Table 7 Diagnostic performance of X-DOT-TB, QFT-Plus and TST assays.
Diagnostic Parameters | Assays | ||
|---|---|---|---|
X-DOT-TB (n = 325) | QFT-Plus (n = 126*) | TST (n = 322) | |
Sensitivity % (95% CI) | 79.5 (77.4–81.5) | 55.7 (52.9–58.5) | 35.8 (34.4–37.1) |
Overall Specificity % (95% CI) | 85.1 (83.2–87.0) | 78.1 (75.3–80.9) | 82.2 (80.3–84.1) |
PPV % (95% CI) | 81.1 (79.0-83.2) | 64.5 (61.4–67.5) | 60.4 (58.5–62.2) |
NPV % (95% CI) | 79.6 (77.5–81.6) | 70.3 (67.1–73.4) | 60.2 (58.4–62.0) |
LR+ (95% CI) | 6.4 (5.8–6.9) | 2.2 (1.7–2.6) | 1.9 (1.6–2.2) |
LR− (95% CI) | 0.24 (-) | 0.6 (0.5–0.7) | 0.79 (0.63–0.95) |
Accuracy % (95% CI) | 79.1 (77.0-81.2) | 56.6 (53.7–59.4) | 36.6 (35.1–38.0) |
Diagnostic Odd Ratio | 26.6 (27.7–25.4) | 3.6 (3.02–4.3) | 2.4 (2.1–2.7) |