Table 1 CFUs of biofilm-embedded bacteria prior (control) and after steam disinfection. The results are expressed as the mean ± the standard deviation.

From: Disinfection of medical devices with a steam machine that operates at atmospheric pressure and is suitable for home usage

CFU/ml

Control

1 min

5 min

10 min

S. aureus ATCC

2 ± 0.27*108

14 ± 2*103

0

0

E75

2.18 ± 0.14*108

13 ± 1*103

0

0

E166

1.39 ± 0.36*108

11.3 ± 2.3*103

0

0

E276

2.68 ± 0.42*108

15.3 ± 2.2*103

0

0

USA300

1.82 ± 0.27*108

4.67 + 1.67*103

0

0

HG001

1.52 ± 0.34*108

10.67 ± 0.67*103

0

0

S. epidermidis ATCC

3.9 ± 0.43*108

12.3 ± 3.3*103

0

0

E. coli ATCC

9.6 ± 1.07*108

9 ± 3*103

0

0

P. aeruginosa ATCC

2.8 ± 0.58*108

13.3 ± 2.7*103

0

0

K. oxytoca

1.27 ± 0.16*108

11 ± 2*103

0

0

K. pneumonia ATCC

2.34 ± 0.54*108

19.67 ± 3.5*103

0

0

A. baumanni ATCC

2 ± 0.32*108

8.67 ± 1.67*103

0

0