Abstract
Carbon paste sensor based on p-chlorophenyl maleanilic acid (pCl-MA) as an ionophore was synthesized and investigated as an Al(III) ion selective sensor. The developed sensor shows Nernstian slope of 20.32 ± 1.18 mV decade− 1 over the working concentration range of 1.0 × 10− 6 − 1.0 × 10− 1 mol L− 1 of Al (III) ion with a detection limit of 3.3 × 10− 7 mol L− 1. The sensor exhibited a stable potential response in the pH range from 4.0 to 6.0 and a fast response time of 8 s. The prepared sensor indicated a very good selectivity to Al (III) over a variety of cations. The sensor could be easily employed for determination of Al (III) ions in pharmaceutical products, soft drink and biological fluids of kidney failure and Alzheimer disease patients. The reaction mechanism between pCL-MA and Al (III) was confirmed by FT-IR spectroscopic analysis, SEM and EDX.
Introduction
Aluminum is considered as vital and one of the most essential trace elements and its concentration determines whether it will be effective or toxic. To help the vital body physiological functions working well the standard concentration should be underneath an allowable level1. Aluminum is the most abundant metal in the earth crust and one of the rich elements in soil minerals and rocks. It is widely used in transport, architecture, household and industrial machines, also in potable water treatment units as a flocculating agent2,3,4. Because of its activity it is rarely found in the metallic form. Aluminum is naturally found in plants, such as potatoes, rice, and mushrooms5, the kind of plant, the soil’s acidity, and the irrigation water used, affect impact how much aluminum is found in vegetables and fruits6. Also, it is used as food additives like sodium aluminum sulphate, sodium aluminum phosphate, and sodium aluminosilicate which serve as stabilizers and extend the shelf life of most processed foods.
Recent studies reported that the increased concentration of aluminum in bone and in the central nervous system is a main source of neurotoxicity. Also, Aluminum plays a major role in Parkinson’s disease pathology, dialysis and Alzheimer’s disease7,8,9. Humans may ingest a large amount of aluminum because it may be found in food, or transferred to food from kitchen materials, food packaging, foils, cooking utensils, and other product structures, and from food to the human body. Aluminum containers are coated with a protective oxide layer, which is susceptible to erosion due to any physical effect or chemical reaction. Aluminum can seep into the cooked food because of this layer’s deterioration and the use of these components in cooking foods10.
Other sources of aluminum are vaccination, antiperspirants, and some drugs, such as toothpaste, antacids, and buffered aspirin11 that makes aluminum determination a vital demand. The methods that have been established for aluminum determinations; containing the use of atomic absorption12,13,14, flow injection analysis15, inductively coupled plasma atomic emission spectrometry16. Most of the previously mentioned methods are costly and require qualified personnel for pre-processing, execution of the analysis, data collection and data analysis and interpretation. Therefore, developing low-cost, easily manipulated methods with a high level of sensitivity is essential. So, ion–selective sensors are considered as alternatives to other analytical methods in the detection of many ionic species. The promising use of sensors in the fields of agriculture, environmental, pharmaceutical and biological analysis is what stimulated that work to develop carbon sensors precise, fast, reproducible and selective for potentiometric investigation of aluminum ions in real samples such as pharmaceutical, canned beverages and biological fluids of kidney failure and Alzheimer disease patients. Sensitivity and selectivity of the ion–selective sensors toward the ion under investigation are controlled by the modifier or electroactive material which considered the main component of the paste. Several chemical compounds were utilized as a modifier (electroactive material) in the preparation of the electrochemical sensors to detect aluminum ions like Schiff bases17, crown18, polymer19,20,21 however these sensors have many drawback such as time consumed in preparation, long response time, narrow working range, presence of internal solution, the PVC membrane cannot withstand the mechanical stress. The proposed carbon sensor was modified with p-chlorophenyl maleanilic acid (pCl-MA) which is one of the amide group ligands offers two coordination sites O of carboxylic group and N of the amide22.
Experimental
Reagents and apparatus
The constituents of the paste mixture such as graphite powder and tricresyl phosphate (TCP) were supplied by Aldrich while other plasticizers as dibutyl phthalate (DBP) and o-Nitrophenyloctylether (o-NPOE) were purchased from Fluka and BDH Industries Ltd, respectively. NiCl2.6H2O, MnCl2.4H2O, and Al2(SO4)3.16H2O were supplied from BDH chemical Ltd. SrCO3, Pb(CH3COO)2.3H2O, CaCL2 were purchased from laboratory chemicals, Fluka, Fischer scientific, El Madina For chemicals, Adwic, and Scientific Bismarco Office (S.B.O) respectively; while CuCl2.2H2O, CoCl2.6H2O, and FeCl3.6H2O was purchased from LOBA Chemie (laboratory reagents for chemicals). (NH4)6 Mo7 O24. 4H2O was purchased from Aldrich.
Fresh stock solutions (1.0 × 10− 1 mol L− 1) of the metal ions were made by dissolving the precisely weighed quantity of metal salts in the proper solvent then complete to the mark with the distilled water to achieve the desired concentration. By serially diluting stock solutions, other diluted solutions (1.0 × 10− 2 − 1.0 × 10− 8 mol L− 1) were created. Throughout the whole work, all solutions were maintained in dark quick-fit bottles in the refrigerator.
The potential measurements were performed using a digital multimeter DT.9205 A. Ag/AgCl reference electrode (Hanna) in conjugation with the developed carbon sensors as working electrode was used. The pH measurements were made using Hanna mV/pH/Temperature meter (Model: pHS – 3CW). Automatic Micropipettes, within volume range of 100–1000 µL (Model: Accupipette USA) was used to accurately measure the small volumes. The FT-IR spectra were carried out on a Perkin-Elmer 1650 spectrometer (4000–400 cm− 1) in potassium bromide pellets at the Micro-analytical Center, Cairo University Egypt. The Al (III) content was determined using inductively coupled plasma (ICP) (National research center, Egypt).
Preparations of real samples
Pharmaceutical
Maalox 460 mg Al (OH)3 Anti Acid, produced by Sanofi. To 250 mL measuring flask, 1.824 mL of Maalox was added and diluted to the mark using double distilled water to prepare 1.0 × 10− 2 mol L− 1 solution, then subjected to estimate Al (III) ions content by the carbon sensor and ICP methods.
Canned beverages
To estimate the aluminum content in the soft drink which packed in cane and plastic bottle, commercially available canned and bottles products were purchased from a local market in Giza, Egypt, were examined using the carbon sensor and ICP.
Biological fluids: Certified serum samples were supplied from El Rahma Hospital (El Rahma Laboratories, Giza, Egypt). Human serum; patient 1: female, Age: 56, patient suffer from chronic kidney failure, Patient 2: male, Age: 63, patient suffer from acute kidney failure and patient 3: male, Age: 65, patient suffer from Alzheimer.
500 µL of the serum sample was mixed with 1mL of 0.2 mol L−1 HNO3 in 10 mL measuring flask then completed to the mark using double distilled water, then analyzed by the proposed method and ICP23,24. To reduce matrix effects, samples were prepared by using HNO3 which acting as an oxidizing agent to break down the organic matrix and as a matrix modifier to reduce interferences and diluted with water to obtain a clear sample suitable for analysis25,26,27,28.
All experiments were performed in compliance with the relevant laws and institutional guidelines (Profession and Ethics Regulations, Resolution of the Ministry of Health and Population No. 238/2003). The protocol was approved by Faculty of Science (Cairo University) institutional committee(s) and all participants provided written informed consent to approve the experiments done with human subjects.
Preparation of working carbon sensor
CPEs were prepared by grinding 500 mg pure graphite powder and 0.0–20 mg of pCl-MA as modifier. The content was transferred to mortar and mixed well with plasticizer (0.2 mL of o-NPOE, TCP, or DBP) until homogenization of this mixture was achieved, then the resulted paste was used to fill the electrode body, as shown in Table (2).The surface of the resulting CPE was buffed using a filter paper to get a new working surface.
Results and discussion
Paste composition and selectivity of the developed carbon sensor
To determine which metal ion, the proposed modifier pCl-MA has a high affinity for it, calibration graphs were constructed for several inorganic cations using the pCl-MA based carbon sensor. As shown in Table 1 the proposed sensor shows a non-Nernstian attitude to all the examined cations except for Al (III) ions.
Several carbon sensors were prepared by mixing different mass ratios of graphite powder, pCl-MA as an electroactive material and TCP as plasticizer, the potential of them was measured as function of Al (III) concentration and constructed in calibration plot. To achieve the repeatability and precision of the measurements, the calibration plot for each composition was repeated 5 times. It is clear from the data listed in Table 2 that the performance of the sensors improved as the amount of pCl-MA increase till the best Nernstian slope of 20.32 ± 1.18 mV decade− 1 was obtained by sensor (IV) of composition 2.0% (wt/wt) pCl-MA, 31.24% (wt/wt) TCP and 66.67% (wt/wt) graphite powder, over working concentration range 1.0 × 10− 6 to 1.0 × 10− 1 mol L− 1 of Al(III) with 3.3 × 10− 7 mol L− 1 as a detection limit. Sensor (IV) was used for subsequent study.
The proportion and nature of the plasticizer is a truly important factor; that must be enhanced to support the optimal activity of the synthesized paste and to eliminate the electrical asymmetry. Several carbon pastes were prepared with different plasticizers, namely, TCP, o-NPOE and DBP to determine the optimum plasticizer integrating in the construction of the proposed electrode. It is clear from the results presented in Table 2, that the sensor plasticized by TCP shows the best trivalent Nernstian slope, reproducibility and good stability; this may be attributed to its high dielectric constant, relatively high molecular weight and the homogenized paste which improve flexibility and durability and diminish the electrical asymmetry of the paste.
Where the proposed sensor achieves Nernstian response, selectivity is a second challenge which might be solved. Theoretically, that the potentiometric selectivity of the sensor is based on the composition of the paste and the complexation specificity of the modifier involved. Depending on the cation size, modifier structure and extractability or complex formation is the determining factor for selectivity29. This was confirmed by FT-IR, and UV-spectroscopic analysis.
FT-IR spectrum of the sensor (IV) surface before immersion in aluminum solution was performed and shows the major functional groups of pCl-MA at 1612, 3433, 1214, 1142 and 2924 Cm− 1 are assigned to ν (C = O) stretching, ν (O-H) of carboxylic acid, ν (C-O) stretching, ν (C-N-C) asymmetric and ν(C-H) stretching vibration, respectively.
While after immersing the same sensor surface in Al (III) ion solution, a variation in the FT-IR spectrum was noticed as depicted in Fig. 1. The complex formation takes place through chelation via ν (O-H) of carboxylic and -N- of the amide group (C-N-C) which are shifted to 3480 and 1150 cm− 1, respectively30,31. Also, the bands due to ν (C-O), ν (C = O) and ν(C-H) were shifted to 1204, 1628 and 2909 cm− 1, respectively upon complexation.
A complex formed at the surface of carbon sensor by extraction of Al (III) ions from the test solution into the paste phase by the aid of suitable modifier content and plasticizer. This was established by using scanning electron microscope (SEM) and energy dispersive X-ray analysis (EDX) before and after immersion in Al (III) solution. The complex formation was proofed by appearance of brightened spots light the cavities between carbon particles, as shown in Figs. 2 and 3.
The proposed sensor was prepared and its response against the primary Al (III) ion and different cationic species was examined separately (Table 1). Sensor’s selectivity can be powerfully expressed in terms of selectivity coefficients, and they were determined by separate solution method (SSM)32,33. If the value of selectivity coefficient equal to 1.0 indicates equal response to both primary ion and interfering ions, while if the value of selectivity coefficient is smaller than 1.0 it shows that the sensor is selective to the primary ion over the interfering ions. It is obvious from the data presented in Table 1, that the selectivity coefficients values are very small in most examined cations, which proved high selectivity of the proposed sensor toward Al (III) ions over other cations. This can be attributed to the selectivity, interaction between pCl-MA ionophore and Al (III) ion; also, due to the fast exchange34,35.
pH effect
The pH effect on the proposed Al (III) sensor was examined in the pH range of 2–11 on 1.0 × 10− 5 and 1.0 × 10− 3 mol L− 1 of Al (III) solutions. The potential measured by the proposed sensor (IV) was plotted against the pH of Al (III) ions solution as presented in Fig. 4. From the Figure it is found that the potential constant over pH is from 4 to 6. The potential readings increase at pH less than 4 due to response of the sensor to the H+ ion in addition to Al (III) ions in the solution. While the observed decline in the potential at pH greater than 6, may be attributed to hydrolysis of Al (III) ions and formation of some insoluble or soluble hydroxide36,37.
Effect of temperature
The effect of temperature on the performance of the developed sensor was studied by constructing the potential calibration graphs in temperature range from 10 to 60 °C, to select the optimum temperature for the aluminum determination.
The isothermal coefficient (dE°/dt) cell was calculated by plotting the standard electrode potential against (t-25) using Antropov’s equation38. It is found to be 4.291 × 10− 3 V °C− 1. According to the isothermal coefficient value the sensor exhibited excellent thermal stability up to 60 °C without deviating from the trivalent Nernstian slope.
Response time
The response time of the sensor can be expressed as the average time of the sensor’s needed to reach steady state potential reading39. Dynamic response of ion selective electrode is generated by selective complexation of the pCl-MA as neutral carrier molecule dispersed in a paste matrix to the primary ion, Al (III). The response time of the proposed carbon sensor was measured through the immersion of the carbon sensor in conjugation with the reference electrode in aluminum solution which has been successively changed from lower concentration 1.0 × 10− 6 to higher concentration 1.0 × 10− 1 mol L− 1. The potential readings were plotted against time as graphically represented in Fig. 5. It was found that the average response time of the sensor (IV) is 8 s which reflect the fast exchange kinetics of complexation–decomplexation of aluminum ions with the pCl-MA at the paste- test solution interface.
Analytical applications
The proposed carbon sensor was applied for the micro- determination of Al (III) ions concentrations in a variety of real samples including pharmaceutical formulation, soft drink and biological fluids to prove the applicability of the selected method.
The proposed electrode is shown to be sufficiently useful in the direct potentiometric determination of Al (III) in Maalox using calibration method.
The Al (III)-carbon sensor based on pCl-MA was used to determine the amount of Al (III) in the soft drink which is packed in cane and plastic bottle. The results indicated the elevation of aluminum level in the soft drink which was packed in cane because of the aluminum swept from cane to it.
Moreover, the proposed sensor was successfully applied to determine the concentration of Al (III) in different serum samples of kidney failure and Alzheimer disease patients, which are expected to have high levels of aluminum.
The obtained results indicated great agreement between the data obtained by the proposed Al-carbon sensor based on pCl-MA with those obtained by ICP as presented in Table 3.
Comparison study
The proposed sensor presented an acceptable and good performance when compared with the previously published Al (III) - ISE concerning linear ranges Nernstian slope, limit of detection (LOD), response time and pH, as shown in Table 4.
Conclusion
Al (III)-carbon sensor based on p-chlorophenyl maleanilic acid (pCl-MA) was developed. The proposed sensor exhibits precise, fast response of 8 s., reproducible and selective potentiometric investigation of aluminum ions. It shows trivalent Nernstian slope (20.32 ± 1.18) over a linear range 1.0 × 10− 6 −1.0 × 10− 1 mol L− 1 and high selectivity toward aluminum ions over the other cationic species. Moreover, the developed sensor successfully applied to detect the trace level of aluminum in real samples such as pharmaceutical, canned beverages and biological fluids of kidney failure and Alzheimer disease patients, without the need to complicated instrumentation or extra separation techniques like coprecipitation and ion extraction.
Data availability
The data supporting this article has been included as part of the manuscript.
References
Zayed, M. A., Abbas, A. A., Mahmoud, W. H., Ali, A. E. & Mohamed, G. G. Development and surface characterization of a Bis (aminotriazoles) derivative based renewable carbon paste electrode for selective potentiometric determination of cr (III) ion in real water samples. Microchem. J. 159, 105478 (2020).
Zatta, P., Drago, D., Bolognin, S. & Sensi, S. L. Alzheimer’s disease, metal ions and metal homeostatic therapy. Trends Pharmacol. Sci. 30 (7), 346–355 (2009).
Good, P. F., Olanow, C. W. & Perl, D. P. Neuromelanin-containing neurons of the substantia Nigra accumulate iron and aluminum in parkinson’s disease: a LAMMA study. Brain Res. 593 (2), 343–346 (1992).
Parkinson, I. S., Ward, M. K. & Kerr, D. N. Dialysis encephalopathy, bone disease and anaemia: the aluminum intoxication syndrome during regular haemodialysis. J. Clin. Pathol. 34 (11), 1285 (1981).
Fawzy, H., Zayed, M. A. & Frag, E. Y. Chemically modified Flubendazole carbon electrochemical sensor for aluminum determination in food stuff, multivitamin syrup and real water samples supported by DFT calculations, IR and morphological tools. Food Chem. 421, 136197 (2023).
Hue, N. Soil acidity: Development, impacts, and management. In Structure and Functions of Pedosphere 1st edn. 103–131 (Springer Nature Singapore, 2022).
Kawahara, M., Muramoto, K., Kobayashi, K., Mori, H. & Kuroda, Y. Aluminum promotes the aggregation of Alzheimer′ s amyloid β-Protein in vitro. Biochem. Biophys. Res. Commun. 198 (2), 531–535 (1994).
Paik, S. R., Lee, J. H., Kim, D. H., Chang, C. S. & Kim, J. Aluminum-induced structural alterations of the precursor of the non-Aβ component of alzheimer’s disease amyloid. Arch. Biochem. Biophys. 344 (2), 325–334 (1997).
Lin, J. L., Kou, M. T. & Leu, M. L. Effect of long-term low-dose aluminum-containing agents on hemoglobin synthesis in patients with chronic renal insufficiency. Nephron 74 (1), 33–38 (1996).
Koo, Y. J. et al. Determination of toxic metal release from metallic kitchen utensils and their health risks. Food Chem. Toxicol. 145, 111651 (2020).
Sanajou, S., Şahin, G. & Baydar, T. Aluminium in cosmetics and personal care products. J. Appl. Toxicol. 41 (11), 1704–1718 (2021).
Rains, T. C. Determination of aluminum, barium, calcium, lead, magnesium, and silver in ferrous alloys by atomic emission and absorption spectrometry. In New Analytical Techniques for Trace Constituents of Metallic and Metal-Bearing Ores (ASTM International, 1981).
Johansson, C. G. & Analysis Committee of the European Brewery Convention. DETERMINATION OF ALUMINIUM IN BEER BY GF-AAS. J. Inst. Brew. 100 (4), 255–255 (1994).
Craney, C. L., Swartout, K., Smith, F. W. & West, C. D. Improvement of trace aluminum determination by electrothermal atomic absorption spectrophotometry using phosphoric acid. Anal. Chem. 58 (3), 656–658 (1986).
Hoch, R. L. Determination of aluminum in beverages by automated non-segmented continuous flow analysis with fluorescent detection of the lumogallion complex. Analyst 124 (5), 793–796 (1999).
Sanz-Medel, A., Cabezuelo, A. B. S., Milačič, R. & Polak, T. B. The chemical speciation of aluminium in human serum. Coord. Chem. Rev. 228 (2), 373–383 (2002).
Yao, H., Wang, S., Ma, X., Ren, L. & Yan, F. A novel aluminum (III)-selective PVC membrane electrode based on a schiff base complex of Bis (5-sulphonate salicylaldehyde) 2, 3-diaminobenzene. Int. J. Electrochem. Sci. 9 (4), 2158–2170 (2014).
Esmaelpourfarkhani, M., Rounaghi, G. H. & Arbab-Zavar, M. H. Construction of a new aluminum (III) cation selective electrode based on 12-crown-4 as an ionophore. J. Braz. Chem. Soc. 26 (5), 963–969 (2015).
Mohan, C., Sharma, K. & Chandra, S. Aluminium (III) selective PVC membrane sensor based on zinc complex of thiophenealdehyde thiosemicarbazone as an ionophore. Der Pharma Chem. 9 (7), 133–139 (2017).
Mizani, F. et al. Design and construction of new potentiometric sensors for determination of Al3+ ion based on (Z)-2-(2-methyl benzylidene)-1-(2, 4-dinitrophenyl) hydrazine. Mater. Sci. Engineering: C. 49, 861–868 (2015).
Arvand, M., Kermanian, M. & Zanjanchi, M. A. Direct determination of aluminium in foods and pharmaceutical preparations by potentiometry using an AlMCM-41 modified polymeric membrane sensor. Electrochim. Acta. 55 (23), 6946–6952 (2010).
Himeno, Y., Hotatsu, T., Arisawa, K., Koshimura, S. & Hirata, R. Fundamental studies in chemotherapy of tuberculosis: part 49. Tuberculo-bacteriostatic experiments with o-aminophenol nacyl derivatives, and other compounds. Annu. Rep. Res. Inst. Tuberc Kanazawa Univ. 12, 5–8 (1945).
Sinikka, V. & Antero, A. ; analysis of aluminium in serum and urine for the biomonitoring of occupational exposure. Sci. Total Environ. 199, 103–110 (1997).
Magalhães, C. G., Liles, A., Rocha, K. L., da Silva, C. A. & J. B. B Direct determination of aluminium in serum and urine by electrothermal atomic absorption spectrometry using ruthenium as permanent modifier. Anal. Chim. Acta. 464, 323–330 (2002).
Janez, S., Radmila, M., Ingrid, F., Maja, C. & Peter, B. Use of nitric acid in sample pretreatment for determination of trace elements in various biological samples by ETAAS. J. Pharm. Biomed. Anal. 22, 993–1002 (2000).
Taylor, A. & Walker, A. W. Ann. Clin. Biochem. 29, 377–389 (1992).
Sanz-Medel, A., Rodrigues Roza, R., Gonzales Alonso, R., Vallina, N. & Canatta, A. J J. Anal. Spectrom. 2, 177–184. (1987).
Taylor, A., Starkey, B. J. & Walkers, A. Ann. Clin. Biochem. 22, 351–358 (1985).
Zayed, M. A., El-Desawy, M. & Eladly, A. A. Experimental and theoretical spectroscopic studies in relation to molecular structure investigation of Para chloro, Para fluoro and Para nitro maleanilinic acids. Comput. Biol. Chem. 76, 338–356 (2018).
Abdel Basset, M. H. Structure investigation of synthesized novel Retinoid-Iodine derivatives and its biological applications. Egypt. J. Chem. 64 (1), 211–224 (2021).
Lewenstam, A. & Hulanicki, A. Selectivity coefficients of ion-sensing electrodes. Selective Electrode Rev. 12, 161 (1990).
Fekri, M. H., Darvishpour, M. & Baghdar, E. Chromium(III) ion selective electrode based on 2H–1, 4-benzothioazine-2,3(4H) dione dioxime as a neutral carrier. J. Phys. Theo Chem. 8, 17–21 (2011).
Abu Shawish, H., Elhabiby, M., Abu Aziz, H., Saadeh, S. & Tbaza, A. Determination of trihexyphenidyl hydrochloride drug in tablets and urine using a potentiometric carbon paste electrode. Sens. Actuators B. 235, 18–26 (2016).
Shamsipur, M., Kazemi, S. Y. & Sharghi, H. Design of a selective and sensitive PVC-membrane potentiometric sensor for strontium ion based on 1, 10-diaza-5, 6-benzo-4, 7-dioxacyclohexadecane-2, 9-dione as a neutral ionophore. Sensors 7 (4), 438–447 (2007).
Bakker, E., Bühlmann, P. & Pretsch, E. Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics. Chem. Rev. 97 (8), 3083–3132 (1997).
Kumar, P., Sharma, H. K. & Shalaan, K. G. Development of chromium (III) selective potentiometric sensor by using synthesized Triazole derivative as an ionophore. J. Chem. 2013 (1), 142752 (2013).
Abu-Shawish, H. M., Saadeh, S. M., Hartani, K. & Dalloul, H. M. A comparative study of chromium (III) ion-selective electrodes based on N, N-bis (salicylidene)-o-phenylenediaminatechromium (III). J. Iran. Chem. Soc. 6, 729–737 (2009).
Ghoreishi, S. M., Behpour, M. & Nabi, M. A novel naphazoline-selective membrane sensor and its pharmaceutical applications. Sens. Actuators B. 113 (2), 963–969 (2006).
Kazemi, S. Y., Hamidi, Asanjarani, A., Zolgharnein, N. & J Optimization of a new polymeric chromium (III) membrane electrode based on Methyl Violet by using experimental design. Talanta 81 (4-5), 1681–1687 (2010).
Ying, K. S., Heng, L. Y., Hassan, N. I. & Hasbullah, S. A. A new and All-Solid-State potentiometric aluminium ion sensor for water analysis. Sensors 20, 6898 (2020).
Shamsipur, M., Ershad, S., Yari, A., ,Sharghi, H. & Salimi, A. R. Hydroxythioxanthones as suitable neutral ionophores for the Preparation of PVC-membrane potentiometric sensors for al (III) ion. Anal. Sci. 20, 301 (2004).
OZBE, O. ISILDAK, O., Use of 5,10,15,20-tetrakis(p-chlorophenyl)porphyrin as sensor material: potentiometric determination of aluminium(III) ions. Bull. Mater. Sci. (2022) 45:114 (2022).
Aglan, R. F., Mohamed, G. G. & Mohamed, H. A. Determination of aluminum (III) by using a modified carbon paste selective electrode. J. Pharm. Res. 5 (9), 4748–4754 (2012).
Nahed, A., Abd El Ghany, E. Y., Frag & Mohammed Abd El Fattah. Fabrication of chemically modified carbon paste electrode based on functionalized biopolymer for potentiometric determination of al (III) ion in real water and pharmaceutical samples. J. Iran. Chem. Soc. 15, 1987–1997 (2018).
Khoshnood, R., Akbari, S. & Chenarbou, T. M. Determination of aluminium by potentiometry using carbon paste ion selective electrode based on N,N’-bis(salicylidene)-1,3-Propanediamine ligand. J. Anal. Chem. 77, 1057–1061 (2022).
Ali, A. E., Mohamed, A. M., Zayed, M. A., Omar, M. M. & Mohamed, G. G. Exploitation of bis (tert‐butylaminomethyl) pyridine as a chemical modifier in a carbon paste electrode for sensitive and selective potentiometric determination of Al3+ ion in pharmaceutical and water samples. Electroanalysis 35 (12), 1–14 (2023).
ali, n. m. Novel Quercetin-Based Sensors for Aluminum Detection in Complex Matrices [Master’s Thesis, the American University in Cairo]. AUC Knowledge Fountain. (2025).
Sandeep, K. et al. S., Sandeep K.
Funding
Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).
Author information
Authors and Affiliations
Contributions
Marwa and Eman wrote the main manuscript text and prepared figures and tables while D. Zayed and Eman reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Basset, M.H.A., Zayed, M.A. & Frag, E.Y. Quantitative determination of al (III) traces in soft drink, pharmaceutical products and biological fluids of kidney failure and alzheimer disease patients using carbon sensor. Sci Rep 16, 626 (2026). https://doi.org/10.1038/s41598-025-32308-z
Received:
Accepted:
Published:
Version of record:
DOI: https://doi.org/10.1038/s41598-025-32308-z