Table 1 Comparison with recently reported virus detection methods
From: Electrical pulse-induced electrochemical biosensor for hepatitis E virus detection
Analyte | Linear range | LOD | References |
|---|---|---|---|
Influenza virus (Fluorometric) | 1–10 × 10−11 g mL−1 | 3 × 10−10 g mL−1 | |
Norovirus RNA (Colorimetric) | 102–106 copies mL−1 | 13.2 copies mL−1 | |
Influenza A virus (Colorimetric) | 5 × 10−15–5 × 10−6 g mL−1 | 44.2 × 10−15 g mL−1 | |
Influenza A virus (Fluorometric) | 50–1.0 × 104 PFU mL−1 | 50 PFU mL−1 | |
Dengue RNA (Fluorometric) | 5–500 × 10−9 g mL−1 | 5.2 × 10–9 g mL−1 | |
HBV (Fluorometric) | >264 × 10−9 g mL−1 | 8.3 × 10–9 g mL−1 | |
Norovirus RNA (Microfluidic) | 1–3.5 × 10−9 M | 1 × 10−11 M | |
Norovirus RNA (Fluorometric) | 2–18 copies mL−1 | 1.2 copies mL−1 | |
HBV (Fluorometric) | 0.01–1 IU mL−1 | 0.4 IU mL−1 | |
HEV RNA (RT-qPCR) | 103–106 IU mL−1 | 2.1 × 104 IU mL−1 | |
HEV RNA (RT-qPCR) | 8.75 × 103–8.75 × 104 copies mL−1 | 8.75 × 103 copies mL−1 | |
HEV RNA (RT-qPCR) | 10–109 copies mL−1 | 10 copies mL−1 | |
HEV RNA (RT-qPCR) | 2.01 × 103–1.71 × 105 GC g−1 | 10 GC g−1 | |
HEV-LP | 10−12 –10−15 g mL−1 | 8 × 10−14 g mL−1 | This work |
HEV (pulse impedance) | 102–107 RNA copies mL−1 | 96.7 RNA copies mL−1 | This work |