Abstract
The implementation of Home Blood Pressure Monitoring (HBPM) has improved the diagnosis and treatment of hypertension. Current clinical guidelines advocate the use of validated HBPM devices. However, few studies have examined the current use of validated HBPM devices. We aimed to investigate the current status of the use of validated HBPM devices in real-world hypertensive patients. This study was conducted on Korean patients with hypertension using CareforMe®, a smartphone healthcare application which allows the patients to record their HBPM data which can be shared with their physicians. The validation status of HBPM devices was identified based on four registries: STRIDE BP, Medaval, the Japanese Society of Hypertension, and dabl® Educational Trust. From January 2022 through August 2024, 2731 patients entered the model numbers of their HBPM devices. 56.4% (n = 1539) of the population were male, and the mean age was 53.5 ± 10.8 years old. 110 models from 33 different manufacturers were identified. 97.0% (n = 2649) of the patients used upper-arm devices and 3.0% (n = 82) used wrist devices. Based on the four registries, 32.7% (n = 36) out of 110 devices were validated, and 51.5% (n = 1407) out of 2731 patients used validated devices. Patients using validated devices tend to use more expensive devices than those using non-validated devices (P < 0.0001). Among Korean patients with hypertension undergoing HBPM, 97.0% of the patients used upper-arm devices and 51.5% of the patients used validated devices. Our results proposed the urgent need for patient education and public accessibility to validated devices for the better management of patients with hypertension.

The validation status of HBPM devices was identified based on four registries: STRIDE BP, Medaval, the JSH, and dabl® Educational Trust. 51.5% (n = 1,407) out of 2,731 patients used validated devices and 32.7% (n = 36) out of 110 devices were validated. Patient education and public accessibility on validated devices are urgently needed. HBPM home blood pressure monitoring; JSH Japanese Society of Hypertension
Similar content being viewed by others
Introduction
Hypertension is the most important modifiable risk factor for cardiovascular diseases causing atherosclerosis and major organ damage [1]. In Korea, 28.0% of adults aged 20 and older are affected by hypertension, resulting in approximately 12.3 million people with hypertension [2]. However, the control rate (Systolic Blood Pressure [SBP] < 140 mmHg or Diastolic Blood Pressure [DBP] < 90 mmHg) among patients with hypertension in Korea is only 58.6% [3]. Efforts to increase the control rate have faced several obstacles, including inaccurate blood pressure (BP) measurements, lack of awareness, therapeutic inertia, and nonadherence to treatment [4,5,6].
Office BP measurement is essential in the diagnosis and treatment of hypertension [7]; however, it has several limitations including failure to capture BP variations [8], inability to differentiate BP phenotypes, and the potential of mistreatment in individuals with white-coat or masked hypertension. These limitations have led to the complementary use of out-of-office BP measurements such as 24-h ambulatory BP monitoring (ABPM) and home BP monitoring (HBPM). HBPM has several advantages, including its ease of use, repeated measurement capability, cost-effectiveness, and long-term BP monitoring ability in individuals with hypertension. Consequently, HBPM aids in identifying BP phenotypes, predicting cardiovascular morbidity and mortality more effectively than office BP measurements, and improving BP control rates [8,9,10,11]. Therefore, reliable devices and standardized techniques are essential to avoid the overestimation or underestimation of home BP.
Current clinical guidelines recommend the use of validated HBPM devices [7, 12,13,14], and lists of validated BP devices have been provided by various hypertension societies and organizations [15]. However, in most countries, a new BP measurement device does not require validation or a guarantee of its clinical accuracy before being released into the market. Furthermore, only a small number of automated upper-arm and wrist devices globally have evidence of validation [16,17,18]. In addition, few studies have examined the current use of validated HBPM devices in Korea, and these have been conducted in hospital settings with a small number of patients and in limited regions [19, 20]. Therefore, to ensure the proper use of accurate HBPM devices, it is necessary to assess their validation status in larger populations in various regions and different healthcare settings.
We aimed to investigate the current status of validated HBPM device use among a large population of real-world patients with hypertension treated in hospitals and clinics across different regions in Korea.
Methods
Study population
Adult patients aged ≥20 years and diagnosed with hypertension were eligible for enrollment. Individuals who did not use smartphones were excluded from the study. A total of 3002 patients were enrolled between January 2022 and August 2024. (Supplemental Fig. 1). A total of 2731 patients were eligible for analysis, after excluding 206 patients with missing data on HBPM devices or SBP values and 65 patients using kiosk devices. This study was approved by the Institutional Review Board of Wonkwang University Sanbon Hospital (No: WMCSB 202404-04), and all patients signed a written informed consent. This study was conducted in accordance with the ethical principles of the Declaration of Helsinki.
Home BP data collection
CareforMe® is a smartphone healthcare application in which patients can input their HBPM data enabling physicians to see them. Patients were asked to input their age, sex, and information of the medical institutions where they were receiving treatment (a hospital or a clinic, and the region where the medical institution was located) along with their HBPM data including the model numbers and manufacturers of the HBPM devices. The region was classified as capital (Seoul) or non-capital, when the medical institution was not located in Seoul.
Validation status of the HBPM devices
HBPM devices included upper-arm and wrist devices. The validation status of the HBPM devices was identified according to different general and regional registries: STRIDE BP, Medical device assessment and validation (Medaval, https://medaval.ie/blood-pressure-monitors/), the dabl® Educational Trust, and the Japanese Society of Hypertension [21, 22]. For STRIDE BP, we defined devices which are Validated (devices that have passed established validation procedures), Preferred (devices with at least one STRIDE BP-approved validation study), and Equivalent (equivalent to a device that has fulfilled the criteria for validated devices) or Identical devices as validated devices. HBPM devices were classified as validated in Medaval when the device was recommended with the Medaval assessment for clinical validation. The recommended devices which met the validation requirements of the dabl® Educational Trust were defined as validated devices in the dabl® Educational Trust. HBPM devices found to be validated or equivalent in the JSH were considered validated devices according to the JSH. In this study, we defined validated devices as HBPM devices found to be validated in at least one of these four registries. The authors of this study have checked validation status of each HBPM device based on these four registries. The price of each HBPM device was determined based on the online market.
Statistical analysis
Continuous variables were summarized using means and standard deviations, while categorical variables were presented as frequencies and proportions. The use of a validated device for each characteristic was assessed using independent t-tests for continuous variables and chi-square tests for categorical variables. A significance level (α) of 0.05 was applied. All data management and analyses were performed using SAS software, version 9.4.
Results
Baseline characteristics
Baseline characteristics of the study population (n = 2731) are shown in Table 1. 56.4% (n = 1539) of the population were male, and the mean age was 53.5 ± 10.8 years old. Patients were enrolled from all over South Korea including the capital (15.5%, n = 368). Three university hospitals and 162 primary clinics were identified. A total of 24.8% (n = 676) of the patients were treated at hospitals, while 72.8% (n = 1989) of the population were receiving treatment at primary clinics. In total, 110 HBPM models from 33 different manufacturers were identified. 97.0% (n = 2649) of the patients used upper-arm devices and 3.0% (n = 82) were using wrist devices. 55.0% (n = 1478) of the patients were using HBPM devices priced 35–70 USD.
Validation status of the HBPM devices
Table 2 shows the validation status of the HBPM devices according to different registries. Among the four registries, validation status was highest (number of people: 46.0%, n = 1255, number of devices: 22.7%, n = 25) with STRIDE BP, while showing the lowest validation status with JSH registry (number of people: 6.4%, n = 174, number of devices: 7.3%, n = 8). When the four registries were combined, 32.7% (n = 36) out of 110 devices were validated, and 51.5% (n = 1407) out of 2731 patients were using validated devices (Fig. 1).
Use of Validated HBPM Devices According to Patients and Devices. Out of 2731 patients (110 devices), 1407 (51.5%) used validated devices (devices: n = 36, 32.7%), including 1377 (50.4%) using validated upper-arm devices (devices: n = 32, 29.1%) and 30 patients (1.1%) using validated wrist devices (devices: 4, 3.6%). HBPM home blood pressure monitoring
Patient characteristics according to the HBPM devices are presented in Table 1. There were no significant differences between patients using validated devices and those using non-validated devices according to age and sex. The proportion of the patients treated in the capital was higher in patients using validated devices although it was not statistically significant (14.4% vs. 12.5%; P = 0.133). The proportion of patients using upper-arm devices was higher in those using validated devices than in those using non-validated devices (97.9% vs. 96.1%, P = 0.006). Patients using validated devices used more expensive devices than those using non-validated devices (P < 0.0001).
We categorized the HBPM devices according to their manufacturers and analyzed the 3 main manufacturers chosen by the patients (Table 3). Most patients using validated devices (1217 out of 1407 patients) used HBPM devices from these three manufacturers.
Discussion
Among Korean patients with hypertension undergoing HBPM, 97.0% of the patients used upper-arm devices, while 3.0% used wrist devices. Based on the four registries, 32.7% of the devices were validated. Overall, 51.5% of the patients were using validated devices, including 50.4% of upper-arm device users and 1.1% of wrist device users. To the best of our knowledge, this is the first large-scale, multi-center study to evaluate the current status of the use of validated HBPM devices among Korean patients with hypertension.
HBPM devices are automated instruments which enable users to measure their BP with a simple pressing of a button [14]. However, automated devices have limitations. These devices detect arterial oscillations and estimate the mean BP when the oscillation amplitude is maximal, subsequently calculate BP indirectly using proprietary algorithms known only by the manufacturer. Because these algorithms are not interchangeable among different devices and are not publicly disclosed, they should be tested rigorously on clinical accuracy and precision of devices against a reference standard with globally established protocols, known as validation [15]. The use of non-validated devices could undermine optimal medical practice through misdiagnosis and inappropriate management of hypertension [11, 23].
Currently, formal clinical validation of BP devices is not mandatory in all countries to be put on the market. In 2020, approximately 38% of the HBPM devices in the world were validated [17]. Few studies have been conducted regarding the current use of validated HBPM devices. An Australian study found out that only 18.3% of the upper-arm devices and 8.0% of wrist devices were found to be validated [18]. A study based on the Medaval database showed that approximately 19.9% of the devices were validated or equivalent [16]. There were two Korean regional studies about validated BP devices. Jung et al. conducted a single-center study of 212 patients and found that 38.7% of the patients used validated upper-arm devices [19]. Recently, Rhee et al. [20] conducted a survey study in two cities (number of patients = 359) and found that 52.0% of the patients were using validated upper-arm devices. As such, the validation status of HBPM devices varies between studies, and the current status of the use of validated HBPM devices is unknown. The validation status of HBPM devices in our study differs from that of two previous Korean regional studies [19, 20], although all three studies showed a higher validation status of HBPM devices in Korea than in other countries.
The validation status of HBPM devices can differ between studies according to the validation protocols adapted by the researchers. Jung et al. [19] conducted a study in a tertiary cardiology outpatient clinic located in Suwon on 212 patients assessed the validation status of HBPM devices brought to the clinic by the patients based on dabl® Educational Trust and British Hypertension Society websites. The community-based study by Rhee et al. [20], which included 673 participants residing in Goyang and Paju assessed the validation status of HBPM devices provided as pictures by participants, based on the websites (STRIDE BP, US Blood Pressure Validate Device Listing, JSH, and Medaval) and queries to the manufacturers. Our study could represent real-world data of Korean patients with hypertension because of the large number of patients comprising more heterogeneous and general populations from various regions and healthcare settings. Furthermore, we have checked the validation status of each HBPM device based on the four renowned global and regional registries, with accurate model numbers of the HBPM devices provided by the patients.
Over the past 30 years, various scientific societies and associations have developed validation protocols for BP devices, but fewer than 10% of devices have been published established protocols [7]. Several registries are operated by national or continental scientific societies, including the Japanese Society of Hypertension (JSH), STRIDE BP (affiliated with the International Society of Hypertension, European Society of Hypertension, and the World Hypertension League), Medaval (affiliated with the World Hypertension League), and the Dabl® Educational Trust (the first widely used device registry established by a group of international investigators) [24]. Internationally established validation protocols include the Association for the Advancement of Medical Instrumentation (AAMI), the American National Standards Institute (ANSI), the British Hypertension Society (BHS), the European Society of Hypertension International Protocol (ESH-IP), and the International Organization for Standardization Universal Standard (ISO). Each registry has its own lists of validated devices according to accepted validation protocols [15]: STRIDE BP accepts AAMI/ ANSI/ ISO, BHS, and ESH-IP; Medaval includes AAMI/ ANSI/ ISO, BHS, ESH-IP, and German Hypertension League Quality Seal Protocol; Dabl® Educational Trust uses AAMI, BHS, and ESH-IP; and JSH lists devices submitted by manufacturers. Because of the variety of internationally established validation protocols and accepted validation protocols by each registry, we assessed the validation status of HBPM devices across multiple global and regional registries.
Even though South Korea has the highest treatment and control rates in the world [3, 25], there are currently no Korean websites or associations on which patients can find validated devices. As our study is the first large-scale, multi-center study to analyze the current status of the use of validated HBPM devices among real-world hypertensive patients in Korea including a large number of patients with heterogeneous backgrounds and precise data, we sincerely hope that our study could become an initiative for improving the validation status of HBPM devices to help patients measure their home BP more accurately to control their BP better for minimizing their cardiovascular risks. It would be ideal but impossible to conduct accredited validation testing for all BP monitors. Nevertheless, encouraging the endorsement of clinically validated devices can help manufacturers to care more about clinical validation for accuracy as well as device safety. Furthermore, we strongly agree with an international statement of developing a single universally accredited list of BP monitors [13, 26,27,28] in the future for accurate BP measurement and better control of hypertension. Accredited websites that allow not only physicians but also patients to easily assess and check the validation status of HBPM devices are needed to enhance the public accessibility of validated devices for accurate HBPM and better management of hypertension.
This study has several limitations. First, the population using CareforMe® were included in this study. Although this study included a large number of patients across South Korea, selection bias may have been present because we did not apply formal sampling procedure among the population according to the region. Second, the socioeconomic status of each patient could affect the type of HBPM device used because our study found that patients using validated devices tended to use more expensive devices than patients using non-validated devices.
Perspective of Asia
While few studies have evaluated the current use of validated HBPM devices, our study holds particular significance as the first large-scale, multi-center investigation to assess their use among real-world Korean patients with hypertension. In Asian populations, blood pressure is more strongly associated with cardiovascular diseases such as stroke and heart failure, and greater benefits from blood pressure reduction have been observed compared to Western populations [29,30,31,32]. However, substantial heterogeneity in hypertension awareness and control rates exists across Asian countries, largely due to differences in healthcare infrastructure, public health strategies, and socioeconomic conditions. In this context, the use of clinically validated HBPM devices is essential for improving hypertension diagnosis and management in Asia. Despite these needs, our study reveals a low rate of validated device use in actual clinical practice, highlighting the urgent need to improve the validation status and accessibility of HBPM devices through coordinated efforts across Asian countries.
Conclusions
To the best of our knowledge, this is the first study to analyze the current status of the use of validated HBPM devices among hypertensive patients in Korea. Among Korean patients with hypertension undergoing HBPM, 97.0% of the patients used upper-arm devices and 51.5% of the patients were using validated devices. Our results proposed the urgent need for patient education on accurate HBPM and public accessibility to validated devices for better management of patients with hypertension.
References
Olsen MH, Angell SY, Asma S, Boutouyrie P, Burger D, Chirinos JA, et al. A call to action and a lifecourse strategy to address the global burden of raised blood pressure on current and future generations: the lancet commission on hypertension. Lancet. 2016;388:2665–712.
Kim HC, Lee H, Lee HH, Son D, Cho M, Shin S, et al. Korea hypertension fact Sheet 2023: analysis of nationwide population-based data with a particular focus on hypertension in special populations. Clin Hypertens. 2024;30:7.
Kim HC, Lee H, Lee HH, Ahn SV, Lee JM, Cheon DY, et al. Korea Hypertension Fact Sheet 2024: nationwide population-based analysis with a focus on young adults. Clin Hypertens. 2025;31:e11.
Smith AP, Overton K, Rakotz M, Wozniak G, Sanchez E. Target: BP: a National initiative to improve blood pressure control. Hypertension. 2023;80:2523–32.
Sakhuja S, Colvin CL, Akinyelure OP, Jaeger BC, Foti K, Oparil S, et al. Reasons for uncontrolled blood pressure among US adults: data from the US National Health and Nutrition Examination Survey. Hypertension. 2021;78:1567–76.
Nilsson PM. Sustainable hypertension care - How can it be achieved?. J Intern Med. 2023;294:242–4.
Mancia G, Kreutz R, Brunström M, Burnier M, Grassi G, Januszewicz A, et al. 2023 ESH Guidelines for the management of arterial hypertension. The Task Force for the management of arterial hypertension of the European Society of Hypertension: Endorsed by the International Society of Hypertension (ISH) and the European Renal Association (ERA). J Hypertens. 2023;41:1874–2071.
George J, MacDonald T. Home blood pressure monitoring. Eur Cardiol. 2015;10:95–101.
Ntineri A, Kalogeropoulos PG, Kyriakoulis KG, Aissopou EK, Thomopoulou G, Kollias A, et al. Prognostic value of average home blood pressure and variability: 19-year follow-up of the Didima study. J Hypertens. 2018;36:69–76.
Ihm SH, Park JH, Kim JY, Kim JH, Kim KI, Lee EM, et al. Home blood pressure monitoring: a position statement from the Korean Society of Hypertension Home Blood Pressure Forum. Clin Hypertens. 2022;28:38.
Lee EM. When and how to use ambulatory blood pressure monitoring and home blood pressure monitoring for managing hypertension. Clin Hypertens. 2024;30:10.
McEvoy JW, McCarthy CP, Bruno RM, Brouwers S, Canavan MD, Ceconi C, et al. 2024 ESC Guidelines for the management of elevated blood pressure and hypertension. Eur Heart J. 2024;45:3912–4018.
Sharman JE, O’Brien E, Alpert B, Schutte AE, Delles C, Hecht Olsen M, et al. Lancet commission on hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressure. J Hypertens. 2020;38:21–9.
Kim HL, Lee EM, Ahn SY, Kim KI, Kim HC, Kim JH, et al. The 2022 focused update of the 2018 Korean Hypertension Society Guidelines for the management of hypertension. Clin Hypertens. 2023;29:11.
Picone DS, Padwal R, Stergiou GS, Cohen JB, McManus RJ, Eckert S, et al. How to find and use validated blood pressure measuring devices. J Hum Hypertens. 2023;37:108–14.
Picone DS, Campbell NRC, Schutte AE, Olsen MH, Ordunez P, Whelton PK, et al. Validation status of blood pressure measuring devices sold globally. Jama. 2022;327:680–1.
Murthy S, Rajwar E, Johnson R, Shrivastava K, Kamath P, Rahman R, et al. Validation of blood pressure devices as per 2020 World Health Organization technical specifications: a scoping review of global literature. Hypertension. 2023;80:1110–6.
Picone DS, Deshpande RA, Schultz MG, Fonseca R, Campbell NRC, Delles C, et al. Nonvalidated home blood pressure devices dominate the online marketplace in Australia: major implications for cardiovascular risk management. Hypertension. 2020;75:1593–9.
Jung MH, Kim GH, Kim JH, Moon KW, Yoo KD, Rho TH, et al. Reliability of home blood pressure monitoring: in the context of validation and accuracy. Blood Press Monit. 2015;20:215–20.
Lee KJ, Rhee MY. Status of home blood pressure measurement in treated hypertensive patients. Results of a survey from two cities in Korea. J Clin Hypertens. 2024;26:825–31.
Stergiou GS, O’Brien E, Myers M, Palatini P, Parati G. STRIDE BP: an international initiative for accurate blood pressure measurement. J Hypertens. 2020;38:395–9.
Atkins N, O’Brien. The dabl Educational Trust device equivalence procedure. Blood Press Monit. 2007;12:246–9.
Whelton PK, Picone DS, Padwal R, Campbell NRC, Drawz P, Rakotz MK, et al. Global proliferation and clinical consequences of non-validated automated BP devices. J Hum Hypertens. 2023;37:115–9.
Cohen JB, Padwal RS, Gutkin M, Green BB, Bloch MJ, Germino FW, et al. History and justification of a national blood pressure measurement validated device listing. Hypertension. 2019;73:258–64.
Collaboration NCDRF. Worldwide trends in hypertension prevalence and progress in treatment and control from 1990 to 2019: a pooled analysis of 1201 population-representative studies with 104 million participants. Lancet. 2021;398:957–80.
Campbell NR, Gelfer M, Stergiou GS, Alpert BS, Myers MG, Rakotz MK, et al. A call to regulate manufacture and marketing of blood pressure devices and cuffs: a position statement from the world hypertension league, international society of hypertension and supporting hypertension organizations. J Clin Hypertens. 2016;18:378–80.
Ringrose JS, Padwal R. Automated blood pressure measuring devices: how are they clinically validated for accuracy?. J Hum Hypertens. 2023;37:101–7.
Sharman JE, Padwal R, Campbell NRC. Global marketing and sale of accurate cuff blood pressure measurement devices. Circulation. 2020;142:321–3.
Kario K, Chen CH, Park S, Park CG, Hoshide S, Cheng HM, et al. Consensus document on improving hypertension management in Asian patients, taking into account Asian characteristics. Hypertension. 2018;71:375–82.
Kario K, Okura A, Hoshide S, Mogi M. The WHO Global report 2023 on hypertension warning the emerging hypertension burden in globe and its treatment strategy. Hypertens Res. 2024;47:1099–102.
Kario K, Hoshide S, Mogi M. Hypertension treatment up-date on World Hypertension Day 2024: current status and future prospects in Asia. Hypertens Res. 2024;47:1763–5.
Kario K, Chia YC, Siddique S, Turana Y, Li Y, Chen CH, et al. Seven-action approaches for the management of hypertension in Asia - The HOPE Asia network. J Clin Hypertens. 2022;24:213–23.
Funding
This study was partially funded by the Korea Health Industry Development Institute (RS-2023-00264340) and grant by Wonkwang University (number: 2024-03-24-213).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Ethic approval
This study was approved by the Institutional Review Board of Wonkwang University Sanbon Hospital (No: WMCSB 202404-04), and all patients signed a written informed consent. This study was conducted in accordance with the ethical principles of the Declaration of Helsinki.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lee, J., Kim, YM., Yoo, S. et al. Current status of the use of validated home blood pressure monitoring devices among Korean patients with hypertension. Hypertens Res 48, 2811–2818 (2025). https://doi.org/10.1038/s41440-025-02364-z
Received:
Revised:
Accepted:
Published:
Version of record:
Issue date:
DOI: https://doi.org/10.1038/s41440-025-02364-z