Table 1 Property comparison of representative nanomaterial modifications for bidirectional neural interfaces in the last 3 years
Material | Fabrication | Diameter (μm) | CIC/CIL (mC/cm2) | Cathodic CSC | Impedance | Biological experiment | Reference |
|---|---|---|---|---|---|---|---|
cGO/PEDOT:PSS | Electrodeposited | 30 | 3.11 ± 0.25 | 7.53 ± 0.34 mC/cm2 | 7.26 ± 0.29 kΩ (1 kHz) | In vitro | |
Nanostructured platinum | Electrodeposited | 35 | 3.0 ± 0.1 | 51.3 ± 0.2 nC | 66.71 ± 0.44 kΩ (100 Hz) | In vivo | |
PEDOT:PSS/ PtNPs | Electrodeposited | 30 | 4.37 ± 0.22 | 14.84 ± 2.72 mC/cm2 | 10.94 ± 0.49 kΩ; −12.54 ± 0.51° (1k Hz) | In vitro | |
PEDOT:CNF | Electrodeposited | 20 | 10.03 | 7.89 mC/cm2 | 1.28 MΩ/μm2 (1k Hz) | In vitro | |
PEDOT:PSS | Electrodeposited | 40 | / | 31.5 ± 1.3 mC /cm2 | 9.4 ± 0.9 kΩ (1k Hz) | In vitro and in vivo | |
Pt-Ir | Electrodeposited | 75 | 3.85 | 12.5 + 0.75 mC/cm2 | 50 ± 9 kΩ (1k Hz) | In vivo |