Table 1 Performance of comparable cavity ring-down spectrometers operating near 4.5 μm
From: Author Correction: Mid-infrared supermirrors with finesse exceeding 400 000
Reference | α0 at 1 s (cm−1 Hz−1/2) | L (m) | α0 at 1 s for L = 10 cm (cm−1 Hz−1/2) |
|---|---|---|---|
INO-CNR18 | 5.0 × 10−9 | 1 | 5.0 × 10-7 |
VTT19 | 2.1 × 10−9 | 0.4 | 3.4 × 10-8 |
Nagoya20 | 1.1 × 10−8 | 0.11 | 1.3 × 10-8 |
NIST21 | 2.6 × 10−11 | 1.5 | 5.9 × 10-9 |
LLNL15,22 | 1.2 × 10−10 | 0.67 | 5.3 × 10-9 |
This work | 6.0 × 10−11 | 0.79 | 3.7 × 10-9 |