Table 3 SPECT acquisition protocol for GC-1 and GC-2.

From: Diagnostic reference level, achievable dose, and effective dose estimation in adult hybrid SPECT myocardial perfusion imaging in Institut Jantung Negara

System

GE NM 530c (GC-1)

GE Ventri (GC-2)

Collimator

19 Pinhole Collimators

Low Energy High Resolution (LEHR)

Radionuclide

Tc-99 m

Tc-99 m

Energy window width

140.5 ± 10% keV

140 ± 10% keV

Phase

Stress

Rest

Stress

Rest

Protocol

SGATE LD

RGATE LD

SGATE EVOLUTION

RGATE EVOLUTION

Gated

Yes

Yes

Yes

Yes

Position

Supine

Supine

Supine

Supine

Contour

Not Applicable

Not Applicable

Circular

Circular

Acquisition Mode

Static

Static

Step and Shoot, 3 intervals over a 180° arc

Step and Shoot, 3 intervals over a 180° arc

Angle

315° (fixed)

315° (fixed)

45° right anterior oblique to 45° left posterior oblique

45° right anterior oblique to 45° left posterior oblique

Direction

Not Applicable

Not Applicable

Counter-clockwise

Counter-clockwise

Gating parameter

8 frames per cycle

8 frames per cycle

8 frames per cycle

8 frames per cycle

Zoom

1.00

1.00

1.00

1.00

Matrix size

32 × 32

32 × 32

64 × 64

64 × 64

Time per view

5 min

3 min

13 s

10 s

Total view

Not applicable

Not applicable

30 steps, 60 views

30 steps, 60 views

Total acquisition time

5 min

3 min

7 min 7 s

8 min 40 s