. & Denis de Senneville, B. Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn. Reson. Med. 63, 1080–1087 (2010).
Cady, E. B., D’Souza, P. C., Penrice, J. & Lorek, A. The Estimation of Local Brain Temperature by in Vivo 1H Magnetic Resonance Spectroscopy. Magn. Reson. Med. 33, 862–867 (1995).
Rzechorzek, N. M. et al. A daily temperature rhythm in the human brain predicts survival after brain injury. Brain 145, 2031–2048 (2022).
Childs, C., Vail, A., Protheroe, R., King, A. T. & Dark, P. M. Differences between brain and rectal temperatures during routine critical care of patients with severe traumatic brain injury*. Anaesthesia 60, 759–765 (2005).
Holbrook, A. B., Santos, J. M., Kaye, E., Rieke, V. & Pauly, K. B. Real-time MR thermometry for monitoring HIFU ablations of the liver. Magn. Reson. Med. 63, 365–373 (2010).
Thrippleton, M. J. et al. Reliability of MRSI brain temperature mapping at 1.5 and 3. T. NMR Biomed. 27, 183–190 (2014).
Ohnishi, M. & Urry, D. W. Temperature dependence of amide proton chemical shifts: The secondary structures of gramicidin S and valinomycin. Biochem. Biophys. Res. Commun. 36, 194–202 (1969).
Blackwell, J. et al. Proton Resonance Frequency Shift Thermometry: A Review of Modern Clinical Practices. J. Magn. Reson. Imaging 55, 389–403 (2022).
Bruns, C. et al. Organic Fluorine Compoundsand Their Uses as Molecular MR‐Based Temperature Sensors. Chem. Phys. Chem 24, e202300512 (2023).
Schilling, F. et al. MRI Thermometry Based on Encapsulated Hyperpolarized Xenon. ChemPhysChem 11, 3529–3533 (2010).
Odéen, H. & Parker, D. L. Magnetic resonance thermometry and its biological applications – Physical principles and practical considerations. Prog. Nucl. Magn. Reson. Spectrosc. 110, 34–61 (2019).
Ardenkjær-Larsen, J. H. et al. Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMR. Proc. Natl. Acad. Sci. 100, 10158–10163 (2003).
Golman, K. Metabolic Imaging by Hyperpolarized 13C Magnetic Resonance Imaging for In vivo Tumor Diagnosis | Cancer Research | American Association for Cancer Research. https://aacrjournals.org/cancerres/article/66/22/10855/526545/Metabolic-Imaging-by-Hyperpolarized-13C-Magnetic.
Lee, P. M. et al. Whole-Abdomen Metabolic Imaging of Healthy Volunteers Using Hyperpolarized [1- 13C]pyruvate MRI. J. Magn. Reson. Imaging 56, 1792–1806 (2022).
Grashei, M. et al. Simultaneous magnetic resonance imaging of pH, perfusion and renal filtration using hyperpolarized 13C-labelled Z-OMPD. Nat. Commun. 14, 5060 (2023).
Düwel, S. et al. Imaging of pH in vivo using hyperpolarized 13C-labelled zymonic acid. Nat. Commun. 8, 15126 (2017).
Gallagher, F. A. et al. Magnetic resonance imaging of pH in vivo using hyperpolarized 13C-labelled bicarbonate. Nature 453, 940–943 (2008).
Qin, H. et al. Clinical translation of hyperpolarized 13C pyruvate and urea MRI for simultaneous metabolic and perfusion imaging. Magn. Reson. Med. 87, 138–149 (2022).
Brindle, K. Watching tumours gasp and die with MRI: the promise of hyperpolarised 13C MR spectroscopic imaging. Br. J. Radiol. 85, 697–708 (2012).
Warburg, O. On the Origin of Cancer Cells. Science 123, 309–314 (1956).
Zaccagna, F. et al. Imaging Glioblastoma Metabolism by Using Hyperpolarized [1-13C]Pyruvate Demonstrates Heterogeneity in Lactate Labeling: A Proof of Principle Study. Radiol. Imaging Cancer 4, e210076 (2022).
Gallagher, F. A. et al. Imaging breast cancer using hyperpolarized carbon-13 MRI. Proc. Natl. Acad. Sci. 117, 2092–2098 (2020).
Tang, S. et al. Metabolic imaging with hyperpolarized 13C pyruvate magnetic resonance imaging in patients with renal tumors—Initial experience. Cancer 127, 2693–2704 (2021).
Nelson, S. J. et al. Metabolic Imaging of Patients with Prostate Cancer Using Hyperpolarized [1-13C]Pyruvate. Sci. Transl. Med. 5, 198ra108–198ra108 (2013).
Larson, P. E. Z. et al. Investigation of analysis methods for hyperpolarized 13C-pyruvate metabolic MRI in prostate cancer patients. NMR Biomed. 31, e3997 (2018).
Larson, P. E. Z. et al. Current methods for hyperpolarized [1-13C]pyruvate MRI human studies. Magn. Reson. Med. 91, 2204–2228 (2024).
Bowers, C. R. & Weitekamp, D. P. Parahydrogen and synthesis allow dramatically enhanced nuclear alignment. J. Am. Chem. Soc. 109, 5541–5542 (1987).
Hövener, J.-B. et al. Parahydrogen-Based Hyperpolarization for Biomedicine. Angew. Chem. Int. Ed. 57, 11140–11162 (2018).
Reineri, F., Boi, T. & Aime, S. ParaHydrogen Induced Polarization of 13C carboxylate resonance in acetate and pyruvate. Nat. Commun. 6, 5858 (2015).
Nagel, L. et al. Parahydrogen-Polarized [1-13C]Pyruvate for Reliable and Fast Preclinical Metabolic Magnetic Resonance Imaging (Adv. Sci. 30/2023). Adv. Sci. 10, 2370208 (2023).
Adams, R. W. et al. Reversible Interactions with para-Hydrogen Enhance NMR Sensitivity by Polarization Transfer. Science 323, 1708–1711 (2009).
de Maissin, H. et al. In Vivo Metabolic Imaging of [1-13C]Pyruvate-d3 Hyperpolarized By Reversible Exchange With Parahydrogen**. Angew. Chem. Int. Ed. 62, e202306654 (2023).
Reitman, M. L. Of mice and men – environmental temperature, body temperature, and treatment of obesity. FEBS Lett. 592, 2098–2107 (2018).
Kaggie, J. D. et al. Deuterium metabolic imaging and hyperpolarized 13C-MRI of the normal human brain at clinical field strength reveals differential cerebral metabolism. NeuroImage 257, 119284 (2022).
Ma, J. et al. Dynamic 13C MR spectroscopy as an alternative to imaging for assessing cerebral metabolism using hyperpolarized pyruvate in humans. Magn. Reson. Med. 87, 1136–1149 (2022).
Ursprung, S. et al. Hyperpolarized 13C-Pyruvate Metabolism as a Surrogate for Tumor Grade and Poor Outcome in Renal Cell Carcinoma—A Proof of Principle Study. Cancers 14, 335 (2022).
Cuddihee, R. W. & Fonda, M. L. A mathematical model describing the effect of temperature and substrate concentration on the activities of M4 and H4 lactate dehydrogenase from the big brown bat (Eptesicus fuscus). Arch. Biochem. Biophys. 212, 705–716 (1981).
Grashei, M., Hundshammer, C., van Heijster, F. H. A., Topping, G. J. & Schilling, F. pH Dependence of T2 for Hyperpolarizable 13C-Labelled Small Molecules Enables Spatially Resolved pH Measurement by Magnetic Resonance Imaging. Pharmaceuticals 14, 327 (2021).
Hundshammer, C., Grashei, M., Greiner, A., Glaser, S. J. & Schilling, F. pH Dependence of T for 13C-Labelled Small Molecules Commonly Used for Hyperpolarized Magnetic Resonance Imaging. ChemPhysChem 20, 798–802 (2019).
Gordon, C. J. Temperature Regulation in Laboratory Rodents. (Cambridge University Press, Cambridge, 1993). https://doi.org/10.1017/CBO9780511565595.
Zhang, L. et al. Absolute thermometry of human brown adipose tissue by magnetic resonance with laser polarized 129Xe. Commun. Med. 3, 147 (2023).
Gordon, C. J. Thermal physiology of laboratory mice: Defining thermoneutrality. J. Therm. Biol. 37, 654–685 (2012).
Liu, C. et al. Quantitative Susceptibility Mapping: Contrast Mechanisms and Clinical Applications. Tomography 1, 3–17 (2015).
Schilling, F. et al. Diffusion of hyperpolarized 13C-metabolites in tumor cell spheroids using real-time NMR spectroscopy. NMR Biomed. 26, 557–568 (2013).
Søgaard, L. V., Schilling, F., Janich, M. A., Menzel, M. I. & Ardenkjær-Larsen, J. H. In vivo measurement of apparent diffusion coefficients of hyperpolarized 13C-labeled metabolites. NMR Biomed. 27, 561–569 (2014).
Wespi, P., Steinhauser, J., Kwiatkowski, G. & Kozerke, S. Overestimation of cardiac lactate production caused by liver metabolism of hyperpolarized [1-13C]pyruvate. Magn. Reson. Med. 80, 1882–1890 (2018).
Breukels, V. et al. Direct dynamic measurement of intracellular and extracellular lactate in small-volume cell suspensions with 13C hyperpolarised NMR. NMR Biomed. 28, 1040–1048 (2015).
Tu, Q. et al. Perioperative administration of dexamethasone to prevent postoperative shivering: a systematic review and meta-analysis of randomized controlled trials. J. Int. Med. Res. 51, 3000605231187805 (2023).
Skinner, J. G. et al. Spectrally selective bSSFP using off-resonant RF excitations permits high spatiotemporal resolution 3D metabolic imaging of hyperpolarized [1- 13C]Pyruvate-to-[1- 13C] lactate conversion. Magn. Reson. Med. 90, 894–909 (2023).
Grist, J. T. et al. Creating a clinical platform for carbon-13 studies using the sodium-23 and proton resonances. Magn. Reson. Med. 84, 1817–1827 (2020).
Topping, G. J. et al. Hyperpolarized 13C Spectroscopy with Simple Slice-and-Frequency-Selective Excitation. Biomedicines 9, 121 (2021).
Margolis, S. A. & Coxon, B. Identification and quantitation of the impurities in sodium pyruvate. Anal. Chem. 58, 2504–2510 (1986).
Diehl, K.-H. et al. A good practice guide to the administration of substances and removal of blood, including routes and volumes. J. Appl. Toxicol. 21, 15–23 (2001).
Serrao, E. M. et al. Analysis of 13C and 14C labeling in pyruvate and lactate in tumor and blood of lymphoma-bearing mice injected with 13C- and 14C-labeled pyruvate. NMR Biomed. 31, e3901 (2018).
Sharma, R. & Sharma, S. Physiology, Blood Volume. in StatPearls. StatPearls Publishing, Treasure Island (FL), (2024).
Wiesinger, F. et al. IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1- 13C]pyruvate. Magn. Reson. Med. 68, 8–16 (2012).
Maudsley, A. A., Matson, G. B., Hugg, J. W. & Weiner, M. W. Reduced phase encoding in spectroscopic imaging. Magn. Reson. Med. 31, 645–651 (1994).
Virtanen, P. et al. SciPy 1.0: fundamental algorithms for scientific computing in Python. Nat. Methods 17, 261–272 (2020).
Acknowledgements
We acknowledge support from Alina Schreier, Hermine Mohr, Natalie Röder and Hussein Trabulssi in helping with animal blood extraction, Irina Skuratovska for animal breeding, Luisa Willner for helping with blood extraction and Lasse Relker and Christoph Schulz for assistance with blood gas analysis. This research was funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation, Emmy Noether program, grant number 391523415), the European Union’s Horizon 2020 research and innovation program under Grant Agreement No 820374 and German Federal Ministry of Education and Research (BMBF 13N16450/QuE-MRT).
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
W.G., L.N., J.G.S., M.Gr., G.J.T., S.S., N.S., W.E., I.H., and F.H.A.v.H. conducted and planned pre-clinical and in vitro experiments. Z.A., M.Gi., S.Ka., I.S., and S.Kn. provided and operated the PHIP polariser. W.G. and L.N. performed data analysis. M.A.M., J.M.P., and F.A.G. provided clinical data. W.G., L.N., M.Gr., G.J.T., and F.S. wrote the manuscript with input from all authors. W.G. and F.S. designed the study.
Corresponding author
Ethics declarations
Competing interests
The authors declare the following competing financial interests: Z.A., S.Kn, M.Gi., S.Ka., I.H., and I.S. are or were employed by NVision Imaging Technologies GmbH. F.S. serves on the scientific advisory board of NVision Imaging Technologies GmbH. F.A.G. has research support from GE Healthcare, NVision, and AstraZeneca.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gottwald, W., Nagel, L., Skinner, J.G. et al. Temperature dependent chemical shifts of pyruvate and lactate enable in vivo hyperpolarized 13C MRSI thermometry. npj Imaging 3, 19 (2025). https://doi.org/10.1038/s44303-025-00081-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s44303-025-00081-3