Table 3 Bias, standard deviation (STDV), and limits of agreement (LoAs) in paw thickness assessment by edema severity.

From: 3D scanner’s potential as a novel tool for lymphedema measurement in mouse hindlimb models

Caliper - CT

Bias

STDV

Lower LoA

Upper LoA

None

− 0.15

0.23

− 0.60

0.30

Moderate

− 0.11

0.37

− 0.83

0.61

Severe

− 0.30

0.53

− 1.34

0.75

Scanner - CT

Bias

STDV

Lower LoA

Upper LoA

None

− 0.14

0.22

− 0.58

0.30

Moderate

− 0.05

0.14

− 0.33

0.24

Severe

− 0.10

0.17

− 0.43

0.22

  1. This table details the bias, standard deviation, and limits of agreement in the assessment of paw thickness across different levels of edema severity. The paw thickness ranged from 2.35 mm to 3.43 mm. It provides critical insights into the accuracy and consistency of paw thickness measurements under different pathological conditions, highlighting how edema severity influences measurement precision and reliability.
  2. Caliper: Paw thickness measured using the silk method.
  3. Scanner: Paw thickness measured using the scanner method.
  4. CT: Paw thickness measured using micro-CT.
  5. Unit: mm.