Table 1 Performance of conventional X-ray detectors for medical imaging

From: Halide lead perovskites for ionizing radiation detection

 

Atomic number

Applied electric field (V μm−1)

μτ product (cm2 V−1)

Sensitivity (μC  Gy−1air cm−2)

Spatial resolution (lp mm−1)

Lowest detectable dose rate (μGyair s−1)

Refs

Sia

14

0.5

>1

8

4.5

<8300

24,26

a-Sea

34

10

10−7

20

~15

5.5

21,70

HgI2a

53,80

10

10−4

1600

3.93

10

84

CZTa

48, 52

0.1–1

0.01

318

10

50

85,86

MAPbBr3a

35, 82

0.05

0.012

2.1 × 104

10

0.039

18,66

MAPbI3a

53, 82

0.24

0.010

1.1 × 104

3

<5000

64

Cs2AgBiBr6a

47, 53, 55, 83

0.025

0.0060

105

0.060

47

CsI(Tl)b

53, 55, 81

5370

10

0.18

27,28

GOS(Tb)b

64, 65

7.3

4.75

29

  1. aThe direct X-ray detection by semiconductor
  2. bThe indirect X-ray detection by scintillator