Table 1 Comparison of multimodal implantable Si- and fiber-based neural probes with light emission, electrophysiological recording, and microfluidic functionalities

From: Nanophotonic neural probes for in vivo photostimulation, electrophysiology, and microfluidic delivery

Device design

Implant cross-sectional area

No. of electrodes

No. of emitters

No. of microfluidic channels

Cross-sectional area of the channel

Ref.

Four-shank Si probe

128 μm × 40 μm × 4 shanks

32 (8/shank)

1 polymer waveguide

1 multi-cavity channel

10 μm × 12 μm for each cavity

13

Assembly of a glass fiber and a Si probe

150 μm × 50 μm + 225-μm diameter fiber

32

1 fiber core

1

11 μm × < 5 μm

24

Polymer fiber neural probe

≈160 μm diameter

4

8 fiber cores

1

≈80-μm diameter

19

Polymer fiber neural probe

306 μm × 342 μm

6

1 fiber core

1

100 μm × 40 μm

12

Si nanophotonic neural probe

107 μm × 100 μm

18

16 SiN grating coupler emitters

1

≈160 μm2

This work