Table 3 Image quality parameters: mean SNR and CNR with standard deviation.
From: Deep learning reconstruction enhances tophus detection in a dual-energy CT phantom study
Raster phantom | Bio-phantom | |||
|---|---|---|---|---|
SNR (80 kVp/60 keV) | CNR (80 kVp/60 keV) | SNR (80 kVp/60 keV) | CNR (80 kVp/60 keV) | |
FBP | 11.90 ± 0.77 | 11.38 ± 0.71 | 36.53 ± 13,93 | 16.31 ± 5.83 |
IR | 12.26 ± 0.41 | 11.69 ± 0.36 | 55.39 ± 20,57 | 24.86 ± 9.09 |
DLR mild | 63.52 ± 24.70 | 59.99 ± 23.66 | 53.58 ± 16.98 | 28.42 ± 7.69 |
DLR strong | 112.80 ± 35.98 | 107.20 ± 34.30 | 84.03 ± 20.65 | 43.80 ± 9.98 |
P | < 0.0001 | < 0.0001 | < 0.0001 | < 0.0001 |