Table 4 Recovery and relative standard of the proposed sensors against official method (HPLC).

From: Electrochemical investigation of oxytrol in fizzy drinks and juices via graphene and polymer printed PVC by ANOVA and design expert

Electrode Type

[oxytrol]

Taken

ppm

Calibration method

Standard addition method

Found

ppm

Recovery

(%)

RSD

%

Found

ppm

Recovery

(%)

RSD

%

I

7.878

7.756

98.45

0.52

7.716

97.94

0.45

II

7.814

99.14

0.32

7.856

99.72

0.28

III

7.809

99.12

0.24

7.796

98.95

0.38

Official method HPLC29

7.877

99.84

0.31

7.806

99.09

0.25