Table 3 Repeatability test for the intraocular pressure obtained with the IC01 at visit 2.

From: Agreement and repeatability of a novel portable transpalpebral tonometer for home tonometry versus non-contact and rebound tonometry

IOP at visit-2

Right eye (n = 69)

Left eye (n = 69)

ICC (95% CI)

0.77 (0.44–0.79)

0.75 (0.48–0.78)

CV (%)

  

Mean ± SD (range)

8.65 ± 4.15 (0–39.99)

7.68 ± 2.61 (0.5–39.358)

  1. IOP,intraocular pressure ; ICC, intraclass correlation coefficient; CI,confidence interval; CV, coefficient of variation; SD, standard deviation.