Table 4 Upper and lower limits of agreement between IC01 and TOPCON CT-800.

From: Agreement and repeatability of a novel portable transpalpebral tonometer for home tonometry versus non-contact and rebound tonometry

Subgroup

n

MD ± SD

Significance

Limits of agreement

Upper

Upper

Right eyes

69

&-0.69 ± 3.66

 

& + 8.9

&-9.8

Left eyes

68

&-0.72 ± 2.99

 

& + 7.8

&-9.2

Right eyes

     

IC01 then TOPCON CT-800

35

&-0.42 ± 2.48

p = 0.362

& + 8.1

&-9.2

TOPCON CT-800 then IC01

34

&-0.78 ± 4.22

& + 8.7

&-11.02