Table 4 System accuracy results for glucose concentration \(\ge\)100 mg/dL.
From: The open-source multi-wavelength non-invasive blood glucose sensing system
Method | n (\(\ge\)100 mg/dL) | ± 5 % | ± 10 % | ± 15 % | ± 20 % |
|---|---|---|---|---|---|
SR1 | 77 | 15/77 (19.5 %) | 25/77 (32.5 %) | 38/77 (49.4 %) | 50/77 (64.9 %) |
SR2 | 77 | 15/77 (19.5 %) | 30/77 (39.0 %) | 39/77 (50.6 %) | 48/77 (62.3 %) |
MR | 77 | 18/77 (23.4 %) | 35/77 (45.5 %) | 47/77 (61.0 %) | 56/77 (72.7 %) |
ST1 | 77 | 25/77 (32.5 %) | 49/77 (63.3 %) | 73/77 (94.8 %) | 77/77 (100.0 %) |
ST2 | 77 | 27/77 (35.1 %) | 53/77 (68.8 %) | 74/77 (96.1 %) | 77/77 (100.0 %) |
MT | 77 | 34/77 (44.2 %) | 58/77 (75.3 %) | 76/77 (98.7 %) | 77/77 (100.0 %) |