Table 1 Agreement of absolute values for both Capno-SvO2 and fiberoptic SvO2 as compared to the reference method, for the various different phases of the experiment.

From: Capnodynamic assessment of mixed venous oxygen saturation in a porcine experimental endotoxemic model

Hemodynamic intervention

Compared methods

Bias (percentage points)

ULOA (CI) (percentage points)

LLOA (CI) (percentage points)

Baseline/precision

Capno vs CO-oxi

1

12

– 10

 

Fiber vs CO-oxi

– 2

7

– 11

Sepsis induction

Capno vs CO-oxi

1

19

– 18

 

Fiber vs CO-oxi

– 4

6

– 14

Crystalloid bolus

Capno vs CO-oxi

4

15

– 8

 

Fiber vs CO-oxi

– 1

11

– 13

FiO2

Capno vs CO-oxi

1

16

– 15

 

Fiber vs CO-oxi

– 3

6

– 12

PEEP

Capno vs CO-oxi

1

15

– 14

 

Fiber vs CO-oxi

– 6

5

– 17

Dobutamine

Capno vs CO-oxi

5

22

– 13

 

Fiber vs CO-oxi

– 1

17

– 18

Norepinephrine

Capno vs CO-oxi

1

16

– 14

 

Fiber vs CO-oxi

– 5

6

– 17

Hemorrhage

Capno vs CO-oxi

– 1

16

– 17

 

Fiber vs CO-oxi

– 5

18

– 27

Restitution of whole blood

Capno vs CO-oxi

4

21

– 13

 

Fiber vs CO-oxi

– 3

6

– 11