Table 2 Comparison of visual functions and subjective appraisals on life quality before and after wearing Onyx.
From: Clinical efficacy of a head-mounted device for central vision loss
Metric | Without Onyx, mean ± SD | With Onyx, mean ± SD | P value | Percentage of improvement, % | Number of patients with significant improvement, n (%) |
|---|---|---|---|---|---|
Visual functions | |||||
Near VA, LogRAD | 1.10 ± 0.13 | 0.99 ± 0.14 | < 0.001* | 10.00% | 25/41 (60.98%) |
Distance VA, LogMAR | 1.58 ± 0.27 | 1.36 ± 0.29 | < 0.001* | 13.92% | 33/41 (80.49%) |
Combined recognition score, point | 7.32 ± 2.53 | 12.68 ± 1.63 | < 0.001* | 73.22% | 37/41(90.24%) |
Face recognition score | 7.10 ± 2.64 | 12.48 ± 2.17 | < 0.001* | 75.77% | N/A |
Object recognition score | 7.54 ± 3.01 | 12.88 ± 1.71 | < 0.001* | 70.82% | N/A |
Scores of the LVQOL questionnaire | |||||
Total score, point | 52.05 ± 3.75 | 60.95 ± 3.80 | < 0.001* | 17.10% | 28/41 (68.29%) |
Distance vision, mobility and lighting | 26.80 ± 4.28 | 30.02 ± 3.37 | < 0.001* | 12.01% | N/A |
Adjustment | 9.00 ± 1.69 | 10.41 ± 1.76 | < 0.001* | 15.67% | N/A |
Reading and fine work | 7.93 ± 1.33 | 9.68 ± 1.37 | < 0.001* | 22.07% | N/A |
Activities of daily living | 8.00 ± 1.50 | 9.33 ± 1.28 | < 0.001* | 16.63% | N/A |