Table 1 Comparison to IFCN guidelines. IRN for [0.5, 100] Hz bandwidth. CMRR measure for a 50 Hz input sinusoidal signal of 100 \(mV_P\). Measurements executed with for PGA=24 V/V and \(F_S=250~Hz\).
From: EEG glasses for real-time brain electrical activity monitoring
Fs (Hz) | Resolution (\(\mu\)V/bit) | Impedance (\(M\Omega\)) | CMRR (dB) | \(IRN\) (\(\mu\)\(V_P\)) | |
|---|---|---|---|---|---|
IFCN24guidelines | \(\ge 200\) | \(\le 0.5\) | \(\ge 100\) | \(\ge 110^1\) | \(\le 1.5\) |
e-Glass system | 250 | 0.023 | \(>1000\) | \(110^2\) | 1.07 |