Table 1 Image acquisition parameters per scan.
From: Deep learning reconstruction enhances tophus detection in a dual-energy CT phantom study
Scan number | I [mA] at 80 kVp | I [mA] at 135 kVp | Q [mAs] at 80 kVp | Q [mAs] at 135 kVp | CTDIvol [mGy] | DLP [mGy* cm] | EED [mSv] |
|---|---|---|---|---|---|---|---|
1 | 60 | 10 | 16.5 | 2.75 | 0.5 | 7.4 | 0.003 |
2 | 90 | 15 | 24.75 | 4.13 | 0.7 | 11.1 | 0.004 |
3 | 110 | 20 | 30.25 | 5.5 | 0.9 | 14.1 | 0.006 |
4 | 170 | 30 | 46.75 | 8.25 | 1.3 | 21.5 | 0.009 |
5 | 230 | 40 | 63.25 | 11 | 1.8 | 28.9 | 0.012 |
6 | 290 | 50 | 79.75 | 13.75 | 2.3 | 36.3 | 0.015 |
7 | 400 | 70 | 110 | 19.25 | 3.2 | 50.5 | 0.02 |
8 | 510 | 90 | 140.25 | 24.75 | 4 | 64.6 | 0.026 |
9 | 630 | 110 | 173.25 | 30.25 | 5 | 79.4 | 0.032 |
10 | 800 | 140 | 220 | 38.5 | 6.3 | 100.9 | 0.04 |